Wearable MagnetoCardioGraphy (MCG) Sensor Using a Single Coil

被引:0
作者
Kaiss, A. [1 ]
Kiourti, A. [1 ]
机构
[1] Ohio State Univ, Columbus, OH 43210 USA
来源
2024 INTERNATIONAL APPLIED COMPUTATIONAL ELECTROMAGNETICS SOCIETY SYMPOSIUM, ACES 2024 | 2024年
基金
美国国家科学基金会;
关键词
Coils; heart beats; magnetic field; magnetocardiography; sensor;
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We have previously reported a wearable magnetocardiography (MCG) sensor that utilizes a coil array capable of detecting the weak magnetic fields generated by the heart. In this work, we improve upon our previous design by reducing the number of coils from seven (7) to one (1) while also achieving significantly better detection results. Experimental validation on a human subject shows an increase in heart beat detection accuracy from 28.4% to 96.8%. Key to this improved performance is a new signal processing method that significantly reduces noise. Given the increasing importance of MCG as a reliable and non-contact method of assessing cardiac health and cognitive workload, the proposed sensor is expected to have significant clinical impact.
引用
收藏
页数:2
相关论文
共 9 条
[1]   DETECTION OF MAGNETIC FIELD OF HEART [J].
BAULE, G ;
MCFEE, R .
AMERICAN HEART JOURNAL, 1963, 66 (01) :95-&
[2]   Clinical magnetocardiography: the unshielded bet-past, present, and future [J].
Brisinda, D. ;
Fenici, P. ;
Fenici, R. .
FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
[3]   Magnetocardiography and 32-lead potential mapping: Repolarization in normal subjects during pharmacologically induced stress [J].
Brockmeier, K ;
Schmitz, L ;
Chavez, JB ;
Burghoff, M ;
Koch, H ;
Zimmermann, R ;
Trahms, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (06) :615-626
[4]   Clinical application of magnetocardiography [J].
Fenici, R ;
Brisinda, D ;
Meloni, AM .
EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2005, 5 (03) :291-313
[5]  
Mooney JW, 2017, BIOMED PHYS ENG EXPR, V3, DOI 10.1088/2057-1976/3/1/015008
[6]   Heart rate variability [J].
Stauss, HM .
AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 2003, 285 (05) :R927-R931
[7]   Quantifying Cognitive Workload Using a Non-Contact Magnetocardiography (MCG) Wearable Sensor [J].
Wang, Zitong ;
Zhu, Keren ;
Kaur, Archana ;
Recker, Robyn ;
Yang, Jingzhen ;
Kiourti, Asimina .
SENSORS, 2022, 22 (23)
[8]   A Review of Magnetic Field Emissions From the Human Body: Sources, Sensors, and Uses [J].
Zhu, Keren ;
Kiourti, Asimina .
IEEE OPEN JOURNAL OF ANTENNAS AND PROPAGATION, 2022, 3 :732-744
[9]   Miniature Coil Array for Passive Magnetocardiography in Non-Shielded Environments [J].
Zhu, Keren ;
Shah, Adhvait M. ;
Berkow, Jan ;
Kiourti, Asimina .
IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2021, 5 (02) :124-131