Background Total hip arthroplasty (THA) using a portable navigation system (PNS) incurs costs per procedure. However, it does not require a large console. This study aimed to compare the accuracy of acetabular cup placement using a pelvic alignment guide (PAG) attached to the pelvis and an accelerometer-based PNS in THA performed in the lateral decubitus position. Methods We retrospectively analyzed 100 hips that underwent primary THA in the lateral decubitus position between July 2018 and January 2021. The PAG was used in 50 hips, whereas the PNS was used in the other 50. Cup placement accuracy was measured using postoperative computed tomography scans, comparing errors in inclination and anteversion angles. The surgical time, blood loss, and complications were recorded. The follow-up period was at least 2 years in all cases. Results The mean absolute error of the inclination angle was similar between the groups (the PAG group: 3.7 degrees +/- 2.3 degrees [range, 0.0-9.0]; the PNS group: 3.7 degrees +/- 2.3 degrees [range, 0.2-10.5], p = 0.705). The mean absolute error of the anteversion angle was significantly smaller in the PAG group than in the PNS group (3.0 degrees +/- 2.4 degrees [range, 0.0-9.7] vs. 6.5 degrees +/- 4.8 degrees [range, 0.3-17.3], p < 0.001). The PAG group had a higher proportion of hips within 5 degrees and 10 degrees of the target angle (64 vs. 42%, P = 0.028, and 100 vs. 74%, p < 0.001, respectively). The PNS group had six hips with anteversion errors of 15 degrees or more. Surgical time and blood loss were lower in the PAG group. The PNS group had one dislocation, whereas the PAG group did not. Conclusion The accelerometer-based PNS did not demonstrate superior cup alignment accuracy compared to the PAG in THA performed in the lateral decubitus position. This finding informs surgeons that computer-assisted surgery is not necessarily superior to conventional THA using a PAG.
引用
收藏
页码:4849 / 4856
页数:8
相关论文
共 50 条
  • [31] Combined Anteversion Technique for Total Hip Arthroplasty With Handheld Accelerometer-Based Navigation System
    Perico, Diego Alarcon
    Warne, Christopher N.
    Lee, Sheng-Hsun
    Roberts, Heather J.
    Sierra, Rafael J.
    ARTHROPLASTY TODAY, 2023, 23
  • [32] Does Augmented Reality-based Portable Navigation Improve the Accuracy of Cup Placement in THA Compared With Accelerometer-based Portable Navigation? A Randomized Controlled Trial
    Kurosaka, Kenji
    Ogawa, Hiroyuki
    Hirasawa, Naoyuki
    Saito, Masayoshi
    Nakayama, Tsutomu
    Tsukada, Sachiyuki
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2023, 481 (08) : 1515 - 1523
  • [33] The Accuracy of Cup Placement in Total Hip Arthroplasty (THA) Using an Augmented Reality-Based Navigation System
    Shimizu, Junya
    Nagoya, Satoshi
    Kosukegawa, Ima
    Kanaizumi, Arata
    Nakahashi, Naoya
    Teramoto, Atsushi
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (04)
  • [34] Accuracy of A Handheld Accelerometer-Based Navigation System for Femoral and Tibial Resection in Total Knee Arthroplasty
    Huang, Eddie H.
    Copp, Steven N.
    Bugbee, William D.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (11) : 1906 - 1910
  • [35] Portable imageless navigation system and surgeon’s estimate for accurate evaluation of acetabular cup orientation during total hip arthroplasty in supine position
    Ryohei Takada
    Tetsuya Jinno
    Kazumasa Miyatake
    Masanobu Hirao
    Toshitaka Yoshii
    Atsushi Okawa
    European Journal of Orthopaedic Surgery & Traumatology, 2020, 30 : 707 - 712
  • [36] Cadaveric analysis of an accelerometer-based portable navigation device for distal femoral cutting block alignment in total knee arthroplasty
    Nam, Denis
    Jerabek, Seth A.
    Cross, Michael B.
    Mayman, David J.
    COMPUTER AIDED SURGERY, 2012, 17 (04) : 205 - 210
  • [37] Accuracy of cup placement using computed tomography-based navigation system in total hip arthroplasty through the direct anterior approach
    Matsuki, Yuta
    Imagama, Takashi
    Tokushige, Atsunori
    Yamazaki, Kazuhiro
    Sakai, Takashi
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2023, 28 (02) : 370 - 375
  • [38] Two- and three-dimensional measurements of leg length change using an accelerometer-based portable navigation system in total hip arthroplasty
    Masahiro Hasegawa
    Yohei Naito
    Shine Tone
    Akihiro Sudo
    BMC Musculoskeletal Disorders, 24
  • [39] Two- and three-dimensional measurements of leg length change using an accelerometer-based portable navigation system in total hip arthroplasty
    Hasegawa, Masahiro
    Naito, Yohei
    Tone, Shine
    Sudo, Akihiro
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [40] Relationship between the Accuracy of the Acetabular Cup Angle and BMI in Posterolateral Total Hip Arthroplasty with CT-Based Navigation
    Ishikura, Hisatoshi
    Nakamura, Masaki
    Nakamura, Shigeru
    Tanaka, Takeyuki
    Kawano, Hirotaka
    Tanaka, Sakae
    MEDICINA-LITHUANIA, 2022, 58 (07):