Comparison of apparent diffusion coefficient (ADC) values obtained by echo planar imaging diffusion-weighted imaging (DWI) and radial acquisition regime DWI in low field MRI systems: A phantom study

被引:1
作者
Sakoda, K. [1 ]
Baba, S. [1 ]
机构
[1] Kagoshima Med Technol Coll, Dept Radiol Technol, Kagoshima, Japan
关键词
Diffusion-weighted imaging (DWI); Apparent diffusion coefficient (ADC); MRI; Low-field MRI; RADAR (RADAR-DWI); Echo planar imaging DWI (EPI-DWI); TURBO SPIN-ECHO; SEGMENTED EPI; NORMAL BRAIN; RICE PADS; TUMORS; DIFFERENTIATION; LESIONS; AGE;
D O I
10.1016/j.radi.2024.07.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Diffusion-weighted imaging (DWI) with radial acquisition regime (RADAR; RADAR-DWI) is a fast spin echo (FSE)-based DWI imaging technique that is known to be robust to magnetic susceptibility artifacts and distortions as compared with echo planar imaging DWI (EPI-DWI). Several reports have suggested that the apparent diffusion coefficient (ADC) values obtained with FSE-based DWI are different from those obtained with EPI-DWI. The purpose of this study was to create phantoms that mimic the T-2 and ADC values of various tissues and to demonstrate the ADC values obtained with RADAR-DWI and EPI-DWI in low-field magnetic resonance imaging (MRI) systems. Methods: Several phantoms were created using sucrose and manganese (II) chloride tetrahydrate mimicking various tissues. RADAR-DWI and EPI-DWI were used to scan the phantoms, and the obtained ADC values were compared. Results: The ADC values obtained with RADAR-DWI were significantly higher than those obtained with EPI-DWI for all phantoms (P < 0.05). The ADC values obtained by RADAR-DWI ranged from 0.70 +/- 0.01 to 1.21 +/- 0.02 ( x 10(-3)mm(2)s(-1)). Meanwhile, the ADC values obtained with EPI-DWI ranged from 0.59 +/- 0.01 to 1.08 +/- 0.05 ( x 10(-3)mm(2)s(-1)). Conclusions: We created phantoms mimicking T-2 and ADC values of various tissues and demonstrated the differences in ADC values obtained with RADAR-DWI and EPI-DWI using low-field MRI systems. Implications for practice: ADC values obtained by RADAR-DWI are significantly higher than those obtained by EPI-DWI, with different cutoff values for various tumor malignancies between them. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those for text and data mining, AI training, and similar technologies.
引用
收藏
页码:1290 / 1296
页数:7
相关论文
共 43 条
[1]  
Atlas SW, 2000, AM J NEURORADIOL, V21, P1190
[2]   A REVIEW OF H-1 NUCLEAR-MAGNETIC-RESONANCE RELAXATION IN PATHOLOGY - ARE T1 AND T2 DIAGNOSTIC [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
ARGERSINGER, RE ;
ALLENMOORE, G .
MEDICAL PHYSICS, 1987, 14 (01) :1-37
[3]   Characterization of focal liver lesions by ADC measurements using a respiratory triggered diffusion-weighted single-shot echo-planar MR imaging technique [J].
Bruegel, Melanie ;
Holzapfel, Konstantin ;
Gaa, Jochen ;
Woertler, Klaus ;
Waldt, Simone ;
Kiefer, Berthold ;
Stemmer, Alto ;
Ganter, Carl ;
Rummeny, Ernst J. .
EUROPEAN RADIOLOGY, 2008, 18 (03) :477-485
[4]   Variabilities in apparent diffusion coefficient (ADC) measurements of the spleen and the paraspinal muscle: A single center large cohort study [J].
Chen, Yukun ;
Yang, Panpan ;
Fu, Caixia ;
Bian, Yun ;
Shao, Chengwei ;
Ma, Chao ;
Lu, Jianping .
HELIYON, 2023, 9 (07)
[5]   Differentiation between pelvic abscesses and pelvic tumors with diffusion-weighted MR imaging: a preliminary study [J].
Chou, Chen-Pin ;
Chiou, Shih-Hwa ;
Levenson, Robin B. ;
Huang, Jer-Shyung ;
Yang, Tsung-Lung ;
Yu, Chia-Cheng ;
Chiang, An-Jen ;
Pan, Huay-Ben .
CLINICAL IMAGING, 2012, 36 (05) :532-538
[6]   Multishot diffusion-weighted PROPELLER magnetic resonance imaging of the abdomen [J].
Deng, Jie ;
Miller, Frank H. ;
Salem, Riad ;
Omary, Reed A. ;
Larson, Andrew C. .
INVESTIGATIVE RADIOLOGY, 2006, 41 (10) :769-775
[7]   NMR-NEUROPATHOLOGIC CORRELATION IN STROKE [J].
DEWITT, LD ;
KISTLER, JP ;
MILLER, DC ;
RICHARDSON, EP ;
BUONANNO, FS .
STROKE, 1987, 18 (02) :342-351
[8]   BLADE in Sagittal T2-Weighted MR Imaging of the Cervical Spine [J].
Fellner, C. ;
Menzel, C. ;
Fellner, F. A. ;
Ginthoer, C. ;
Zorger, N. ;
Schreyer, A. ;
Jung, E. M. ;
Feuerbach, S. ;
Finkenzeller, T. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (04) :674-681
[9]   Single shot echo planar imaging (ssEPI) vs single shot turbo spin echo (ssTSE) DWI of the orbit in patients with ocular melanoma [J].
Gumeler, Ekim ;
Parlak, Safak ;
Yazici, Gozde ;
Karabulut, Erdem ;
Kiratli, Hayyam ;
Oguz, Kader K. .
BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1118)
[10]   A new phantom and empirical formula for apparent diffusion coefficient measurement by a 3 Tesla magnetic resonance imaging scanner [J].
Hara, Marina ;
Kuroda, Masahiro ;
Ohmura, Yuichi ;
Matsuzaki, Hidenobu ;
Kobayashi, Tomoki ;
Murakami, Jun ;
Katashima, Kazunori ;
Ashida, Masakazu ;
Ohno, Seiichiro ;
Asaumi, Jun-Ichi .
ONCOLOGY LETTERS, 2014, 8 (02) :819-824