Dual-chamber leadless pacemaker in complex adult congenital heart disease: a case report

被引:1
作者
Ngan, Ho Ting Abe [1 ]
Fabbricatore, Davide [1 ]
Regan, William [2 ]
Rosenthal, Eric [2 ]
Wong, Tom [1 ]
机构
[1] Royal Brompton & Harefield Hosp, Guys & St Thomas NHS Fdn Trust, Dept Cardiac Electrophyiosl, Sydney St, London SW3 6NP, England
[2] Evelina London Childrens Hosp, Guys & St ThomasNHS Fdn Trust, Dept Cardiac Electrophysiol, Westminster Bridge Rd, London SE1 7EH, England
关键词
ACHD; Atrioventricular synchrony; Dual-chamber pacing; Leadless pacemaker; Univentricular heart; Case report; DEFIBRILLATOR; EXPERIENCE;
D O I
10.1093/ehjcr/ytae506
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Atrioventricular block is common with adult congenital heart disease and pacemaker implantation is challenging. Atrioventricular synchronous pacing is important for better haemodynamics. This case reports the implantation of a dual-chamber leadless pacemaker in a patient with univentricular heart physiology and contributes to the literature regarding the management option in complex adult congenital heart disease patients with conduction abnormalities.Case summary A 25-year-old male with double inlet left ventricular, transposition of great arteries, hypoplastic aortic arch receive multiple surgeries including the Glenn shunt at the age of 1. He presented with 2:1 and 3:1 heart block at the age of 13 with a transvenous dual-chamber pacemaker implanted by pacing the superior vena cava stump and puncturing the Glenn shunt for the ventricular lead. A decade later, lead malfunctioned and the patient progressed to complete heart block. A subcutaneous implantable cardioverter defibrillator was implanted when he was 23 for monomorphic ventricular tachycardia. Given the anticipated challenges with transvenous lead extraction and epicardial pacemaker implantation, we implanted the novel dual-chamber leadless pacemakers which resulted in satisfactory atrioventricular synchronous pacing performance immediately post-op and 2 weeks after the procedure.Discussion We present a case of a novel dual-chamber leadless pacemaker implantation to maintain atrioventricular synchrony in the patient with complete heart block and univentricular physiology. This case illustrates an additional pacing option in complex adult congenital heart to maintain atrioventricular synchrony.
引用
收藏
页数:6
相关论文
共 15 条
[1]   Ambulatory atrioventricular synchronous pacing over time using a leadless ventricular pacemaker: Primary results from the AccelAV study [J].
Chinitz, Larry A. ;
El-Chami, Mikhael F. ;
Sagi, Venkata ;
Garcia, Hector ;
Hackett, Kevin ;
Leal, Miguel ;
Whalen, Patrick ;
Henrikson, Charles A. ;
Greenspon, Arnold J. ;
Sheldon, Todd ;
Stromberg, Kurt ;
Wood, Nicole ;
Fagan, Dedra H. ;
Chan, Joseph Yat Sun .
HEART RHYTHM, 2023, 20 (01) :46-54
[2]  
Chubb H, 2016, ARRHYTH ELECTROPHYSI, V5, P57, DOI [10.15420/aer.2016.2.3, 10.15420/AER.2016.2.3]
[3]  
Ferrero Paolo, 2016, Indian Pacing Electrophysiol J, V16, P201, DOI 10.1016/j.ipej.2016.10.007
[4]   Concomitant Use of the Subcutaneous Implantable Cardioverter Defibrillator and a Permanent Pacemaker [J].
Huang, Jason ;
Patton, Kristen K. ;
Prutkin, Jordan M. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2016, 39 (11) :1240-1245
[5]   Clinical experience of entirely subcutaneous implantable cardioverterdefibrillators in children and adults: cause for caution [J].
Jarman, Julian W. E. ;
Lascelles, Karen ;
Wong, Tom ;
Markides, Vias ;
Clague, Jonathan R. ;
Till, Janice .
EUROPEAN HEART JOURNAL, 2012, 33 (11) :1351-1359
[6]   Canadian trial of physiological pacing - Effects of physiological pacing during long-term follow-up [J].
Kerr, CR ;
Connolly, SJ ;
Abdollah, H ;
Roberts, RS ;
Gent, M ;
Yusuf, S ;
Gillis, AM ;
Tang, ASL ;
Talajic, M ;
Klein, GJ ;
Newman, DM .
CIRCULATION, 2004, 109 (03) :357-362
[7]   A Dual-Chamber Leadless Pacemaker [J].
Knops, Reinoud E. ;
Reddy, Vivek Y. ;
Ip, James E. ;
Doshi, Rahul ;
Exner, Derek V. ;
Defaye, Pascal ;
Canby, Robert ;
Bongiorni, Maria Grazia ;
Shoda, Morio ;
Hindricks, Gerhard ;
Neuzil, Petr ;
Rashtian, Mayer ;
Breeman, Karel T. N. ;
Nevo, Jordan R. ;
Ganz, Leonard ;
Hubbard, Chris ;
Cantillon, Daniel J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2023, 388 (25) :2360-2370
[8]   EFFECT OF ATRIOVENTRICULAR SYNCHRONY ON STROKE VOLUME DURING VENTRICULAR-TACHYCARDIA IN MAN [J].
MALONEY, J ;
KHOURY, D ;
SIMMONS, T ;
WILKOFF, B ;
MORANT, V ;
TROHMAN, R ;
CASTLE, L .
AMERICAN HEART JOURNAL, 1992, 123 (06) :1561-1568
[9]   Pacing in congenital heart disease - A four-decade experience in a single tertiary centre [J].
Midha, Disha ;
Chen, Zhong ;
Jones, David G. ;
Williams, Howell J. ;
Lascelles, Karen ;
Jarman, Julian ;
Clague, Jonathan ;
Till, Janice ;
Dimopoulos, Konstatinos ;
Babu-Narayan, Sonya V. ;
Markides, Vias ;
Gatzoulis, Michael A. ;
Wong, Tom .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 241 :177-181
[10]   Conduction System Pacing Versus Conventional Cardiac Resynchronization Therapy in Congenital Heart Disease [J].
Moore, Jeremy P. ;
de Groot, Natasja M. S. ;
Connor, Matthew ;
Cortez, Daniel ;
Su, Jonathan ;
Burrows, Austin ;
Shannon, Kevin M. ;
Leary, Edward T. O. ;
Shah, Maully ;
Khairy, Paul ;
Atallah, Joseph ;
Wong, Tom ;
Lloyd, Michael S. ;
Taverne, Yannick J. H. J. ;
Dubin, Anne M. ;
Nielsen, Jens C. ;
Evertz, Reinder ;
Czosek, Richard J. ;
Madhavan, Malini ;
Chang, Philip M. ;
Aydin, Alper ;
Cano, Oscar .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2023, 9 (03) :385-393