Effects of water and other filters on fracture resolution in industrial micro-CT scanning

被引:1
作者
Mani, Mary R. [1 ]
Christensen, Angi M. [1 ]
Rafiq, Amani [1 ]
机构
[1] Fed Bur Invest, Lab Div, Quantico, VA 22135 USA
来源
FORENSIC IMAGING | 2024年 / 37卷
关键词
MicroCT; Water immersion; Filter; Resolution; Beam hardening; BEAM HARDENING FILTERS; COMPUTED-TOMOGRAPHY; IMAGE QUALITY;
D O I
10.1016/j.fri.2024.200588
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Filtering is known to improve image quality of CT scans. Water immersion is one type of filtering that has been used for CT scanning dry (skeletonized) bones, and it has been suggested that this approach can increase measurement accuracy and improve fracture resolution. These tests have previously involved clinical CT scanners. Here we use an industrial micro -CT scanner and test whether water immersion and other x-ray filtering options increases fracture resolution in reconstructed scans of dry bones. Eleven dry non-human bones were CT scanned using the same acquisition parameters, while varying filter options. Bones were scanned (1) in an unfiltered "dry" air environment, (2) using metal filters at the x-ray source, and (3) with the bones immersed in water. A small subset of bones (N = 3) was also scanned using the same parameters except increasing the number of projections acquired from 500 to 1500. Reconstructed scans were evaluated by the authors, in part using a Likert scale comparing filtered with unfiltered scans, to assess fracture resolution (overall appearance and extent). Results showed that increasing the projections resulted in the greatest improvement in fracture resolution, followed by filtering at the x-ray source. Water immersion performed poorly overall, possibly due to movement artifacts that result from this type of scanning, in which the specimen rotates on a stage during the scan. When using this type of CT scanner, if increased fracture resolution is desired, water immersion is not recommended; increasing the number of projections or filtering at the x-ray source is suggested instead.
引用
收藏
页数:6
相关论文
共 17 条
[1]   Experimental assessment of the influence of beam hardening filters on image quality and patient dose in volumetric 64-slice X-ray CT scanners [J].
Ay, Mohammad Reza ;
Mehranian, Abolfazi ;
Maleki, Asghar ;
Ghadiri, Hossien ;
Ghafarian, Pardis ;
Zaidi, Habib .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (03) :249-260
[2]  
Christensen A., 2018, FORENSIC ANTHR, V1, P124, DOI [10.5744/fa.2018.0013, DOI 10.5744/FA.2018.0013]
[3]   Virtual anthropology and forensic identification using multidetector CT [J].
Dedouit, F. ;
Savall, F. ;
Mokrane, F-Z ;
Rousseau, H. ;
Crubezy, E. ;
Rouge, D. ;
Telmon, N. .
BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1036)
[4]   Virtual anthropology: useful radiological tools for age assessment in clinical forensic medicine and thanatology [J].
Dedouit, Fabrice ;
Saint-Martin, Pauline ;
Mokrane, Fatima-Zohra ;
Savall, Frederic ;
Rousseau, Herve ;
Crubezy, Eric ;
Rouge, Daniel ;
Telmon, Norbert .
RADIOLOGIA MEDICA, 2015, 120 (09) :874-886
[5]  
Garvin Heather M, 2016, Acad Forensic Pathol, V6, P499, DOI 10.23907/2016.050
[6]   Post-mortem computed tomography: Technical principles and recommended parameter settings for high-resolution imaging [J].
Gascho, Dominic ;
Thali, Michael J. ;
Niemann, Tilo .
MEDICINE SCIENCE AND THE LAW, 2018, 58 (01) :70-82
[7]   RADid: A pictorial review of radiologic identification using postmortem CT [J].
Hatch, Gary M. ;
Dedouit, Fabrice ;
Christensen, Angi M. ;
Thali, Michael J. ;
Ruder, Thomas D. .
JOURNAL OF FORENSIC RADIOLOGY AND IMAGING, 2014, 2 (02) :52-59
[8]   A study to improve the image quality in low-dose computed tomography (SPECT) using filtration [J].
Kheruka, S. C. ;
Naithani, U. C. ;
Maurya, A. K. ;
Painuly, N. K. ;
Aggarwal, L. M. ;
Gambhir, S. .
INDIAN JOURNAL OF NUCLEAR MEDICINE, 2011, 26 (01) :14-21
[9]   Image quality produced by different cone-beam computed tomography settings [J].
Kwong, Jeffrey C. ;
Palomo, J. Martin ;
Landers, Michael A. ;
Figueroa, Alex ;
Hans, Mark G. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (02) :317-327
[10]  
North Star Imaging, 2021, How many projections do I need for my CT scan?