Evaluation of bone integrity around the acetabular cup using noninvasive laser resonance frequency analysis

被引:1
|
作者
Hatakeyama, Takuto [1 ]
Nakashima, Daisuke [1 ,2 ]
Mikami, Katsuhiro [3 ]
Oya, Akihito [1 ]
Fujie, Atsuhiro [1 ]
Sujino, Asahi [1 ]
Nakamura, Masaya [1 ]
Nagura, Takeo [1 ,2 ]
机构
[1] Keio Univ, Sch Med, Dept Orthopaed Surg, 35 Shinanomachi,Shinjuku Ku, Tokyo 1608582, Japan
[2] Keio Univ, Sch Med, Dept Clin Biomech, Shinjuku Ku, Tokyo, Japan
[3] Kindai Univ, Fac Biol Oriented Sci & Technol, Kinokawa, Wakayama, Japan
关键词
bone integrity; bone mineral density; laser resonance frequency analysis; noninvasive; resonance frequency; TOTAL HIP-ARTHROPLASTY; IMPLANT STABILITY; INITIAL STABILITY; MINERAL DENSITY; 2-YEAR RSA; DESIGN;
D O I
10.1002/jor.25925
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Resonance frequency analysis (RFA) is valuable for assessing implant status. In a previous investigation, acetabular cup fixation was assessed using laser RFA and the pull-down force was predicted in an in vitro setting. While the pull-down force alone is sufficient for initial fixation evaluation, it is desirable to evaluate the bone strength of the foundation for subsequent fixation. Diminished bone quality causes micromotion, migration, and protracted osseointegration, consequently elevating susceptibility to periprosthetic fractures and failure of ingrained trabecular bone. Limited research exists on the evaluation of bone mineral density (BMD) around the cup using RFA. For in vivo application of laser RFA, we implemented the sweep pulse excitation method and engineered an innovative laser RFA device having low laser energy and small dimensions. We focused on a specific frequency range (2500-4500 Hz), where the peak frequency was presumed to be influenced by foundational density. Quantitative computed tomography with a phantom was employed to assess periprosthetic BMD. Correlation between the resonance frequency within the designated range and the density around the cup was evaluated both in the laboratory and in vivo using the novel laser RFA device. The Kruskal-Wallis test showed robust correlations in both experiments (laboratory study: R = 0.728, p < 0.001; in vivo study: R = 0.619, p < 0.001). Our laser RFA system can assess the quality of bone surrounding the cup. Laser RFA holds promise in predicting the risk of loosening and might aid in the decision-making process for additional fixation through screw insertion.
引用
收藏
页码:2552 / 2561
页数:10
相关论文
共 50 条
  • [31] Noninvasive Evaluation of Kidney Hypoxia and Fibrosis Using Magnetic Resonance Imaging
    Inoue, Tsutomu
    Kozawa, Eito
    Okada, Hirokazu
    Inukai, Kouichi
    Watanabe, Shinichi
    Kikuta, Tomohiro
    Watanabe, Yusuke
    Takenaka, Tsuneo
    Katayama, Shigehiro
    Tanaka, Junji
    Suzuki, Hiromichi
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2011, 22 (08): : 1429 - 1434
  • [32] The reconstruction of extended acetabular bone defects in revision hip arthroplasty - Risk factors for migration and loosening using the pedestal cup
    Tohtz, S.
    Katterle, H.
    Matziolis, G.
    Drahn, T.
    Perka, C.
    ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2007, 145 (02): : 176 - 180
  • [33] Evaluation of the role of pudendal nerve integrity in female sexual function using noninvasive techniques
    Connell, K
    Guess, MK
    La Combe, J
    Wang, A
    Powers, K
    Lazarou, G
    Mikhail, M
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2005, 192 (05) : 1712 - 1717
  • [34] Ex Vivo Evaluation of Cementless Acetabular Cup Stability Using Impact Analyses with a Hammer Instrumented with Strain Sensors
    Tijou, Antoine
    Rosi, Giuseppe
    Hernigou, Philippe
    Flouzat-Lachaniette, Charles-Henri
    Haiat, Guillaume
    SENSORS, 2018, 18 (01):
  • [35] A parametric study of acetabular cup design variables using finite element analysis and statistical design of experiments
    Mantell, SC
    Chanda, H
    Bechtold, JE
    Kyle, RF
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1998, 120 (05): : 667 - 675
  • [36] Accuracy of acetabular cup placement using an angle-adjusting alignment guide with laser pointer in total hip arthroplasty
    Okanoue, Yusuke
    Aso, Koji
    Dan, Junpei
    Takaya, Shogo
    Izumi, Masashi
    Kawakami, Teruhiko
    Ikeuchi, Masahiko
    JOURNAL OF ORTHOPAEDIC SURGERY, 2020, 28 (03)
  • [37] A parametric study of acetabular cup design variables using finite element analysis and statistical design of experiments
    Mantell, S.C.
    Chanda, H.
    Bechtold, J.E.
    Kyle, R.F.
    Journal of Biomechanical Engineering, 1998, 120 (05): : 667 - 675
  • [38] A noncontact resonance frequency detection technique for the assessment of the interfacial bone defect around a dental implant
    Pan, Min-Chun
    Zhuang, Han-Bo
    Chen, Chin-Sung
    Wu, Jer-Whey
    Lee, Shyh-Yuan
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (12) : 1825 - 1830
  • [39] Host bone coverage requirements for press-fit acetabular cup fixation. 3D-CT analysis
    Takao, M
    Sugano, N
    Nishii, T
    Miki, H
    Koyama, T
    Yoshikawa, H
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1299 - 1299
  • [40] Finite element analysis of shear stresses at the implant-bone interface of an acetabular press-fit cup during impingement
    Voigt, Christian
    Kloehn, Carsten
    Bader, Rainer
    von Salis-Soglio, Georg
    Scholz, Roger
    BIOMEDIZINISCHE TECHNIK, 2007, 52 (02): : 208 - 215