Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment

被引:2
|
作者
Mihailovic, Jelena M. [1 ]
Bruesewitz, Michael R. [1 ]
Swicklik, Joseph R. [1 ]
Yalon, Mariana [1 ]
Rajiah, Prabhakar S. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2024年 / 25卷 / 11期
关键词
dual source computed tomography; photon-counting-detector computed tomography; pulmonary embolism; PERFUSED BLOOD-VOLUME; ENERGY CT; EMBOLISM;
D O I
10.1002/acm2.14496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA. Methods: A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images. Results: When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images. Conclusion: PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
引用
收藏
页数:8
相关论文
共 41 条
  • [1] High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT
    Ahmed, Zaki
    Campeau, David
    Gong, Hao
    Rajendran, Kishore
    Rajiah, Prabhakar
    McCollough, Cynthia
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (03) : 1428 - 1435
  • [2] Multi-energy CT imaging for large patients using dual-source photon-counting detector CT
    Tao, Shengzhen
    Marsh, Jeffrey F.
    Tao, Ashley
    Michalak, Greg J.
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17)
  • [3] Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study
    Liu, Peng
    Zhou, Shanshui
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    JAPANESE JOURNAL OF RADIOLOGY, 2025, 43 (02) : 309 - 318
  • [4] Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24)
  • [5] High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population
    Sabel, Bastian O.
    Buric, Kristijan
    Karara, Nora
    Thierfelder, Kolja M.
    Dinkel, Julien
    Sommer, Wieland H.
    Meinel, Felix G.
    PLOS ONE, 2016, 11 (02):
  • [6] Computed tomography pulmonary angiography using high-pitch dual-source scanner technology
    Al Hassan, Donya A.
    Waheed, Khawaja B.
    El Sirafy, Mohamed N.
    Khattab, Mohamad A.
    Al-Hammadi, Hassan I.
    Ibrahim, Motassem F.
    Ibrahim, Motassem F.
    Arulanantham, Zechariah J.
    SAUDI MEDICAL JOURNAL, 2019, 40 (03) : 230 - 237
  • [7] High-pitch Dual-source Computed Tomography Pulmonary Angiography in Freely Breathing Patients
    Bauer, Ralf W.
    Schell, Boris
    Beeres, Martin
    Wichmann, Julian L.
    Bodelle, Boris
    Vogl, Thomas J.
    Kerl, Josef Matthias
    JOURNAL OF THORACIC IMAGING, 2012, 27 (06) : 376 - 381
  • [8] Improved Pulmonary Artery Evaluation Using High-Pitch Photon-Counting CT Compared to High-Pitch Conventional or Routine-Pitch Conventional Dual-Energy CT
    Yalon, Mariana
    Hoodeshenas, Safa
    Chan, Alex
    Horst, Kelly K.
    Crum, Isaac
    Thorne, Jamison E.
    Lee, Yong S.
    Yu, Lifeng
    Mccollough, Cynthia H.
    Fletcher, Joel G.
    Rajiah, Prabhakar Shantha
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2024, 48 (06) : 897 - 905
  • [9] Feasibility of very low iodine dose aortoiliac CT angiography using dual-source photon-counting detector CT
    Oechsner, Tim
    Soschynski, Martin
    Schlett, Christopher L.
    Krauss, Tobias
    Schupppert, Christopher
    Mueller-Peltzer, Katharina
    Vecsey-Nagy, Milan
    Kravchenko, Dmitrij
    Varga-Szemes, Akos
    Emrich, Tilman
    Scheu, Raphael
    Taron, Jana
    Bamberg, Fabian
    Hagar, Muhammad Taha
    EUROPEAN JOURNAL OF RADIOLOGY, 2025, 183
  • [10] Ultra-low dose contrast CT pulmonary angiography in oncology patients using a high-pitch helical dual-source technology
    Rajiah, Prabhakar
    Ciancibello, Leslie
    Novak, Ronald
    Sposato, Jennifer
    Landeras, Luis
    Gilkeson, Robert
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (03) : 195 - 203