Development of a rat biomagnetic measurement system using a high-TC SQUID magnetometer

被引:6
作者
Kim, In-Seon [1 ]
Lee, Chul-Ho [2 ]
Lee, Yong-Ho [1 ]
机构
[1] Korea Res Inst Stand & Sci, Taejon 305600, South Korea
[2] Korea Res Inst Biosci & Biotechnol, Taejon 305806, South Korea
关键词
HEART;
D O I
10.1088/0953-2048/23/8/085001
中图分类号
O59 [应用物理学];
学科分类号
摘要
We have developed a rat magnetocardiograph (MCG) system employing a high-T-C SQUID magnetometer and a tabletop magnetic shield. We obtained clear MCG signals from a healthy Wistar Kyoto rat with a relatively high peak amplitude of 50 pT by virtue of the small gap cryostat developed in this study. Well defined P-, QRS- and T-waves were observed on the MCG of the healthy rat. In the case of a spontaneously hypertensive rat measurement, the MCG showed quite a disturbed wave pattern thought to be caused by the hypertensive heart abnormality. The results suggest that the rat biomagnetic measurement system has a strong potential for monitoring the progress of the heart disease model.
引用
收藏
页数:5
相关论文
共 8 条
[1]   Magnetocardiogram of a rat observed by a high-Tc SQUID sensor system [J].
Aita, Y ;
Kambe, S ;
Nagaishi, T ;
Itozaki, H .
PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2002, 378 (PART 2) :1400-1403
[2]  
Bazett HC, 1920, HEART-J STUD CIRC, V7, P353
[3]   Cine magnetic resonance microscopy of the rat heart using cardiorespiratory-synchronous projection reconstruction [J].
Brau, ACS ;
Hedlund, LW ;
Johnson, GA .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (01) :31-38
[4]   Contactless magnetocardiographic mapping in anesthetized Wistar rats: evidence of age-related changes of cardiac electrical activity [J].
Brisinda, Donatella ;
Caristo, Maria Emiliana ;
Fenici, Riccardo .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2006, 291 (01) :H368-H378
[5]   Correlation of repolarization of ventricular monophasic action potential with ECG in the murine heart [J].
Danik, S ;
Cabo, C ;
Chiello, C ;
Kang, S ;
Wit, AL ;
Coromilas, J .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2002, 283 (01) :H372-H381
[6]  
Goldberger AL., 1999, CLIN ELECTROCARDIOGR
[7]   High-TC SQUID magnetometers for low noise measurements of magnetocardiograms [J].
Kim, IS ;
Yu, KK ;
Lee, YH ;
Kim, KW ;
Park, YK .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2005, 15 (02) :652-655
[8]  
Kuriki S, 2004, CHINESE J PHYS, V42, P501