Employing a switching valve to automate external standard calibration in inductively coupled plasma optical emission spectrometry

被引:1
作者
Ingham, Jesse R. [1 ]
Fontoura, Beatriz M. [1 ]
Jones, Bradley T. [1 ]
Donati, George L. [1 ,2 ]
机构
[1] Wake Forest Univ, Dept Chem, Salem Hall,Box 7486, Winston Salem, NC 27109 USA
[2] Wadsworth Ctr, New York State Dept Hlth, Albany, NY 12208 USA
关键词
Gradient dilution; Automated calibration; Single standard calibration; Green chemistry; ICP-OES; ICP-OES; FLOW-INJECTION; MULTINEBULIZER; BIODIESEL; SAMPLES; METALS;
D O I
10.1016/j.talanta.2024.126749
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The traditional external standard calibration method (EC) is automated and simplified using a four-port switching valve (SV4) and a multi-signal approach that enables the generation of several calibration points from a single calibration solution. The SV4-EC method is applied to inductively coupled plasma optical emission spectrometry (ICP-OES) and is based on gradient dilution taking place within the instrument's sample introduction tubing. Both the calibration solution and the samples are diluted by a blank solution containing an internal standard species. Forty-five dilution points are collected over time while the solutions are mixed. Instrument responses from the calibration solution are then plotted against those from the samples, and the slope of the calibration curve is used to determine the unknown analyte concentrations in the samples. The method is used to determine Ba, Co, Cr, Cu, Fe, Mn, Ni, V and Zn in coconut water, creek water, green tea, mouthwash, soft drink, vinegar, and vodka. Limits of detection are in the 0.0002-0.009 mg L- 1 (n = 10) range, with precision on the order of 0.4 %-3 % RSD. Analyte percent recoveries from a 0.5 mg L- 1 spike are in the ranges of 88.4 %-111 %, 88.9 %-111 %, and 88.0 %-111 % for EC, SV4-EC, and the internal-standard-corrected method (SV4-EC/Sc), respectively. No statistically significant difference is observed between EC and SV4-EC recoveries for any of the sample matrices evaluated. Comparable results between EC and SV4-EC were also found for the analysis of two certified reference materials, Bovine Liver and Oyster Tissue. Based on a single calibration solution, the SV4-EC method requires caution when preparing the calibration standard to minimize measurement bias.
引用
收藏
页数:9
相关论文
共 38 条
[1]   Analysis of biodiesel and oil samples by on-line calibration using a Flow Blurring® multinebulizer in ICP OES without oxygen addition [J].
Angel Aguirre, Miguel ;
Kovachev, Nikolay ;
Hidalgo, Montserrat ;
Canals, Antonio .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (12) :2102-2110
[2]   Compensation for matrix effects on ICP-OES by on-line calibration methods using a new multi-nebulizer based on Flow Blurring® technology [J].
Angel Aguirre, Miguel ;
Kovachev, Nikolay ;
Almagro, Beatriz ;
Hidalgo, Montserrat ;
Canals, Antonio .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2010, 25 (11) :1724-1732
[3]  
[Anonymous], 2019, Reduce costs and boost productivity with the advanced valve system (AVS) 6 or 7 port switching valve system
[4]  
[Anonymous], Cheminert injectors and selectors low pressure injectors and switching valves
[5]  
[Anonymous], 2018, VICI cheminert injectors, valves, fittings, tubing, syringes, detectors catalog
[6]  
[Anonymous], 2016, Reduce costs and boost productivity with the advanced valve system (AVS 4) four port switching valve
[7]  
[Anonymous], Design principles for sustainable green chemistry and engineering
[8]   Analysis of metals and phosphorus in biodiesel B100 from different feedstock using a Flow Blurring® multinebulizer in inductively coupled plasma-optical emission spectrometry [J].
Avila Orozco, Francisco D. ;
Kovachev, Nikolay ;
Aguirre Pastor, Miguel Angel ;
Domini, Claudia E. ;
Fernandez Band, Beatriz S. ;
Canals Hernandez, Antonio .
ANALYTICA CHIMICA ACTA, 2014, 827 :15-21
[9]   Plasma-related matrix effects in inductively coupled plasma - atomic emission spectrometry by group I and group II matrix-elements [J].
Chan, GCY ;
Chan, WT .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2003, 58 (07) :1301-1317
[10]   Investigation of plasma-related matrix effects in inductively coupled plasma-atomic emission spectrometry caused by matrices with low second ionization potentials - identification of the secondary factor [J].
Chan, George C. -Y. ;
Hieftje, Gary M. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (06) :642-659