Dosimetric Effect of Target Position Accuracy on Single-Isocenter Multiple Liver Metastases SBRT

被引:1
作者
Zhang, Yun [1 ,2 ,3 ]
Niu, Shanzhou [4 ]
Yuan, Jun [5 ]
Wang, Xiaoping [1 ,2 ,3 ]
Gong, Changfei [1 ,2 ,3 ]
Tang, Chunbo [5 ]
机构
[1] Jiangxi Canc Hosp, Dept Med Oncol, Nanchang 330029, Jiangxi, Peoples R China
[2] Nanchang Univ, Affiliated Hosp 2, Nanchang, Peoples R China
[3] Jiangxi Canc Hosp, Jiangxi Clin Res Ctr Canc, JXHC Key Lab Tumor Microenvironm & Immunoregulat, Nanchang, Jiangxi, Peoples R China
[4] Gannan Normal Univ, Ganzhou Key Lab Computat Imaging, Ganzhou, Jiangxi, Peoples R China
[5] Gannan Med Univ, Affiliated Hosp 1, Dept Oncol, Ganzhou 341000, Peoples R China
关键词
stereotactic body radiotherapy; multiple liver metastases; baseline shifts; rotational errors; 4DCT; BASE-LINE SHIFTS; RESPIRATORY MOTION; RADIATION-THERAPY; BRAIN METASTASES; RADIOTHERAPY; RADIOSURGERY; VARIABILITY; DELIVERY; IMPACT; MODEL;
D O I
10.1177/15330338241257422
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the dosimetric effects of intrafraction baseline shifts combined with rotational errors on Four-dimensional computed tomography-guided stereotactic body radiotherapy for multiple liver metastases (MLMs). Methods: A total of 10 patients with MLM (2 or 3 lesions) were selected for this retrospective study. Baseline shift errors of 0.5, 1.0, and 2.0 mm; and rotational errors of 0.5 degrees, 1 degrees, and 1.5 degrees, were simulated about all axes. All of the baseline shifts and rotation errors were simulated around the planned isocenter using a matrix transformation of 6 degrees of freedom. The coverage degradation of baseline shifts and rotational errors were analyzed according to the dose to 95% of the planning target volume (D95) and the volume covered by 95% of the prescribed dose (V95), and related changes in gross tumor volume were also analyzed. Results: At the rotation error of 0.5 degrees and the baseline offset of less than 0.5 mm, the D95 and V95 values of all targets were >95%. For rotational errors of 1.0 degrees (combined with all baseline shift errors), 36.3% of targets had D95 and V95 values of <95%. Coverage worsened substantially when the baseline shift errors were increased to 1.0 mm. D95 and V95 values were >95% for about 77.3% of the targets. Only 11.4% of the D95 and V95 values were >95% when the baseline shift errors were increased to 2.0 mm. When the rotational error was increased to 1.5 degrees and baseline shift errors increased to 1.0 mm, the D95 and V95 values were >95% in only 3 cases. Conclusions: The multivariate regression model analysis in this study showed that the coverage of the target decreased further with reduced target volume, increasing the baseline drift, the rotation error, and the distance to the target.
引用
收藏
页数:11
相关论文
共 39 条
[1]   External Beam Radiation Therapy for Primary Liver Cancers: An ASTRO Clinical Practice Guideline [J].
Apisarnthanarax, Smith ;
Barry, Aisling ;
Cao, Minsong ;
Czito, Brian ;
DeMatteo, Ronald ;
Drinane, Mary ;
Hallemeier, Christopher L. ;
Koay, Eugene J. ;
Lasley, Foster ;
Meyer, Jeffrey ;
Owen, Dawn ;
Pursley, Jennifer ;
Schaub, Stephanie K. ;
Smith, Grace ;
Venepalli, Neeta K. ;
Zibari, Gazi ;
Cardenes, Higinia .
PRACTICAL RADIATION ONCOLOGY, 2022, 12 (01) :28-51
[2]   INTERFRACTION AND INTRAFRACTION CHANGES IN AMPLITUDE OF BREATHING MOTION IN STEREOTACTIC LIVER RADIOTHERAPY [J].
Case, Robert B. ;
Moseley, Douglas J. ;
Sonke, Jan Jakob ;
Eccles, Cynthia L. ;
Dinniwell, Robert E. ;
Kim, John ;
Bezjak, Andrea ;
Milosevic, Michael ;
Brock, Kristy K. ;
Dawson, Laura A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (03) :918-925
[3]   FEASIBILITY OF SINGLE-ISOCENTER VOLUMETRIC MODULATED ARC RADIOSURGERY FOR TREATMENT OF MULTIPLE BRAIN METASTASES [J].
Clark, Grant M. ;
Popple, Richard A. ;
Young, P. Edward ;
Fiveash, John B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (01) :296-302
[4]   The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy [J].
Dawson, LA ;
Brock, KK ;
Kazanjian, S ;
Fitch, D ;
McGinn, CJ ;
Lawrence, TS ;
Ten Haken, RK ;
Balter, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (05) :1410-1421
[5]   INTERFRACTION LIVER SHAPE VARIABILITY AND IMPACT ON GTV POSITION DURING LIVER STEREOTACTIC RADIOTHERAPY USING ABDOMINAL COMPRESSION [J].
Eccles, Cynthia L. ;
Dawson, Laura A. ;
Moseley, Joanne L. ;
Brock, Kristy K. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (03) :938-946
[6]   Dosimetric consequences of translational and rotational errors in frame-less image-guided radiosurgery [J].
Guckenberger, Matthias ;
Roesch, Johannes ;
Baier, Kurt ;
Sweeney, Reinhart A. ;
Flentje, Michael .
RADIATION ONCOLOGY, 2012, 7
[7]   Radiosurgery of multiple brain metastases with single-isocenter dynamic conformal arcs (SIDCA) [J].
Huang, Yimei ;
Chin, Karen ;
Robbins, Jared R. ;
Kim, Jinkoo ;
Li, Haisen ;
Amro, Hanan ;
Chetty, Indrin J. ;
Gordon, James ;
Ryu, Samuel .
RADIOTHERAPY AND ONCOLOGY, 2014, 112 (01) :128-132
[8]   A method for optimizing LINAC treatment geometry for volumetric modulated arc therapy of multiple brain metastases [J].
Kang, Jun ;
Ford, Eric C. ;
Smith, Koren ;
Wong, John ;
McNutt, Todd R. .
MEDICAL PHYSICS, 2010, 37 (08) :4146-4154
[9]  
Korol RM., 2011, Int J Radiat Oncol Biol Phys, V81, pS781
[10]   Respiratory gated beam delivery cannot facilitate margin reduction, unless combined with respiratory correlated image guidance [J].
Korreman, Stine S. ;
Juhler-Nottrup, Trine ;
Boyer, Arthur L. .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :61-68