Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging

被引:1
|
作者
Li, Da [1 ]
Ren, Yong [2 ]
Liu, Feng [3 ]
机构
[1] Univ Sci & Technol China, Hefei 230031, Peoples R China
[2] Chinese Acad Sci, Hefei Inst Phys Sci, Inst Plasma Phys, Hefei 200031, Peoples R China
[3] Univ Queensland, Sch Elect Engn & Comp Sci, Brisbane, Qld 4072, Australia
关键词
Coils; Magnetic resonance imaging; Superconducting magnets; Magnetic fields; Windings; Stress; Electromagnetic forces; Head scanner; magnetic resonance imaging (MRI); REBCO coil; superconducting magnet;
D O I
10.1109/TASC.2024.3426543
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A 0.5 T REBCO MRI coil with a warm bore of 560 mm was designed for head imaging using linear programming (LP) and nonlinear programming (NLP). The LP method was used to determine the initial distribution of the coil cluster, followed by the NLP method, which optimized the radii and heights of the coil blocks using this initial layout as a reference. Both layer-winding and double-pancake (DP) winding coils were explored for developing MRI magnets. The layer-winding coil enabled the MRI magnet to achieve a uniformity of 10 parts per million (ppm) over a 200-mm diameter spherical volume. In contrast, the DP winding coils attained a field uniformity of 17.6 ppm without employing either iron or superconducting shim coils. The stray magnetic field of the MRI magnet was designed to be within 0.5 mT at radial and axial positions of 3.2 and 4.2 m, respectively. Because the REBCO magnet has a substantial temperature margin, it can be effectively cooled using solid nitrogen at 35 K or through conduction cooling. This article presents the electromagnetic design and mechanical analysis results of the coils.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 50 条
  • [21] 0.5T MRI as a competitor to CT for sinus imaging
    Parker, Mark
    Beyea, Steven
    Rioux, James
    King, Brian
    Abdolell, Mohamed
    Reeve, Sarah
    Lieuwen, Beverly
    Bowen, Chris
    Volders, David
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [22] Structural Design of a 9.4 T Whole-Body MRI Superconducting Magnet
    Dai, Yinming
    Wang, Qiuliang
    Wang, Chunzhong
    Li, Lankai
    Wang, Housheng
    Ni, Zhipeng
    Song, Shousen
    Chen, Shunzhong
    Zhao, Baozhi
    Wang, Hui
    Li, Yi
    Hu, Xinning
    Cui, Chunyan
    Cheng, Junsheng
    Lei, Yuanzhong
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)
  • [23] Quench Protection Design of a 1.5 T Superconducting MRI Magnet
    Li, Yi
    Wang, Qiuliang
    Chen, Shunzhong
    Liu, Feng
    Hu, Xinning
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)
  • [24] Infratentorial brain maturation: A comparison of MRI at 0.5 and 1.5T
    Hittmair, K
    Kramer, J
    Rand, T
    Bernert, G
    Wimberger, D
    NEURORADIOLOGY, 1996, 38 (04) : 360 - 366
  • [25] Design and Construction of a Low-Cryogen, Lightweight, Head-Only 7T MRI Magnet
    Wu, Anbo
    Ricci, Justin
    Conte, Gene
    Van Epps, Chris
    Xu, Minfeng
    Bai, Ye
    Parizh, Michael
    Stautner, Wolfgang
    Hua, Yihe
    Vermilyea, Mark
    Soni, Vijay
    Foo, Thomas K. F.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 5
  • [26] Conceptual Design of a Portable, Solid-Nitrogen-Cooled 0.5-T/560-mm Point-of-Care MRI Magnet
    Park, Dongkeun
    Bascunan, Juan
    Lee, Wooseung
    Iwasa, Yukikazu
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2023, 33 (05)
  • [27] The Iseult/Inumac whole body 11.7 T MRI magnet design
    Schild, Th.
    Aubert, G.
    Berriaud, C.
    Bredy, Ph.
    Juster, F. P.
    Meuris, C.
    Nunio, F.
    Quettier, L.
    Rey, J. M.
    Vedrine, P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2008, 18 (02) : 904 - 907
  • [28] Design and Improvement of Quench Protection for A 3 T MRI Superconducting Magnet
    Shang, Wangnan
    Mei, Yunhao
    Yang, Shige
    Tang, Bohan
    Jiang, Shili
    Yu, Hui
    Xie, Bowen
    Kuang, Guangli
    Jiang, Donghui
    JOURNAL OF LOW TEMPERATURE PHYSICS, 2025, 218 (3-4) : 146 - 165
  • [29] An Optimal Design of Actively Shielded MRI Superconducting Magnet
    Liang, Zhenguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (02)
  • [30] Design of a Halbach Magnet Gradient Coil for High-Resolution Portable MRI
    Lu, Rongsheng
    Song, Xuan
    Bao, Chong
    Jiang, Xiaowen
    Ni, Zhonghua
    Yi, Hong
    IEEE SENSORS JOURNAL, 2025, 25 (05) : 8187 - 8195