Evaluation of imaging dose for the cone beam computed tomography system of the ring-shaped radiotherapy unit using the different International Electrotechnical Commission approaches

被引:1
作者
Wang, Qingxin [1 ,2 ,3 ]
Li, Qifeng [1 ]
Wang, Zhongqiu [2 ]
Qu, Chengbin [2 ]
Wang, Yufei [2 ]
Zhang, Wenwen [2 ]
Li, Chunyin [2 ]
Zhao, Zhonglu [2 ]
Wang, Wei [2 ]
Sa, Yu [4 ]
机构
[1] Tianjin Univ, Sch Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Key Lab Canc Prevent & Therapy, Natl Clin Res Ctr Canc, Dept Radiat Oncol,Tianjins Clin Res Ctr Canc, Tianjin 300060, Peoples R China
[3] Hetian Dist Peoples Hosp, Dept Med Oncol, Hetian 848000, Xinjiang, Peoples R China
[4] Tianjin Univ, Med Sch, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金;
关键词
Cone beam CT; Radiotherapy; Imaging dose; Computed tomography dose index; Halcyon; DNA-DAMAGE; RADIATION; CT; ACCUMULATION; DOSIMETRY; CBCT;
D O I
10.1016/j.radphyschem.2024.112165
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Purpose: Daily imaging is mandatory for the Halcyon and Ethos, which are the new ring-shaped radiotherapy units. This study quantified the imaging dose of the cone beam computed tomography (CBCT) system of the new ring-shaped radiotherapy units using the latest International Electrotechnical Commission (IEC) measurement approach. Furthermore, we investigated the influence of collimation width on the imaging dose and compared the imaging dose values obtained using three IEC measurement approaches. Methods: The IEC has proposed three approaches for measuring CTDI100: traditional, transition, and the latest measurement approach. In addition to the dose measured on the CTDI phantom, the latest IEC approach uses correction factors measured in free air for CTDI100 measurement. The imaging dose was measured for all scanning protocols using the latest IEC approach. To investigate the influence of the collimation width on the imaging dose, the associated dose metric values were evaluated and compared using different IEC approaches to increase the collimation width (2-28 cm). Furthermore, the imaging system output stability was evaluated by repeated measurements. Results: In the repeated measurements, the standard deviation of CTDIw value for the pelvis and head protocol was 0.05 mGy and 0.01 mGy, respectively. And the coefficient of variation was 0.21% and 0.23%, respectively. The measured CTDIw values of the clinical protocols were higher than the vendor-provided value, except for the pelvis protocol. The value of CTDIw changes little with increasing collimation width using the latest IEC approach. The traditional and transition approaches either underestimated or overestimated the value of CTDIw, as measured by the latest IEC approach, with increasing collimation width. Conclusions: The new ring-shaped radiotherapy units imaging system has good output stability. The CTDIw values measured for clinical scanning protocols generally exceeded the vendor-provided values. Compared with the latest IEC CTDI100 measurement approach, the traditional and transition measurement approaches were more influenced by beam width. For wide CBCT scans, rationalizing the choice of imaging dose measurement approach is important.
引用
收藏
页数:8
相关论文
共 50 条
[41]   Evaluation of four-dimensional cone beam computed tomography ventilation images acquired with two different linear accelerators at various gantry speeds using a deformable lung phantom [J].
Kadoya, Noriyuki ;
Nemoto, Hikaru ;
Kajikawa, Tomohiro ;
Nakajima, Yujiro ;
Kanai, Takayuki ;
Ieko, Yoshiro ;
Ikeda, Ryutaro ;
Sato, Kiyokazu ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 :75-83
[42]   Evaluation of the anterior dentoalveolar relationship in skeletal Class III malocclusion patients with different vertical facial patterns using cone-beam computed tomography [J].
Han, Shaobo ;
Fan, Xiangfei ;
Xiao, Danna .
ANGLE ORTHODONTIST, 2024, 94 (02) :187-193
[43]   Quantification of six-degree-of-freedom motion during beam delivery in spine stereotactic body radiotherapy using intra-irradiation cone-beam computed tomography imaging technique [J].
Iramina, Hiraku ;
Nakamura, Mitsuhiro ;
Nakamura, Kiyonao ;
Fujimoto, Takahiro ;
Mizowaki, Takashi .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 110
[44]   Evaluation of interfractional variation of the centroid position and volume of internal target volume during stereotactic body radiotherapy of lung cancer using cone-beam computed tomography [J].
Sun, Yanan ;
Ge, Hong ;
Cheng, Siguo ;
Yang, Chengliang ;
Zhu, Qianqian ;
Li, Dingjie ;
Tian, Yuan .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02) :461-472
[45]   The impact of different cone beam computed tomography and multi-slice computed tomography scan parameters on virtual three-dimensional model accuracy using a highly precise ex vivo evaluation method [J].
Matta, Ragai-Edward ;
von Wilmowsky, Cornelius ;
Neuhuber, Winfried ;
Lell, Michael ;
Neukam, Friedrich W. ;
Adler, Werner ;
Wichmann, Manfred ;
Bergauer, Bastian .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2016, 44 (05) :632-636
[46]   Comparative Evaluation of Three Different Obturating Techniques Lateral Compaction, Thermafil and Calamus for Filling Area and Voids Using Cone Beam Computed Tomography: An Invitro study [J].
Gupta, Ruchi ;
Dhingra, Anil ;
Panwar, Nidhi R. .
JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2015, 9 (08) :ZC15-ZC17
[47]   Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy [J].
Fu, Caofei ;
Ma, Changsheng ;
Shang, Dongping ;
Qiu, Qingtao ;
Meng, Huipeng ;
Duan, Jinghao ;
Yin, Yong .
TRANSLATIONAL CANCER RESEARCH, 2020, 9 (10) :6005-6012
[48]   Quantification of the kV X-ray imaging dose during real-time tumor tracking and from three- and four-dimensional cone-beam computed tomography in lung cancer patients using a Monte Carlo simulation [J].
Nakamura, Mitsuhiro ;
Ishihara, Yoshitomo ;
Matsuo, Yukinori ;
Iizuka, Yusuke ;
Ueki, Nami ;
Iramina, Hiraku ;
Hirashima, Hideaki ;
Mizowaki, Takashi .
JOURNAL OF RADIATION RESEARCH, 2018, 59 (02) :173-181
[49]   Evaluation of three-dimensional setup errors using cone beam computed tomography (CBCT) for mediastinal lymphoma treated with volumetric modulated arc radiotherapy (VMAT) and deep inspiratory breath-hold (DIBH) technique [J].
Mohanty, Samarpita ;
Pradhan, Sangram ;
Patil, Divya ;
Khwaja, Razi A. ;
Jain, Jivanshu ;
Bagul, Kamlesh ;
Tambe, Santosh ;
Kakoti, Sangeeta ;
Goda, Jayant S. .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2025, 21 (01) :104-110
[50]   Interfraction prostate displacement during image-guided radiotherapy using intraprostatic fiducial markers and a cone-beam computed tomography system: A volumetric off-line analysis in relation to the variations of rectal and bladder volumes [J].
Ingrosso, Gianluca ;
Miceli, Roberto ;
Ponti, Elisabetta ;
Lancia, Andrea ;
di Cristino, Daniela ;
de Pasquale, Francesco ;
Bove, Pierluigi ;
Santoni, Riccardo .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 :S69-S75