Establishment of institutional Diagnostic Reference Levels for Computed Tomography examination at the University of Ghana Medical Centre

被引:0
|
作者
Issahaku, Shirazu [1 ,2 ]
Amoh, Simon Mensah [3 ]
Acquah, Isaac Kwesi [4 ]
Nunoo, George [3 ]
Sackey, Theophilus A. [1 ,2 ]
机构
[1] Ghana Atom Energy Commiss, Accra, Ghana
[2] Univ Ghana, Dept Med Phys, Legon, Accra, Ghana
[3] Univ Ghana, Med Ctr, Dept Radiol, Accra, Ghana
[4] Univ Educ, Dept Phys Educ, Winneba, Ghana
关键词
Diagnostic reference levels; Computed tomography dose index; Computed tomography; Radiation exposure; Dose-length product; Radiation protection; CT EXAMINATIONS; ADULT; OPTIMIZATION; CHEST; HEAD;
D O I
10.1007/s12553-024-00902-2
中图分类号
R-058 [];
学科分类号
摘要
IntroductionDiagnostic Reference Levels are important indicators for monitoring and optimizing protocol and procedure in medical imaging between facilities and equipment. This helps to evaluate whether, in routine clinical conditions, the median value obtained for a representative group of patients are within an agreed range from a specified procedure is unusually high or low for that procedure. This study aimed to propose Diagnostic Reference Levels for Computed Tomography examination of the most common routine examination of the head, chest and abdominal pelvis regions at the University of Ghana Medical Centre.MethodsThe Diagnostic Reference Levels were determined based on the investigation of the most common routine examinations including head Computed Tomography examination with and without contrast, abdominopelvic Computed Tomography examination with and without contrast, and chest Computed Tomography examination without contrast. The study was based on two dose indicators; the volumetric Computed Tomography Dose Index and Dose-Length Product.ResultsThe estimated median distribution for head Computed Tomography with contrast for volumetric-Computed Tomography dose index and Dose-Length Product were 38.33 mGy and 829.35 mGy.cm, while without contrast, were 38.90 mGy and 860.90 mGy.cm respectively. For an abdominopelvic Computed Tomography examination with contrast, the estimated volumetric-Computed Tomography dose index and Dose-Length Product values were 40.19 mGy and 2096.60 mGy.cm. While without contrast, the estimated values were 14.65 mGy and 800.40 mGy.cm respectively. Additionally, for chest Computed Tomography examination, the estimated values were: 12.75 mGy and 423.95 mGy.cm for volumetric-Computed Tomography dose index and Dose-Length Product respectively. These median values represent the proposed diagnostic reference values of the head, chest and abdominal pelvis regions.ConclusionsThe proposed Diagnostic Reference Level are comparable to recommended International Atomic Energy Agency and International Commission Radiation Protection Publication 135 and other regional published data by European Commission and Regional National Diagnostic Reference Level in Africa. These reference levels will serve as benchmarks to guide clinicians in optimizing radiation dose levels while ensuring accurate diagnostic image quality at the facility.
引用
收藏
页码:1199 / 1207
页数:9
相关论文
共 50 条
  • [1] Establishment of institutional diagnostic reference levels for 6 adult computed tomography examinations: Results from preliminary data collection
    Albahiti, Sarah K.
    Barnawi, Rashid A.
    Alsafi, Khalid
    Khafaji, Mawya
    Aljondi, Rowa
    Alghamdi, Salem Saeed
    Awan, Zuhier
    Sulieman, Abdelmoneim
    Jafer, Mustafa
    Tamam, Nissren
    Tajaldeen, Abdulrahman
    Mattar, Essam H.
    Al-Malki, Khaled Mudis
    Bradley, David
    RADIATION PHYSICS AND CHEMISTRY, 2022, 201
  • [2] Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan
    Radaideh, Khaldoon
    Al-Radaideh, Ali
    Ramli, Ramzun Maizan
    Saleh, Abdallah
    Alshayeb, Rasha
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2023, 29 (01): : 26 - 34
  • [3] The establishment of computed tomography diagnostic reference levels in Portugal
    Santos, Joana
    Foley, Shane
    Paulo, Graciano
    McEntee, Mark F.
    Rainford, Louise
    RADIATION PROTECTION DOSIMETRY, 2014, 158 (03) : 307 - 317
  • [4] An Analysis of Computed Tomography Diagnostic Reference Levels in India Compared to Other Countries
    Malik, Maajid Mohi Ud Din
    Alqahtani, Mansour
    Hadadi, Ibrahim
    AlQhtani, Abdullah G. M.
    Alqarni, Abdullah
    DIAGNOSTICS, 2024, 14 (15)
  • [5] A Contribution to the Establishment of Diagnostic Reference Levels in Computed Tomography in Brazil
    Narciso, L. D. L.
    Lima, N. W.
    Dartora, C. M.
    Marques da Silva, A. M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 737 - 740
  • [6] Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India
    Saravanakumar, A.
    Vaideki, K.
    Govindarajan, K. N.
    Jayakumar, S.
    JOURNAL OF MEDICAL PHYSICS, 2014, 39 (01) : 50 - 55
  • [7] Estimation of radiation dose and establishment of local diagnostic reference levels for computed tomography of head in pediatric population
    Priyanka
    Kadavigere, Rajagopal
    Sukumar, Suresh
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2022, 30 (05) : 983 - 991
  • [9] PATIENT DOSES IN COMPUTED TOMOGRAPHY: AN ASSESSMENT OF LOCAL DIAGNOSTIC REFERENCE LEVELS IN A LARGE TEACHING HOSPITAL
    Hadnadjev, Darka R.
    Arandjic, Danijela D.
    Stojanovic, Sanja S.
    Ciraj-Bjelac, Olivera F.
    Bozovic, Predrag M.
    Stankovic, Jelena S.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2012, 27 (03): : 305 - 310
  • [10] Establishment and utilisation of national diagnostic reference level for adult computed tomography examinations in Ghana
    Issahaku, Shiraz
    Boadu, Mary
    Inkoom, Stephen
    Hasford, Francis
    Sackey, Theophilus A.
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (06) : 564 - 571