Effects of ionization chamber orientation in photon and neutron beams

被引:0
作者
Kirjanen, Emmi [1 ,2 ]
Wendland, Lauri [1 ,2 ]
Porra, Liisa [1 ,2 ]
Seppala, Tiina [1 ,2 ]
Tenhunen, Mikko [1 ,2 ]
机构
[1] Helsinki Univ Hosp HUS, Comprehens Canc Ctr, Haartmaninkatu 4, Helsinki 00290, Finland
[2] Univ Helsinki, Haartmaninkatu 4, Helsinki 00290, Finland
基金
芬兰科学院;
关键词
Photon beam; Neutron beam; Dosimetry; Ionization chamber; BNCT; Quality assurance; DETECTORS; PHANTOM;
D O I
10.1016/j.nima.2024.169553
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Boron neutron capture therapy (BNCT) is a biologically targeted radiotherapy modality that utilizes a neutron beam. Ionization chambers are used in neutron beam dosimetry which is part of the commissioning process and periodic quality assurance. Ionization chambers used in BNCT are a bit larger in size compared to conventional external beam radiotherapy counterparts and a fixed horizontal beam direction poses its challenges among other considerations. Because of this, more flexible ionization chamber positioning would be preferred even in a large water phantom. In this work, we explain the effects and uncertainties that depend on chamber orientation. We compared depth dose curves measured with Exradin T2 and M2 ion chambers with two radiation sources, 6X photon and epithermal neutron beam, and in two orientations; one where the ionization chamber rod was positioned in perpendicular and the other where it was positioned in parallel orientation compared to the beam central axis. The perpendicular orientation is the reference orientation used in calibration. However, the parallel orientation allowed us to measure depth dose closer to the surface and the measurements had smaller systematic uncertainties because of a more straightforward alignment process. We can convert the data between the two different orientations with two simple correction parameters; correction factor which adjusted amplitude (k) and shift which adjusted the change of the effective point position in depth direction (Delta). All correction factor results were within 1.000 +/- 0.015 and all shift results were smaller or the same order of magnitude compared to the positioning accuracy, 0.05mm - 0.5mm. Even without the usage of corrections, the measurements were already almost in agreement. Because of this, we would recommend taking into consideration 2 % additional uncertainty if measurements are performed in parallel orientation and no such correction parameters are used.
引用
收藏
页数:8
相关论文
共 19 条
[1]  
Absorbed Dose Determination in External Beam Radiotherapy, 2001, Technical Reports Series
[2]  
[Anonymous], 2023, boron neutron capture therapy based on generative adversarial network
[3]  
[Anonymous], 2012, DOC 80652-00
[4]   A toolkit for epithermal neutron beam characterisation in BNCT [J].
Auterinen, I ;
Serén, T ;
Uusi-Simola, J ;
Kosunen, A ;
Savolainen, S .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :587-593
[5]   Evaluation of scanning resolution, detector choice and detector orientation to be used for accurate and time-efficient commissioning of a 6MV clinical linear accelerator [J].
Azhar, Deeba ;
Gul, Attia ;
Javid, Muhamad Arshad ;
Hussain, Muhammad Mubashar ;
Shehzadi, Nazia Neelam .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2023, 62 (01) :83-96
[6]   Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel [J].
Casar, Bozidar ;
Gershkevitsh, Eduard ;
Mendez, Ignasi ;
Jurkovic, Slaven ;
Huq, M. Saiful .
MEDICAL PHYSICS, 2020, 47 (01) :242-259
[7]   Small field output factors: Comparison of measurements with various detectors and effects of detector orientation with primary jaw setting [J].
Godson, Henry Finlay ;
Ravikumar, M. ;
Ganesh, K. M. ;
Sathiyan, S. ;
Ponmalar, Y. Retna .
RADIATION MEASUREMENTS, 2016, 85 :99-110
[8]  
James F., 1998, Technical Report
[9]   A Review of Boron Neutron Capture Therapy: Its History and Current Challenges [J].
Jin, Will H. ;
Seldon, Crystal ;
Butkus, Michael ;
Sauerwein, Wolfgang ;
Giap, Huan B. .
INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2022, 9 (01) :71-82
[10]   Twin ionisation chambers for dose determinations in phantom in an epithermal neutron beam [J].
Kosunen, A ;
Kortesniemi, M ;
Ylä-Mella, H ;
Seppälä, T ;
Lampinen, J ;
Serén, T ;
Auterinen, I ;
Järvinen, H ;
Savolainen, S .
RADIATION PROTECTION DOSIMETRY, 1999, 81 (03) :187-194