Scatter estimation and correction using time-of-flight and deconvolution in x-ray medical imaging

被引:1
作者
Rossignol, Julien [1 ,2 ]
Belanger, Gabriel [1 ,2 ]
Fromont, Maelle [1 ,2 ]
Therrien, Audrey C. [1 ,2 ]
Fontaine, Rejean [1 ,2 ]
Berube-Lauziere, Yves [2 ]
机构
[1] Univ Sherbrooke, Inst Interdisciplinaire Innovat Technol 3IT, Sherbrooke, PQ, Canada
[2] Univ Sherbrooke, Dept Genie Elect & Genie Informat, Sherbrooke, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
time-of-flight; computed tomography; radiography; deconvolution; scatter correction; scatter rejection; medical imaging; BEAM COMPUTED-TOMOGRAPHY; DIGITAL RADIOGRAPHY; GENERAL FRAMEWORK; ANTISCATTER GRIDS; QUALITY; PERFORMANCE; REDUCTION; SYSTEM;
D O I
10.1088/1361-6560/ad700e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Time-of-flight (TOF) scatter rejection requires a total timing jitter, including the detector timing jitter and the x-ray source's pulses width, of 50 ps or less to mitigate most of the effects of scattered photons in radiography and CT imaging. However, since the total contribution of the source and detector to the timing jitter can be retrieved during an acquisition with nothing between the source and detector, it can be demonstrated that this contribution may be partially removed to improve the image quality. Approach. A scatter correction method using iterative deconvolution of the measured time point-spread function estimates the number of scattered photons detected in each pixel. To evaluate the quality of the estimation, GATE was used to simulate the radiography of a water cylinder with bone inserts, and a head and torso in a system with total timing jitters from 100 ps up to 500 ps full-width-at-half-maximum (FWHM). Main results. With a total timing jitter of 200 ps FWHM, 89% of the contrast degradation caused by scattered photons was recovered in a head and torso radiography, compared to 28% with a simple time threshold method. Corrected images using the estimation have a percent root-mean square error between 2% and 14% in both phantoms with timing jitters from 100 to 500 ps FWHM which is lower than the error achieved with scatter rejection alone at 100 ps FWHM. Significance. TOF x-ray imaging has the potential to mitigate the effects of the scattering contribution and offers an alternative to anti-scatter grids that avoids loss of primary photons. Compare to simple TOF scatter rejection using only a threshold, the deconvolution estimation approach has lower requirements on both the source and detector. These requirements are now within reach of state-of-the-art systems.
引用
收藏
页数:14
相关论文
共 45 条
  • [1] Bodi Geoffroy, 2009, Proceedings of the SPIE - The International Society for Optical Engineering, V7369, DOI 10.1117/12.831687
  • [2] Stationary Computed Tomography for Space and other Resource-constrained Environments
    Cramer, Avilash
    Hecla, Jake
    Wu, Dufan
    Lai, Xiaochun
    Boers, Tim
    Yang, Kai
    Moulton, Tim
    Kenyon, Steven
    Arzoumanian, Zaven
    Krull, Wolfgang
    Gendreau, Keith
    Gupta, Rajiv
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [3] X-RAY TRANSMISSION FORMULA FOR ANTISCATTER GRIDS
    DAY, GJ
    DANCE, DR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12) : 1429 - 1433
  • [4] DESIGN OF A PULSED X-RAY SYSTEM FOR FLUORESCENT LIFETIME MEASUREMENTS WITH A TIMING ACCURACY OF 109 PS
    DERENZO, SE
    MOSES, WW
    BLANKESPOOR, SC
    ITO, M
    OBA, K
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (03) : 629 - 631
  • [5] Deconvolution method for recovering the photon time-of-flight distribution from time-resolved measurements
    Diop, Mamadou
    St Lawrence, Keith
    [J]. OPTICS LETTERS, 2012, 37 (12) : 2358 - 2360
  • [6] Comparative Study of Image Quality in Time-Correlated Single-Photon Counting Computed Tomography
    Gaudreault, David
    Rossignol, Julien
    Berube-Lauziere, Yves
    Fontaine, Rejean
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (03) : 343 - 349
  • [7] Experimental time resolution limits of modern SiPMs and TOF-PET detectors exploring different scintillators and Cherenkov emission
    Gundacker, Stefan
    Turtos, Rosana Martinez
    Kratochwil, Nicolaus
    Pots, Rosalinde Hendrika
    Paganoni, Marco
    Lecoq, Paul
    Auffray, Etiennette
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (02)
  • [8] ScatterNet: A convolutional neural network for cone-beam CT intensity correction
    Hansen, David C.
    Landry, Guillaume
    Kamp, Florian
    Li, Minglun
    Belka, Claus
    Parodi, Katia
    Kurz, Christopher
    [J]. MEDICAL PHYSICS, 2018, 45 (11) : 4916 - 4926
  • [9] GATE:: a simulation toolkit for PET and SPECT
    Jan, S
    Santin, G
    Strul, D
    Staelens, S
    Assié, K
    Autret, D
    Avner, S
    Barbier, R
    Bardiès, M
    Bloomfield, PM
    Brasse, D
    Breton, V
    Bruyndonckx, P
    Buvat, I
    Chatziioannou, AF
    Choi, Y
    Chung, YH
    Comtat, C
    Donnarieix, D
    Ferrer, L
    Glick, SJ
    Groiselle, CJ
    Guez, D
    Honore, PF
    Kerhoas-Cavata, S
    Kirov, AS
    Kohli, V
    Koole, M
    Krieguer, M
    van der Laan, DJ
    Lamare, F
    Largeron, G
    Lartizien, C
    Lazaro, D
    Maas, MC
    Maigne, L
    Mayet, F
    Melot, F
    Merheb, C
    Pennacchio, E
    Perez, J
    Pietrzyk, U
    Rannou, FR
    Rey, M
    Schaart, DR
    Schmidtlein, CR
    Simon, L
    Song, TY
    Vieira, JM
    Visvikis, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) : 4543 - 4561
  • [10] Scatter correction of cone-beam CT using a deep residual convolution neural network (DRCNN)
    Jiang, Yangkang
    Yang, Chunlin
    Yang, Pengfei
    Hu, Xi
    Luo, Chen
    Xue, Yi
    Xu, Lei
    Hu, Xiuhua
    Zhang, Luhan
    Wang, Jing
    Sheng, Ke
    Niu, Tianye
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (14)