Effect of auxiliary geometric devices on the accuracy of intraoral scans in full-arch implant-supported rehabilitations: An in vitro study

被引:7
作者
Canullo, Luigi [1 ]
Pesce, Paolo [1 ,4 ]
Caponio, Vito Carlo Alberto [2 ]
Iacono, Roberta [3 ]
Raffone, Carlo [4 ]
Luciani, Francesco Saverio [4 ]
Menini, Maria [1 ]
机构
[1] Univ Genoa, Dept Surg Sci, Genoa, Italy
[2] Univ Foggia, Dept Clin & Expt Med, Foggia, Italy
[3] Univ Sapienza, Sapienza, Italy
[4] Univ Genoa, Dept Surg Sci DISC, Genoa, Italy
关键词
full arch; dental implants; digital impression; intraoral scanner; DIGITAL IMPRESSIONS; OPERATOR EXPERIENCE; MULTIPLE IMPLANTS; PRECISION; TECHNOLOGY; PROSTHESES; SCANNERS; SYSTEM; FRAMEWORKS; IMPROVE;
D O I
10.1016/j.jdent.2024.104979
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim of the present in vitro study was to evaluate the effect of a novel auxiliary geometric device (AGD) on the accuracy of full-arch scans captured with 3 different intraoral scanners (IOS). Methods: An edentulous maxillary model with four internal connection implant replicas was scanned using 3 different IOS: iTero Element 5D (ITERO) (Align Technology, Tempe, AZ, USA), Trios 4 (TRIOS) (3Shape A/S, Copenhagen, Denmark), and Carestream 3700 (CS) (Carestream Dental, Atlanta, USA). Thirty-six scans were taken with each IOS, 18 with the AGD in place, and 18 without the AGD. A digital master model was created using an industrial optical scanner (ATOS compact Scan 5M, GOM GmbH, Braunschweig, Germany). The master and IOS models were aligned using the scan bodies as a reference area. A surface comparison was performed, and deviation labels were exported for each scan body to evaluate the linear and angular deviation. Total body, platform and angular deviations were measured. Results: The use of AGD resulted in a statistically significant increase of angular deviation: 0.87 degrees (SD=0.21) in the AGD group versus 0.64 degrees (SD=0.46) in the no AGD group (p-value=0.005). The difference between the AGD and no AGD groups was not statistically significant for total body and platform deviation values (p-value=0.051 and 0.302 respectively). Using AGD, ITERO showed a statistically significant increase in angular deviation (mean difference=-0.46 mu m, p-value=0.002) and a decrease in mean platform deviation (mean difference=63.19 mu m, p-value<0.001). No statistically significant differences were found for the other IOS. Conclusions: The use of AGD did not add benefit on CS and TRIOS. On ITERO, there was an improvement in platform deviation, that was outweighed by the worsening of the angular deviation. Clinical Significance: In vitro data suggest that intraoral scans can be successfully used in full-arch cases. The use of AGD has no additional benefit on CS and TRIOS. On ITERO there was an improvement in platform deviation that was outweighed by the worsening of the angular deviation. Translational application to clinical practice deserves further investigation, taking into account patient-related and anatomical variables.
引用
收藏
页数:8
相关论文
共 40 条
[1]   Digital Versus Conventional Impressions in Fixed Prosthodontics: A Review [J].
Ahlholm, Pekka ;
Sipila, Kirsi ;
Vallittu, Pekka ;
Jakonen, Minna ;
Kotiranta, Ulla .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2018, 27 (01) :35-41
[2]   A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: effects of tooth irregularity and scanning direction [J].
Anh, Ji-won ;
Park, Ji-Man ;
Chun, Youn-Sic ;
Kim, Miae ;
Kim, Minji .
KOREAN JOURNAL OF ORTHODONTICS, 2016, 46 (01) :3-12
[3]   Non-normal data: Is ANOVA still a valid option? [J].
Blanca, Maria J. ;
Alarcon, Rafael ;
Arnau, Jaume ;
Bono, Roser ;
Bendayan, Rebecca .
PSICOTHEMA, 2017, 29 (04) :552-557
[4]   In vitro comparison of five desktop scanners and an industrial scanner in the evaluation of an intraoral scanner accuracy [J].
Borbola, Daniel ;
Berkei, Gabor ;
Simon, Botond ;
Romanszky, Laszlo ;
Sersli, Gyorgy ;
DeFee, Michael ;
Renne, Walter ;
Mangano, Francesco ;
Vag, Janos .
JOURNAL OF DENTISTRY, 2023, 129
[5]   Trueness of Intraoral Scanners Considering Operator Experience and Three Different Implant Scenarios: A Preliminary Report [J].
Canullo, Luigi ;
Colombo, Marco ;
Menini, Maria ;
Sorge, Patrizia ;
Pesce, Paolo .
INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2021, 34 (02) :250-253
[6]   In vitro assessment of the accuracy of digital impressions prepared using a single system for full-arch restorations on implants [J].
Ciocca, Leonardo ;
Meneghello, Roberto ;
Monaco, Carlo ;
Savio, Gianpaolo ;
Scheda, Lorenzo ;
Gatto, Maria Rosaria ;
Baldissara, Paolo .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2018, 13 (07) :1097-1108
[7]   Accuracy of a Digital Impression System Based on Active Triangulation Technology With Blue Light for Implants: Effect of Clinically Relevant Parameters [J].
Gimenez, Beatriz ;
Oezcan, Mutlu ;
Martinez-Rus, Francisco ;
Pradies, Guillermo .
IMPLANT DENTISTRY, 2015, 24 (05) :498-504
[8]   Accuracy of a Digital Impression System Based on Parallel Confocal Laser Technology for Implants with Consideration of Operator Experience and Implant Angulation and Depth [J].
Gimenez, Beatriz ;
Oezcan, Mutlu ;
Martinez-Rus, Francisco ;
Pradies, Guillermo .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (04) :853-862
[9]   An In Vitro Study of Factors Influencing the Performance of Digital Intraoral Impressions Operating on Active Wavefront Sampling Technology with Multiple Implants in the Edentulous Maxilla [J].
Gimenez-Gonzalez, Beatriz ;
Hassan, Bassam ;
Ozcan, Mutlu ;
Pradies, Guillermo .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (08) :650-655
[10]   Continuous Scan Strategy (CSS): A Novel Technique to Improve the Accuracy of Intraoral Digital Impressions [J].
Imburgia, Mario ;
Kois, John ;
Marino, Elio ;
Lerner, Henriette ;
Mangano, Francesco G. .
EUROPEAN JOURNAL OF PROSTHODONTICS AND RESTORATIVE DENTISTRY, 2020, 28 (03) :128-141