Isotope-dilution-LC-MS/MS candidate reference measurement procedure for cefepime in human serum

被引:0
|
作者
Schaeffler, Judith [1 ]
Vogeser, Michael [1 ]
Habler, Katharina [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, LMU Univ Hosp, Inst Lab Med, Munich, Germany
关键词
Cefepime; Isotope dilution-liquid chromatography-tan-; dem mass spectrometry (ID-LC-MS/MS); Reference measurement procedures; Metrological traceability; Standardization; METROLOGICAL TRACEABILITY; QUANTIFICATION; CIPROFLOXACIN; PIPERACILLIN; MEROPENEM;
D O I
10.1016/j.jmsacl.2024.08.001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Reference measurement procedures are an essential element in the standardization and comparability of analytical measurement results in laboratory medicine. No LC-MS/MS-based reference measurement procedure for cefepime in serum has been published previously. Materials and methods: An isotope-dilution based two-dimensional LC-MS/MS reference measurement procedure for cefepime concentrations in human serum was developed and tested. The value assignment of unknown samples is based on a defined measurement series validation. Six unknown samples can be measured per series. Pass criteria for the run and the samples were determined empirically based on a performance evaluation. For this purpose, a between-run determination of five runs of the defined measurement series with six cefepime samples was carried out and evaluated. The goal was to define rigorous, realistic target limits and minimize measurement uncertainty. The final defined target limits are used for series-based validation and value assignment. The results for the six unknown samples are provided with the associated measurement uncertainty for this series. Results: The developed and extensively studied measurement procedure for the quantification of cefepime in serum was found to be practicable and fit for its purpose. The between-run mean imprecision of the six cefepime samples was <= 2.0 %, for the QCs it was <= 2.3 % and the between-run mean inaccuracy of the QCs was within +/- 1.1 %. Conclusion: The novel isotope-dilution-LC-MS/MS measurement procedure in accordance to ISO 15193 can be recommended as candidate reference measurement procedure for the value assignment of cefepime concentrations in human serum.
引用
收藏
页码:43 / 48
页数:6
相关论文
共 50 条
  • [31] Development of a candidate reference measurement procedure by ID-LC-MS/MS for total tau protein measurement in cerebrospinal fluid (CSF)
    Giangrande, Chiara
    Vaneeckhoutte, Helene
    Boeuf, Amandine
    Lalere, Beatrice
    Hirtz, Christophe
    Lehmann, Sylvain
    Quaglia, Milena
    Delatour, Vincent
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (07) : 1235 - 1244
  • [32] Isotope Dilution-LC-MS/MS Reference Method Assessment of Serum Folate Assay Accuracy and Proficiency Testing Consensus Mean
    Blackmore, Sheena
    Pfeiffer, Christine M.
    Lee, Anne
    Fazili, Zia
    Hamilton, Malcolm S.
    CLINICAL CHEMISTRY, 2011, 57 (07) : 986 - 994
  • [33] MEASUREMENT OF HUMAN CORTISOL PRODUCTION-RATE BY ISOTOPE-DILUTION THERMOSPRAY LC MS
    YERGEY, AL
    ESTEBAN, NV
    LOUGHLIN, T
    LORIAUX, DL
    APPLIED RADIATION AND ISOTOPES, 1988, 39 (06) : 523 - 523
  • [34] Development of LC-MS-MS and SPE-LC-MS-MS methods for measurement of antidepressants in human serum
    Webb, Walter
    Breaud, Autumn
    Clarke, William
    THERAPEUTIC DRUG MONITORING, 2007, 29 (04) : 520 - 520
  • [35] Comparison of a routine LC-MS/MS assay for testosterone against an isotope dilution GCMS reference method
    Keevil, B. G.
    Owen, L. J.
    Blincko, S.
    Ramsay, C. S.
    Sluss, P. M.
    Van Uytfanghe, K.
    Thienpont, L. M.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A104 - A104
  • [36] Measurement of Acetylcholine in Rat Brain Microdialysates by LC–Isotope Dilution Tandem MS
    Laszlo Prokai
    Petr Fryčák
    Stanley M. Stevens
    Vien Nguyen
    Chromatographia, 2008, 68 : 101 - 105
  • [37] Absolute quantitation of human serum cystatin C: candidate reference method by 15N-labeled recombinant protein isotope dilution UPLC-MS/MS
    Zhang, Qiaoxuan
    Zhan, Min
    Peng, Xiongqiang
    Jin, Xing
    Yan, Jun
    Zhang, Pengwei
    Zhuang, Junhua
    Han, Liqiao
    Huang, Xianzhang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024,
  • [38] Simultaneous quantification of cefepime, meropenem, ciprofloxacin, moxifloxacin, linezolid and piperacillin in human serum using an isotope-dilution HPLC-MS/MS method
    Paal, M.
    Zoller, M.
    Schuster, C.
    Vogeser, M.
    Schuetze, G.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 152 : 102 - 110
  • [39] Isotope dilution LC/MS/MS for the detection of nerve agent exposure in urine
    Ciner, Frederic L.
    McCord, Carla E.
    Plunkett, Roy W., Jr.
    Martin, Michael F.
    Croley, Timothy R.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 846 (1-2): : 42 - 50
  • [40] Target analyte quantification by isotope dilution LC-MS/MS directly referring to internal standard concentrations - validation for serum cortisol measurement
    Maier, Barbara
    Vogeser, Michael
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (04) : 833 - 837