Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging

被引:1
|
作者
Li, Da [1 ]
Ren, Yong [2 ]
Liu, Feng [3 ]
机构
[1] Univ Sci & Technol China, Hefei 230031, Peoples R China
[2] Chinese Acad Sci, Hefei Inst Phys Sci, Inst Plasma Phys, Hefei 200031, Peoples R China
[3] Univ Queensland, Sch Elect Engn & Comp Sci, Brisbane, Qld 4072, Australia
关键词
Coils; Magnetic resonance imaging; Superconducting magnets; Magnetic fields; Windings; Stress; Electromagnetic forces; Head scanner; magnetic resonance imaging (MRI); REBCO coil; superconducting magnet;
D O I
10.1109/TASC.2024.3426543
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A 0.5 T REBCO MRI coil with a warm bore of 560 mm was designed for head imaging using linear programming (LP) and nonlinear programming (NLP). The LP method was used to determine the initial distribution of the coil cluster, followed by the NLP method, which optimized the radii and heights of the coil blocks using this initial layout as a reference. Both layer-winding and double-pancake (DP) winding coils were explored for developing MRI magnets. The layer-winding coil enabled the MRI magnet to achieve a uniformity of 10 parts per million (ppm) over a 200-mm diameter spherical volume. In contrast, the DP winding coils attained a field uniformity of 17.6 ppm without employing either iron or superconducting shim coils. The stray magnetic field of the MRI magnet was designed to be within 0.5 mT at radial and axial positions of 3.2 and 4.2 m, respectively. Because the REBCO magnet has a substantial temperature margin, it can be effectively cooled using solid nitrogen at 35 K or through conduction cooling. This article presents the electromagnetic design and mechanical analysis results of the coils.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 50 条
  • [1] Electromagnetic Design of 1.0-T Ultrashort-Cavity Intraoperative Head MRI Superconducting Magnet
    Zhang, Qing
    Xie, Yuchong
    Wang, Pingping
    Liu, Zhengrong
    Du, Huiyu
    Huang, Xiaoyan
    Mo, Lei
    Chen, Lin
    Zhou, Yufu
    Qiu, Bensheng
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2025, 35 (02)
  • [2] An Experimental Study on the Passive Shimming in the 0.5 T Superconducting MRI Magnet
    Qu, Hongyi
    Wang, Hui
    Chen, Qi
    Wang, Cong
    Liu, Xin
    Wang, Qiuliang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2022, 32 (07)
  • [3] Electromagnetic Structure Design of the Whole-Body 14 T MRI Magnet
    Xu, Aihua
    Wu, Kaihong
    Cai, Fuhai
    Zhang, Bo
    Tang, Zhipeng
    Shi, Yi
    Sang, Xiaohu
    Luo, Chunhui
    Shi, Kunju
    Shi, Dongli
    Ye, Sui
    Zhou, Jinyang
    Wu, Yu
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (03) : 1 - 5
  • [4] Design of a 14 T Actively Shielded Superconducting Magnet for Animal MRI
    Huang, Weican
    Jiang, Xiaohua
    Chai, Guolin
    Huang, Xianrui
    Zou, Qingyu
    Li, Ye
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2025, 35 (05)
  • [5] Design and Cooling Properties of High Stable Field REBCO Superconducting Magnet for MRI
    Yokoyama, Shoichi
    Kiss, Takanobu
    Miyagi, Daisuke
    Tsuda, Makoto
    Nakamura, Taketsune
    Shirai, Yasuyuki
    Miura, Hideaki
    Matsuda, Tetsuya
    Inoue, Tatsuya
    Morita, Yusuke
    Eguchi, Ryo
    Otake, Shunsuke
    Tanabe, Hajime
    Sato, Shinji
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2020, 30 (04)
  • [6] Design, Fabrication, and Test of a 16 T 100 mm all-REBCO Superconducting Magnet
    Hu, Shuai
    Liu, Huajun
    Liu, Fang
    Wang, Zhaoran
    Hong, Wenzhe
    Gao, Shuowei
    Guo, Liang
    Ma, Hongjun
    Xue, Shenquan
    Zhang, Shuqing
    Yang, Dongshen
    Shao, Liangjun
    Qu, Timing
    Zhu, Jiamin
    Zhang, Xintao
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2025, 35 (05)
  • [7] Mechanical Stability Influenced by Assembly Errors of a 0.5 T Mobile MRI Superconducting Magnet
    Yao, Lianhao
    Qu, Hongyi
    Wang, Hui
    Wu, Huafang
    Liu, Jianhua
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (08) : 1 - 5
  • [8] Actively-Shielded Superconducting Magnet Design of a Large-Bore 7 T Animal MRI Scanner
    Wang, Yaohui
    Wang, Qiuliang
    Wang, Hui
    Wang, Lei
    Zhai, Yujia
    Qu, Hongyi
    Liu, Yang
    Liu, Feng
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2020, 30 (04)
  • [9] Electromagnetic design of actively shielded MRI magnet for dedicated mice imaging
    Li, Da
    Ren, Yong
    Wang, Hongzhi
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2020, 575
  • [10] Electromagnetic design of a 1.5T open MRI superconducting magnet
    Wang, Yaohui
    Wang, Qiuliang
    Wang, Lei
    Qu, Hongyi
    Liu, Yang
    Liu, Feng
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2020, 570