Identification of material by X-ray fluorescence analysis with a pyroelectric X-ray generator

被引:0
|
作者
Ali, M. [1 ]
Bonnett, E. [1 ]
Karataev, P. [1 ]
Kubankin, A. [2 ]
Oleinik, O. [2 ]
Margaryan, V. [3 ]
机构
[1] Royal Holloway Univ London, John Adams Inst, Egham TW20 0EX, Surrey, England
[2] Belgorod Natl Res Univ, Pobedy St 85, Belgorod 308015, Russia
[3] Inst Appl Problems Phys NAS RA, 25 Hr Nersisyan St, Yerevan 0014, Armenia
来源
JOURNAL OF INSTRUMENTATION | 2024年 / 19卷 / 07期
基金
英国科学技术设施理事会;
关键词
Accelerator Applications; Accelerator mass spectroscopy (AMS); Inspection with x-rays; X-ray generators and sources; CRYSTALS;
D O I
10.1088/1748-0221/19/07/C07003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
By changing the temperature of Lithium Tantalate (LiTaO3) single crystal at moderate vacuum conditions leads to generation of strong electric field. The uncompensated polarization during the heating or cooling of the crystal causes the ejection of electrons from either the dielectric layer on the surface of the crystal or from a metal target depending on the polarity. The electrons are accelerated and gain energy of up to 100 keV. The energy of these electrons can be determined by measuring the end-point energy of the X-ray spectrum that resulted from the electron interactions with the target. The conception of a pyroelectric accelerator enabled us to develop compact (portable) electron source, which does not require an external high-voltage and the use of hazardous materials. The compact and portable nature of pyroelectric-driven particle sources holds significant promise for applications in materials science, particularly for materials analysis methodologies. The research demonstrates the feasibility of utilizing the X-ray signal generated by irradiation with electrons to identify elements in each sample. It is revealed that employing only the electron beam enables the successful acquisition of quantitative information regarding the sample structure through pyroelectric driven PD-PIXE analysis. These findings set the stage for the development of a compact and versatile apparatus for elemental analysis of materials based on a pyroelectric source.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Identification of steel by X-ray fluorescence analysis with a pyroelectric X-ray generator
    Ida, H
    Kawai, J
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 379 (04) : 735 - 738
  • [2] Identification of steel by X-ray fluorescence analysis with a pyroelectric X-ray generator
    Hiroyuki Ida
    Jun Kawai
    Analytical and Bioanalytical Chemistry, 2004, 379 : 735 - 738
  • [3] X-ray fluorescence analysis with a pyroelectric x-ray generator
    Kawai, J
    Ida, H
    Murakami, H
    Koyama, T
    X-RAY SPECTROMETRY, 2005, 34 (06) : 521 - 524
  • [4] Identification of glass and ceramics by X-ray fluorescence analysis with a pyroelectric X-ray generator
    Ida, H
    Kawai, J
    ANALYTICAL SCIENCES, 2004, 20 (08) : 1211 - 1215
  • [5] Identification of Glass and Ceramics by X-ray Fluorescence Analysis with a Pyroelectric X-ray Generator
    Hiroyuki Ida
    Jun Kawai
    Analytical Sciences, 2004, 20 : 1211 - 1215
  • [6] Portable x-ray fluorescence spectrometer with a pyroelectric x-ray generator
    Ida, H
    Kawai, J
    X-RAY SPECTROMETRY, 2005, 34 (03) : 225 - 229
  • [7] PYROELECTRIC X-RAY GENERATOR
    BROWNRIDGE, JD
    NATURE, 1992, 358 (6384) : 287 - 288
  • [8] Focusing of x-rays emitted by a pyroelectric x-ray generator for micro x-ray fluorescence
    Wilke, Markus
    Harnisch, Karsten
    Knapp, Wolfram
    Ecke, Martin
    Halle, Thorsten
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2019, 37 (01):
  • [9] Spatially resolved X-ray fluorescence analysis with a pyroelectric X-ray emitter
    Hatakeyama, So
    Kunimura, Shinsuke
    Sasaki, Nobuharu
    Yamamoto, Takashi
    Kawai, Jun
    ANALYTICAL SCIENCES, 2008, 24 (07) : 847 - 850
  • [10] Spatially Resolved X-ray Fluorescence Analysis with a Pyroelectric X-ray Emitter
    So Hatakeyama
    Shinsuke Kunimura
    Nobuharu Sasaki
    Takashi Yamamoto
    Jun Kawai
    Analytical Sciences, 2008, 24 : 847 - 850