Computational study of transcatheter aortic valve replacement based on patient-specific models-rapid surgical planning for self-expanding valves

被引:3
作者
Meng, Zhuangyuan [1 ]
Zhang, Haishan [2 ]
Cai, Yunhan [1 ]
Gao, Yuan [2 ]
Liang, Changbin [2 ]
Wang, Jun [3 ]
Chen, Xin [4 ]
Guo, Liang [2 ]
Wang, ShengZhang [1 ,5 ]
机构
[1] Fudan Univ, Inst Biomech, Dept Aeronaut & Astronaut, Shanghai, Peoples R China
[2] China Med Univ, Hosp 1, Dept Cardiol, Shenyang, Peoples R China
[3] China Med Univ, Hosp 1, Dept Anesthesia, Shenyang, Peoples R China
[4] China Med Univ, Hosp 1, Dept Cardiovasc Ultrasound, Shenyang, Peoples R China
[5] Fudan Univ, Inst Biomed Engn Technol, Acad Engn & Technol, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
finite element analysis; transcatheter aortic valve replacement; structural simulation; self-expanding valve; computational fluid dynamics; MEDTRONIC-COREVALVE; EDWARDS-SAPIEN; IMPLANTATION; STENOSIS; IMPACT; TAVI; CALCIFICATION; SIMULATION;
D O I
10.3389/fphys.2024.1407215
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Transcatheter aortic valve replacement (TAVR) is a minimally invasive interventional solution for treating aortic stenosis. The complex post-TAVR complications are associated with the type of valve implanted and the position of the implantation. The study aimed to establish a rapid numerical research method for TAVR to assess the performance differences of self-expanding valves released at various positions. It also aimed to calculate the risks of postoperative paravalvular leak and atrioventricular conduction block, comparing these risks to clinical outcomes to verify the method's effectiveness and accuracy. Based on medical images, six cases were established, including the aortic wall, native valve and calcification; one with a bicuspid aortic valve and five with tricuspid aortic valves. The parameters for the stent materials used by the patients were customized. High strain in the contact area between the stent and the valve annulus may lead to atrioventricular conduction block. Postoperatively, the self-expanding valve maintained a circular cross-section, reducing the risk of paravalvular leak and demonstrating favorable hemodynamic characteristics, consistent with clinical observations. The outcomes of the six simulations showed no significant difference in valve frame morphology or paravalvular leak risk compared to clinical results, thereby validating the numerical simulation process proposed for quickly selecting valve models and optimal release positions, aiding in TAVR preoperative planning based on patients'geometric characteristics.
引用
收藏
页数:11
相关论文
共 30 条
  • [1] Need for Permanent Pacemaker as a Complication of Transcatheter Aortic Valve Implantation and Surgical Aortic Valve Replacement in Elderly Patients With Severe Aortic Stenosis and Similar Baseline Electrocardiographic Findings
    Bagur, Rodrigo
    Rodes-Cabau, Josep
    Gurvitch, Ronen
    Dumont, Eric
    Velianou, James L.
    Manazzoni, Juan
    Toggweiler, Stefan
    Cheung, Anson
    Ye, Jian
    Natarajan, Madhu K.
    Bainey, Kevin R.
    DeLarochelliere, Robert
    Doyle, Daniel
    Pibarot, Philippe
    Voisine, Pierre
    Cote, Melanie
    Philippon, Francois
    Webb, John G.
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (05) : 540 - 551
  • [2] Assessing the impact of including leaflets in the simulation of TAVI deployment into a patient-specific aortic root
    Bailey, Jonathon
    Curzen, Nick
    Bressloff, Neil W.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2016, 19 (07) : 733 - 744
  • [3] Fluid Structure Interaction on Paravalvular Leakage of Transcatheter Aortic Valve Implantation Related to Aortic Stenosis: A Patient-Specific Case
    Basri, Adi A.
    Zuber, Mohammad
    Basri, Ernnie, I
    Zakaria, Muhammad S.
    Aziz, Ahmad F. A.
    Tamagawa, Masaaki
    Ahmad, Kamarul A.
    [J]. COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2020, 2020
  • [4] Patient-specific simulation of transcatheter aortic valve replacement: impact of deployment options on paravalvular leakage
    Bianchi, Matteo
    Marom, Gil
    Ghosh, Ram P.
    Rotman, Oren M.
    Parikh, Puja
    Gruberg, Luis
    Bluestein, Danny
    [J]. BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2019, 18 (02) : 435 - 451
  • [5] Biondi-Zoccai G, 2014, HEART LUNG VESSEL, V6, P232
  • [6] Calcific nodule morphogenesis by heart valve interstitial cells is strain dependent
    Fisher, Charles I.
    Chen, Joseph
    Merryman, W. David
    [J]. BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2013, 12 (01) : 5 - 17
  • [7] The balloon-expandable Edwards Sapien 3 valve is superior to the self-expanding Medtronic CoreValve in patients with severe aortic stenosis undergoing transfemoral aortic valve implantation
    Gonska, Birgid
    Seeger, Julia
    Baarts, Justus
    Rodewald, Christoph
    Scharnbeck, Dominik
    Rottbauer, Wolfgang
    Woehrle, Jochen
    [J]. JOURNAL OF CARDIOLOGY, 2017, 69 (5-6) : 877 - 882
  • [8] Kanwar A., 2018, Mayo clinic proceedings
  • [9] Transcatheter or Surgical Aortic-Valve Replacement in Intermediate-Risk Patients
    Leon, Martin B.
    Smith, Craig R.
    Mack, Michael J.
    Makkar, Raj R.
    Svensson, Lars G.
    Kodali, Susheel K.
    Thourani, Vinod H.
    Tuzcu, E. Murat
    Miller, D. Craig
    Herrmann, Howard C.
    Doshi, Darshan
    Cohen, David J.
    Pichard, Augusto D.
    Kapadia, Samir
    Dewey, Todd
    Babaliaros, Vasilis
    Szeto, Wilson Y.
    Williams, Mathew R.
    Kereiakes, Dean
    Zajarias, Alan
    Greason, Kevin L.
    Whisenant, Brian K.
    Hodson, Robert W.
    Moses, Jeffrey W.
    Trento, Alfredo
    Brown, David L.
    Fearon, William F.
    Pibarot, Philippe
    Hahn, Rebecca T.
    Jaber, Wael A.
    Anderson, William N.
    Alu, Maria C.
    Webb, John G.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2016, 374 (17) : 1609 - 1620
  • [10] Comparison of Balloon-Expandable Valve and Self-Expandable Valve in Transcatheter Aortic Valve Replacement: A Patient-Specific Numerical Study
    Li, Jianming
    Yan, Wentao
    Wang, Wenshuo
    Wang, Shengzhang
    Wei, Lai
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2022, 144 (10):