Clinical performance and application of novel serum collection " Ser-Col " device in the practice of laboratory diagnosis of infection diseases and several other immunochemical tests

被引:0
作者
Harutyunyan, Hayk [1 ]
Hovhannisyan, Alvard [2 ]
Torosyan, Hamlet [1 ]
Hakobjanyan, Gohar [3 ]
Grigoryan, Ani [1 ]
Petrosyan, Gayane [3 ]
Movsisyan, Mariam [1 ,4 ]
Poland, Dennis [5 ]
Monkel, Richard [5 ]
Lommen, Joan [5 ]
Tuaillon, Edouard [6 ]
Yenkoyan, Konstantin [1 ]
机构
[1] Yerevan State Med Univ, COBRAIN Ctr, Neurosci Lab, Yerevan 0025, Armenia
[2] Yerevan State Med Univ, Dept Infect Dis, Yerevan, Armenia
[3] Yerevan State Med Univ, Lab Diagnost Ctr Heratsi Clin Hosp, Yerevan, Armenia
[4] Yerevan State Med Univ, Dept Allergol & Clin Immunol, Yerevan, Armenia
[5] Labonovum, Limmen, Netherlands
[6] Antilles Univ, Montpellier Univ, Pathogenesis & Control Chron & Emerging Infect, Montpellier Univ Hosp,INSERM,Etablissement Francai, Montpellier, France
关键词
Ser-Col blood collection device; Serum; Viruses; Hepatitis; SARS-CoV-2; Herpes; Hormones; Vitamins; DRIED BLOOD SPOTS; QUANTIFICATION; POPULATIONS; VALIDATION;
D O I
10.1016/j.cca.2024.119970
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Dried blood collection devices might be beneficial for diagnosing infectious diseases in areas far from the medical facilities and in lockdown situations. There are several reports on the efficacy of such applications for qualitative tests. Here we demonstrated the feasibility of a novel Ser-Col blood collection device as a standardized approach for qualitative and quantitative detection of infectious markers and several over immunochemical tests. Methods: In the current study, we included 395 adult participants, 191 men and 204 women, with a median age of 41 years, as well as 75 children with a median age of 3 years. Serological status was determined by testing serum samples for three groups of infection diseases: hepatitis A and C, SARS-CoV-2, and herpes family viruses, as well as for thyroid peroxidase (TPO), prolactin, vitamin B12, and folate. Blood collected on the Ser-Col device (Labonovum) was eluted using an automated system (SCAUT Ser-Col automation, Blok System Supply) and manually. Ser-Col results were compared with serum sampled via standard venipuncture considered as the reference. Results: High correlation coefficients (r = 0.95-0.99) were observed between serum samples collected with SerCol and via standard venipuncture for the following tests: anti-HCV, anti-SARS-CoV-2 IgG, anti-HSV-2 IgG, and anti-CMV IgM. Correlation coefficients between Ser-Col and standard venipuncture serum for anti-HSV-1 IgG, anti-CMV IgG, and anti-EBV tests were relatively low (r = 0.73-0.77). Correlation coefficients for anti-TPO, prolactin, vitamin B12, and folate were also characterized with high values (r = 0.97-0.99). Conclusions: High accuracy and quantitative correlation were demonstrated between Ser-Col and samples collected by standard venipuncture. Hence, the Ser-Col blood collection device should be considered as a promising alternative for blood collection, storage, and transportation in both adult and pediatric populations.
引用
收藏
页数:8
相关论文
共 34 条
  • [1] A Review of Blood Delivery for Sustainable Supply Chain Management (BSCM)
    Alghamdi, Saleh Y.
    [J]. SUSTAINABILITY, 2023, 15 (03)
  • [2] World Medical Association Declaration of Helsinki Ethical Principles for Medical Research Involving Human Subjects
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2013, 310 (20): : 2191 - 2194
  • [3] Baird Serena, 2022, Methods Mol Biol, V2546, P391, DOI 10.1007/978-1-0716-2565-1_34
  • [4] BASSETT F, 1986, CLIN CHEM, V32, P854
  • [5] Advancements in blood collection devices
    Bush, VJ
    Leonard, L
    Szamosi, DI
    [J]. LABORATORY MEDICINE, 1998, 29 (10) : 616 - 622
  • [6] Quantification of insulin in dried blood spots
    Dowlati, B
    Dunhardt, PA
    Smith, MM
    Shaheb, S
    Stuart, CA
    [J]. JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1998, 131 (04): : 370 - 374
  • [7] Eng J., 2014, ROC analysis: web-based calculator for ROC curves [Internet]
  • [8] Clinical Course of Hyperprolactinemia in Children and Adolescents: A Review of 21 Cases
    Eren, Erdal
    Yapici, Senay
    Cakir, Esra Deniz Papatya
    Ceylan, Latife Aytekin
    Saglam, Halil
    Tarim, Omer
    [J]. JOURNAL OF CLINICAL RESEARCH IN PEDIATRIC ENDOCRINOLOGY, 2011, 3 (02) : 65 - 69
  • [9] How and when to use dried blood spots in psychoneuroendocrinological research
    Fischer, Susanne
    Obrist, Ramona
    Ehlert, Ulrike
    [J]. PSYCHONEUROENDOCRINOLOGY, 2019, 108 : 190 - 196
  • [10] Gadde Ranjeeta, 2023, Bioinformation, V19, P1043, DOI 10.6026/973206300191043