Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment

被引:2
|
作者
Mihailovic, Jelena M. [1 ]
Bruesewitz, Michael R. [1 ]
Swicklik, Joseph R. [1 ]
Yalon, Mariana [1 ]
Rajiah, Prabhakar S. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2024年 / 25卷 / 11期
关键词
dual source computed tomography; photon-counting-detector computed tomography; pulmonary embolism; PERFUSED BLOOD-VOLUME; ENERGY CT; EMBOLISM;
D O I
10.1002/acm2.14496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA. Methods: A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images. Results: When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images. Conclusion: PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
引用
收藏
页数:8
相关论文
共 41 条
  • [21] CT pulmonary angiography: simultaneous low-pitch dualsource acquisition mode with 70kVp and 40ml of contrast medium and comparison with high-pitch spiral dual-source acquisition with automated tube potential selection
    Boos, Johannes
    Kroepil, Patric
    Lanzman, Rotem S.
    Aissa, Joel
    Schleich, Christoph
    Heusch, Philipp
    Sawicki, Lino M.
    Antoch, Gerald
    Thomas, Christoph
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1062)
  • [22] Radiation dose and image quality of high-pitch emergency abdominal CT in obese patients using third-generation dual-source CT (DSCT)
    Forbrig, Robert
    Ingrisch, Michael
    Stahl, Robert
    Winter, Katharina Stella
    Reiser, Maximilian
    Trumm, Christoph G.
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [23] Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT
    Tobias De Zordo
    Klemens von Lutterotti
    Christian Dejaco
    Peter F. Soegner
    Renate Frank
    Friedrich Aigner
    Andrea S. Klauser
    Christoph Pechlaner
    U. Joseph Schoepf
    Werner R. Jaschke
    Gudrun M. Feuchtner
    European Radiology, 2012, 22 : 279 - 286
  • [24] Image Quality and Radiation Dose of CTPA With Iodine Maps: A Prospective Randomized Study of High-Pitch Mode Photon-Counting Detector CT Versus Energy-Integrating Detector CT
    Pannenbecker, Pauline
    Heidenreich, Julius F.
    Grunz, Jan-Peter
    Huflage, Henner
    Gruschwitz, Philipp
    Patzer, Theresa S.
    Feldle, Philipp
    Bley, Thorsten A.
    Petritsch, Bernhard
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (02)
  • [25] High-pitch Helical Dual-source Computed Tomographic Pulmonary Angiography Comparing Image Quality in Inspiratory Breath-hold and During Free Breathing
    Ajlan, Amr M.
    Binzaqr, Salma
    Jadkarim, Dalia A.
    Jamjoom, Lamia G.
    Leipsic, Jonathon
    JOURNAL OF THORACIC IMAGING, 2016, 31 (01) : 56 - 62
  • [26] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lenga, Lukas
    Trapp, Franziska
    Albrecht, Moritz H.
    Wichmann, Julian L.
    Johnson, Addison A.
    Yel, Ibrahim
    D'Angelo, Tommaso
    Booz, Christian
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN RADIOLOGY, 2019, 29 (09) : 4603 - 4612
  • [27] High-pitch, Dual-source Chest Computed Tomography Angiography Without Electrocardiographic Synchronization Evaluation of Cardiac Motion Artifacts on Pulmonary Circulation
    Tacelli, Nunzia
    Darchis, Christine
    Pontana, Francois
    Faivre, Jean-Baptiste
    Deken, Valerie
    Duhamel, Alain
    Murphy, Colm
    Remy, Jacques
    Remy-Jardin, Martine
    JOURNAL OF THORACIC IMAGING, 2013, 28 (06) : 376 - 382
  • [28] The Best of Both Worlds: Ultra-high- pitch Pulmonary Angiography with Free-Breathing Technique by Means of Photon-Counting Detector CT for Diagnosis of Acute Pulmonary Embolism
    Pannenbecker, Pauline
    Heidenreich, Julius F.
    Huflage, Henner
    Gruschwitz, Philipp
    Patzer, Theresa S.
    Weng, Andreas M.
    Grunz, Jan-Peter
    Kunz, Andreas S.
    Bley, Thorsten A.
    Petritsch, Bernhard
    ACADEMIC RADIOLOGY, 2024, 31 (12) : 5280 - 5288
  • [29] Evaluation of CT angiography obstruction score and pulmonary perfusion defect score using the third-generation dual-source CT for pulmonary embolism
    Hu, H.
    Xie, P.
    Li, C.
    Ni, S.
    Wang, H.
    Li, A.
    Wei, H.
    Li, W.
    CLINICAL RADIOLOGY, 2023, 78 (09) : e627 - e634
  • [30] High-pitch non-gated scans on the second and third generation dual-source CT scanners: comparison of coronary image quality
    Lin, Cheng Ting
    Chu, Linda Chi Hang
    Zimmerman, Stefan Loy
    Fishman, Elliot K.
    CLINICAL IMAGING, 2020, 59 (01) : 45 - 49