Simultaneous measurements of morphology, size distribution and concentration of ice crystals in icing wind tunnel with HACPI

被引:0
作者
Wang, Boyi [1 ]
Zhang, Letian [1 ]
Huang, Xinyuan [1 ]
Guo, Xiangdong [1 ,2 ]
Wu, Yingchun [1 ]
Wu, Xuecheng [1 ]
Song, Xiaoming [3 ]
Bai, Pengbo [3 ]
Shi, Pan [3 ]
Zhu, Shimin [4 ]
Liu, Kai [4 ]
Zhang, Qian
机构
[1] Zhejiang Univ, State Key Lab Clean Energy Utilizat, Hangzhou 310027, Peoples R China
[2] China Aerodynam Res & Dev Ctr, Key Lab Icing & Anti Deicing, Mianyang 621000, Peoples R China
[3] AVIC CAPDI Integrat Equipment Co Ltd, Beijing 102206, Peoples R China
[4] AVIC Chengdu CAIC Elect Co Ltd, Chengdu 610091, Sichuan, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
Icing wind tunnel; Ice crystal generation; Morphology; Size distribution; Concentration; HACPI; PARTICLE; MICROSCOPY;
D O I
10.1016/j.expthermflusci.2024.111266
中图分类号
O414.1 [热力学];
学科分类号
摘要
The simulation of ice crystal icing at high altitudes in the icing wind tunnel holds significant value for the study of icing on the aero-engine, pitot probe and other components of aircraft. Concurrently, the measurement of ice crystals in the icing wind tunnel is an aspect that warrants attention. This study introduces the Holographic Airborne for Cloud Particle Imager (HACPI) based on in-line holography, and conducts a series of experiments in the icing wind tunnel at AVIC Chengdu CAIC Electronics Co., LTD. The icing wind tunnel is equipped with an ice crystal generation system (ICGS) that operates on a grinding-based method. The morphology, size distribution and concentration of the ice crystals are simultaneously measured under airspeeds of Mach 0.55 and Mach 0.62 with HACPI, the results indicate that the morphology of ice crystals generated by the ICGS exhibits randomness. It is observed that with an increase in particle size, the irregularity also increases, and the maximum size of the ice crystals produced can approach 1000 mu m. Under two typical test conditions, the MVD are 202.4 . 4 mu m and 247.7 . 7 mu m, respectively, while the MMD are 103.7 . 7 mu m and 166.9 . 9 mu m. In addition, the IWC are 4.47 . 47 g/m3 3 and 6.67 . 67 g/m3. 3 . A close relationship is observed between the particle size distribution and concentration of the ice crystals and the operating frequency of the ICGS components. Furthermore, this work demonstrates the significant potential of digital holography in measuring ice crystals in icing wind tunnels.
引用
收藏
页数:12
相关论文
共 46 条
  • [1] Al-Khalil Kamel, 2003, 41 AER SCI M EXH, P903
  • [2] Ice crystal habits from cloud chamber studies obtained by in-line holographic microscopy related to depolarization measurements
    Amsler, Peter
    Stetzer, Olaf
    Schnaiter, Martin
    Hesse, Evelyn
    Benz, Stefan
    Moehler, Ottmar
    Lohmann, Ulrike
    [J]. APPLIED OPTICS, 2009, 48 (30) : 5811 - 5822
  • [3] [Anonymous], 2011, EASA CS-E 780
  • [4] Improvement in determination of ice water content from two-dimensional particle imagery. Part I: Image-to-mass relationships
    Baker, Brad
    Lawson, R. Paul
    [J]. JOURNAL OF APPLIED METEOROLOGY AND CLIMATOLOGY, 2006, 45 (09) : 1282 - 1290
  • [5] Design, construction and commissioning of the Braunschweig Icing Wind Tunnel
    Bansmer, Stephan E.
    Baumert, Arne
    Sattler, Stephan
    Knop, Inken
    Leroy, Delphine
    Schwarzenboeck, Alfons
    Jurkat-Witschas, Tina
    Voigt, Christiane
    Pervier, Hugo
    Esposito, Biagio
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2018, 11 (06) : 3221 - 3249
  • [6] Bartkus T.P., 2017, AIAA Paper 2017-4243, DOI 10.2514/6.2017-4243
  • [7] Baumert A., 2016, 8 AIAA ATMOSPHERIC, P4053
  • [8] Baumert Arne, 2015, 53 AIAA AER SCI M, P1225
  • [9] HoloGondel: in situ cloud observations on a cable car in the Swiss Alps using a holographic imager
    Beck, Alexander
    Henneberger, Jan
    Schoepfer, Sarah
    Fugal, Jacob
    Lohmann, Ulrike
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2017, 10 (02) : 459 - 476
  • [10] Clark Catherine, 2023, SAE Technical Paper