Novel Use of Motor-Evoked Potential Monitoring During Magnetic Resonance Imaging-Guided Soft-Tissue Cryoablation: A Case Report

被引:0
作者
de Forcrand, Claire [1 ]
Thompson, Scott M. [2 ]
Oishi, Tatsuya [3 ]
Woodrum, David A. [2 ]
Adamo, Daniel A. [2 ]
Lu, Aiming [2 ]
Favazza, Chris P. [2 ]
Hoffman, Ernest M. [3 ]
Pasternak, Jeffrey J. [1 ]
Powell, Garret M. [2 ]
Teixeira, Miguel T. [1 ]
机构
[1] Mayo Clin, Dept Anesthesiol & Perioperat Med, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Neurol, Rochester, MN 55905 USA
关键词
TRANSCRANIAL ELECTRICAL-STIMULATION; SAFETY;
D O I
10.1213/XAA.0000000000001832
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Motor-evoked potential (MEP) monitoring is an electrophysiologic technique useful for testing peripheral motor nerve integrity during cryoablation cases with risk of nerve injury. Previously, neuromonitoring within the magnetic resonance imaging (MRI) suite for cryoablation has not been performed as magnetic needles are used which could cause magnetic field interactions with neuromonitoring leads. We present the first report of a patient who underwent MEP monitoring during MRI-guided cryoablation of a vascular malformation adjacent to the brachial plexus. We demonstrate that MEPs may be safely and accurately performed by interleaving MRI and MEPs during treatment, reducing the risk of postprocedural complications.
引用
收藏
页数:4
相关论文
共 18 条
[1]   Practice Advisory on Anesthetic Care for Magnetic Resonance Imaging An Updated Report by the American Society of Anesthesiologists Task Force on Anesthetic Care for Magnetic Resonance Imaging [J].
Borghi, Luca .
ANESTHESIOLOGY, 2015, 122 (03) :495-520
[2]   Optimal remifentanil dosage for providing excellent intubating conditions when co-administered with a single standard dose of propofol [J].
Bouvet, L. ;
Stoian, A. ;
Rimmele, T. ;
Allaouchiche, B. ;
Chassard, D. ;
Boselli, E. .
ANAESTHESIA, 2009, 64 (07) :719-726
[3]   Neurophysiologic Intraoperative Monitoring for Spine Surgery: A Practical Guide From Past to Present [J].
Buhl, Lauren K. ;
Bastos, Andres Brenes ;
Pollard, Richard J. ;
Arle, Jeffrey E. ;
Thomas, George P. ;
Song, Yinchen ;
Boone, M. Dustin .
JOURNAL OF INTENSIVE CARE MEDICINE, 2021, 36 (11) :1237-1249
[4]   Thermal injuries associated with MRI [J].
Dempsey, MF ;
Condon, B .
CLINICAL RADIOLOGY, 2001, 56 (06) :457-465
[5]  
Dias S, 2018, WORLD NEUROSURG, V115, pE637, DOI [10.1016/j.wneu.2018.04.123, 10.1016/J.WNEU.2018.04.123]
[6]   Intraoperative facial motor evoked potential monitoring with transcranial electrical stimulation during skull base surgery [J].
Dong, CCJ ;
MacDonald, DB ;
Akagami, R ;
Westerberg, B ;
AlKhani, A ;
Kanaan, I ;
Hassounah, M .
CLINICAL NEUROPHYSIOLOGY, 2005, 116 (03) :588-596
[7]   Safety in magnetic resonance units: an update [J].
Farling, P. A. ;
Flynn, P. A. ;
Darwent, G. ;
De Wilde, J. ;
Grainger, D. ;
King, S. ;
McBrien, M. E. ;
Menon, D. K. ;
Ridgway, J. P. ;
Sury, M. ;
Thornton, J. ;
Wilson, S. R. .
ANAESTHESIA, 2010, 65 (07) :766-770
[8]   ACR guidance document on MR safe practices: 2013 [J].
Kanal, Emanuel ;
Barkovich, A. James ;
Bell, Charlotte ;
Borgstede, James P. ;
Bradley, William G., Jr. ;
Froelich, Jerry W. ;
Gimbel, J. Rod ;
Gosbee, John W. ;
Kuhni-Kaminski, Ellisa ;
Larson, Paul A. ;
Lester, James W., Jr. ;
Nyenhuis, John ;
Schaefer, Daniel Joe ;
Sebek, Elizabeth A. ;
Weinreb, Jeffrey ;
Wilkoff, Bruce L. ;
Woods, Terry O. ;
Lucey, Leonard ;
Hernandez, Dina .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 37 (03) :501-530
[9]   A practical guide for anesthetic management during intraoperative motor evoked potential monitoring [J].
Kawaguchi, Masahiko ;
Iida, Hiroki ;
Tanaka, Satoshi ;
Fukuoka, Naokazu ;
Hayashi, Hironobu ;
Izumi, Shunsuke ;
Yoshitani, Kenji ;
Kakinohana, Manabu .
JOURNAL OF ANESTHESIA, 2020, 34 (01) :5-28
[10]   Intraoperative Neurophysiologic Monitoring: Basic Principles and Recent Update [J].
Kim, Sung-Min ;
Kim, Seung Hyun ;
Seo, Dae-Won ;
Lee, Kwang-Woo .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2013, 28 (09) :1261-1269