Development of a 3D-printed phantom for total skin electron therapy dose assessment

被引:1
作者
Lee, Andrew [1 ]
Alkhatib, Zaid [2 ]
Ibrahim, Mounir [2 ]
McCallum-Hee, Broderick [2 ]
Dass, Joshua [2 ]
Fernandez de Viana, Matthew [1 ]
Rowshanfarzad, Pejman [1 ,3 ]
机构
[1] Univ Western Australia, Sch Phys Math & Comp, Crawley, WA, Australia
[2] Sir Charles Gairdner Hosp, Dept Radiat Oncol, Nedlands, WA, Australia
[3] Ctr Adv Technol Canc Res CATCR, Perth, Australia
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2024年 / 25卷 / 12期
关键词
3D-printing; dosimetry; TSET; MYCOSIS-FUNGOIDES; EUROPEAN-ORGANIZATION; IRRADIATION; DOSIMETRY; RECOMMENDATIONS; CONSENSUS;
D O I
10.1002/acm2.14520
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Total skin electron therapy (TSET) is a complex radiotherapy technique, posing challenges in commissioning and quality assurance (QA), especially due to significant variability in patient body shapes. Previous studies have correlated dose with factors such as obesity index, height, and gender. However, current treatment planning systems cannot simulate TSET plans, necessitating heavy reliance on QA methods using standardized anthropomorphic phantoms and in-vivo dosimetry. Given the relatively few studies on rotational techniques, comprehensive data in commissioning could streamline the process. Methods: Developing a full-body phantom would enable a more thorough TSET commissioning process, including testing for position-specific dose distributions and comprehensive measurements across all body surfaces, unlike the typical torso-only phantoms. This was created using digital modeling software, fabricated using 3D-printing FDM technology, and filled with tissue-equivalent gelatine. The phantom was positioned at an SSD of 340 cm and irradiated with a standard rotational TSET plan using the 6E HDTSE mode on a Varian TrueBeam linac at gantry angles of +/- 18 degrees from the horizontal. The dose was measured at over 50 points across the surface using Gafchromic EBT3 film. Results: Dose distributions were generally consistent with existing literature values from in-vivo dosimetry, with several position-specific differences identified, including the hands and scalp compared to conventional positions. Hotspots were observed for the mid-dorsum of the foot and nose, with areas under 80% of the dose identified as the soles of the feet, perineum, vertex of the scalp, top of the shoulder, and palm of the hand. Additionally, analysis using an interpolated dose heatmap found that 90% of the pixel area received a dose within 10% of the prescribed dose, indicating good uniformity with the commissioned technique. Conclusions: With high agreement with the current literature, a 3D-printed phantom proves effective for measuring doses in areas typically unmeasurable in TSET commissioning.
引用
收藏
页数:10
相关论文
共 26 条
[1]  
abs, AUSTR BUREAU STAT 43
[2]   Recumbent Total Skin Electron Beam Therapy [J].
Ackerson, Bradley G. ;
Wu, Qiuwen ;
Craciunescu, Oana ;
Oyekunle, Taofik ;
Niedzwiecki, Donna ;
Gupton, Jennie ;
Laug, Patrick ;
Brumfield, Karibee ;
Crain, Erin ;
Champ, Colin E. ;
Kelsey, Chris R. .
ADVANCES IN RADIATION ONCOLOGY, 2021, 6 (04)
[3]   Utilization of thermoluminescent dosimetry in total skin electron beam radiotherapy of mycosis fungoides [J].
Antolak, JA ;
Cundiff, JH ;
Ha, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (01) :101-108
[4]   In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry [J].
Baba, Misba Hamid ;
Singh, Benoy Kumar ;
Wani, Shaqul Qamar .
JOURNAL OF MEDICAL PHYSICS, 2022, 47 (04) :362-366
[5]  
Basaglia T., 2023, ARXIVORG SELECTED AR, DOI [10.48550/arXiv.2311.10684, DOI 10.48550/ARXIV.2311.10684]
[6]   Long-term effects of total skin electron beam therapy for mycosis fungoides on hair and nail loss and regrowth [J].
Breneman, Debra L. ;
Breneman, Alyssa ;
Ballman, Elaine ;
Breneman, John C. .
JOURNAL OF DERMATOLOGICAL TREATMENT, 2022, 33 (04) :1975-1978
[7]   Clinical implementation of total skin electron beam (TSEB) therapy: A review of the relevant literature [J].
Diamantopoulos, S. ;
Platoni, K. ;
Dilvoi, M. ;
Nazos, I. ;
Geropantas, K. ;
Maravelis, G. ;
Tolia, M. ;
Beli, I. ;
Efstathopoulos, E. ;
Pantelakos, P. ;
Panayiotakis, G. ;
Kouloulias, V. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2011, 27 (02) :62-68
[8]   Monte Carlo study on dose distributions from total skin electron irradiation therapy (TSET) [J].
Ding, George X. ;
Osmundson, Evan C. ;
Shinohara, Eric ;
Newman, Neil B. ;
Price, Michael ;
Kirschner, Austin N. .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (07)
[9]   In-vivo dosimetric analysis in total skin electron beam therapy [J].
Elsayad, Khaled ;
Moustakis, Christos ;
Simonsen, Manuela ;
Baecker, Dagmar ;
Haverkamp, Uwe ;
Eich, Hans Theodor .
PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 6 :61-65
[10]  
Evans Jaden D, 2016, Adv Radiat Oncol, V1, P101, DOI 10.1016/j.adro.2016.03.005