Optimizing the image quality of peripancreatic blood vessels through the clinical application of single-energy spectral computed tomography (CT) imaging

被引:1
|
作者
Liu, Jia-Xiong [1 ]
Zhang, Yu [1 ]
Zhou, Huan [2 ]
Zhao, Ying-Jia [1 ]
Dai, Yi-Si [2 ]
Zuo, Zi-Wei [1 ]
Wang, Jia-Ning [1 ]
Yin, Xiao-Ping [1 ]
机构
[1] Hebei Univ, Affiliated Hosp, Dept Radiol, 212 Yuhua East Rd, Baoding 071000, Peoples R China
[2] Shanghai Jiao Tong Univ, Renji Hosp, Sch Med, Dept Radiol, Shanghai, Peoples R China
关键词
Angiography; optimal energy; computed tomography (CT); X-ray; ITERATIVE RECONSTRUCTION; ADENOCARCINOMA;
D O I
10.21037/qims-23-1734
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: With the increase of pancreatic tumor patients in recent years, there is an urgent need to find a way to treat pancreatic tumors. Surgery is one of the best methods for the treatment of pancreatic tumors, the success of which depends on the evaluation of peripancreatic vessels before surgery. Computed tomography (CT), as a non-invasive, fast, and economical auxiliary examination method, is undoubtedly one of the best means of clinical auxiliary examination. In this study, we investigated the impact of single-energy spectral CT imaging on the image quality of peripancreatic blood vessels and the clinical value of low-keV imaging in enhancing the image quality of peripancreatic arteriovenous vessels. Methods: We prospectively enrolled 103 patients who underwent abdominal vascular-enhanced CT examinations at the Affiliated Hospital of Hebei University between December 2022 and May 2023 and who were all scanned with the dual-energy feature on the United Imaging ATLAS scanner. The images were reconstructed at 70 keV, mixed energy, and optimized single energy in the post-processing station of United Imaging Healthcare Technology Co., Ltd. The CT value and contrast-to-noise ratio (CNR) of the superior mesenteric artery (SMA), gastroduodenal artery (GDA), inferior pancreaticoduodenal artery (IPDA), and superior mesenteric vein (SMV) were compared across energy levels, and then the image quality was subjectively evaluated. One-way analysis of variance and rank-sum tests were utilized for the statistical analysis. Results: The CT values of SMA, GDA, IPDA, and SMV in the optimal single energy group were 358.37 +/- 70.24, 323.36 +/- 88.23, 300.76 +/- 76.27, and 257.74 +/- 20.56 Hounsfield unit (HU), respectively, which were superior to those in the mixed energy (241.66 +/- 47.69, 235.17 +/- 53.71, 207.36 +/- 45.17, and 187.39 +/- 23.21 HU) and 70 keV groups (260.89 +/- 54.27, 252.41 +/- 58.87, 223.17 +/- 43.65, and 203.18 +/- 18.17 HU) (P<0.05). The diagnostic efficacy was greater in the optimal single energy group than in the other 2 groups (4.63 +/- 0.50, 3.91 +/- 0.57, and 4.23 +/- 0.83) (P<0.05). Conclusions: The optimal single energy for showing peripancreatic blood vessels is 62 +/- 7 keV when utilizing single-energy spectral CT imaging.
引用
收藏
页码:3951 / 3958
页数:8
相关论文
共 15 条
  • [1] Dual-Energy Computed Tomography: Is There a Penalty in Image Quality and Radiation Dose Compared With Single-Energy Computed Tomography?
    Thomas, Christoph
    Ketelsen, Dominik
    Tsiflikas, Ilias
    Reimann, Anja
    Brodoefel, Harald
    Claussen, Claus D.
    Heuschmid, Martin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (02) : 309 - 315
  • [2] Single-energy pediatric chest computed tomography with spectral filtration at 100 kVp: effects on radiation parameters and image quality
    Bodelle, Boris
    Fischbach, Constanze
    Booz, Christian
    Yel, Ibrahim
    Frellesen, Claudia
    Kaup, Moritz
    Beeres, Martin
    Vogl, Thomas J.
    Scholtz, Jan-Erik
    PEDIATRIC RADIOLOGY, 2017, 47 (07) : 831 - 837
  • [3] Improving Image Quality of Pancreatic Cancer and Surrounding Vessels with Spectral CT Imaging
    Miao, Qi
    Li, Xiaoling
    Ren, Ke
    Zhao, Yu
    Zhao, Li
    Xu, Ke
    CURRENT MEDICAL IMAGING, 2015, 11 (03) : 185 - 191
  • [4] Evaluation of three-dimensional dual-energy CT cholangiopancreatography image quality in patients with pancreatobiliary dilatation: Comparison with conventional single-energy CT
    Li, Bin
    Ni, Jianming
    Chen, Fangming
    Lu, Fengqi
    Zhang, Lei
    Wu, Wenjuan
    Zhang, Zhuiyang
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2023, 11
  • [5] Virtual monochromatic spectral imaging versus linearly blended dual-energy and single-energy imaging during CT-guided biopsy needle positioning: Optimization of keV settings and impact on image quality
    Do, T. D.
    Heim, J.
    Melzig, C.
    Vollherbst, D. F.
    Kauczor, H. U.
    Skornitzke, S.
    Sommer, C. M.
    PLOS ONE, 2020, 15 (02):
  • [6] Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography
    Alikhani, Babak
    Werner, Martin
    Jamali, Leila
    Wacker, Frank
    Werncke, Thomas
    ACADEMIC RADIOLOGY, 2019, 26 (08) : 1095 - 1101
  • [7] Impact of iodinated contrast media concentration on image quality for dual-energy CT and single-energy CT with low tube voltage settings
    Sookpeng, Supawitoo
    Martin, Colin J.
    ACTA RADIOLOGICA, 2023, 64 (03) : 1047 - 1055
  • [8] Improvement of Image Quality in Unenhanced Dual-Layer CT of the Head Using Virtual Monoenergetic Images Compared With Polyenergetic Single-Energy CT
    Neuhaus, Victor
    Abdullayev, Nuran
    Hokamp, Nils Grosse
    Pahn, Gregor
    Kabbasch, Christoph
    Mpotsaris, Anastasios
    Maintz, David
    Borggrefe, Jan
    INVESTIGATIVE RADIOLOGY, 2017, 52 (08) : 470 - 476
  • [9] Comparison of Radiation Dose and Image Quality Between Split-Filter Twin Beam Dual-Energy Images and Single-Energy Images in Single-Source Contrast-Enhanced Chest Computed Tomography
    Zhou, Li-Na
    Zhao, Shi-Jun
    Wang, Ruo-Bing
    Wang, Ya-Wen
    Yang, Shou-Xin
    Wu, Ning
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2021, 45 (06) : 888 - 893
  • [10] Comparison of Image Quality and Radiation Dose Between Single-Energy and Dual-Energy Images for the Brain With Stereotactic Frames on Dual-Energy Cerebral CT
    Zhao, Xiaojing
    Chao, Wang
    Shan, Yi
    Li, Jingkai
    Zhao, Cheng
    Zhang, Miao
    Lu, Jie
    FRONTIERS IN RADIOLOGY, 2022, 2