Physics-based Modeling of Energy Threshold Induced Spectral Inconsistency and Its Adaptive Correction Scheme for photon-counting CT

被引:0
|
作者
Chen, Yuting [1 ,2 ]
Xing, Yuxiang [1 ,2 ]
Gao, Hewei [1 ,2 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[2] Minist Educ, Key Lab Particle & Radiat Imaging, Beijing 100084, Peoples R China
来源
MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1 | 2024年 / 12925卷
基金
中国国家自然科学基金; 国家重点研发计划;
关键词
Photon Counting CT; Energy Threshold bias; Spectral Inconsistency; Model-based Calibration;
D O I
10.1117/12.3007384
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The spectral inconsistency across detector pixels is still a major challenge that could directly lead to obvious artifacts in reconstructed CT images and severe inaccuracies in material discrimination. This work attempts to develop a novel approach of modeling the energy threshold induced spectral inconsistency for photon-counting CT (PCCT) to adaptively characterize and correct for spectral effects. This proposed method is based on the general model expression of photon counts, adds variables to the integration term to represent spectral distortion and furthermore takes a threshold-related nonuniformity bias factor into account at integration bounds. In this work, we assume that the variables associated with spectral distortion are mainly related to the system geometry and the detector material, there is only threshold-related nonuniformity bias factor to change with the energy-threshold, regardless of some nonideal effects like pileup and charge sharing. After simplification, we solve the linear equations to get the threshold-related nonuniformity bias at different thresholds, and the corrected photon counts can be calculated together with the spectral distortion variables have been worked out at the first reference threshold. Experimental results showed that the ring and band artifacts caused by spectral inconsistency were significantly reduced, the quality of reconstructed images is visibly improve. And the universality and robustness of our method are also demonstrated preliminary.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Adaptive Spectral Inconsistency Modeling for Photon-Counting Detector CT
    Qi, Binxiang
    Gao, Hewei
    MEDICAL PHYSICS, 2020, 47 (06) : E377 - E377
  • [2] Photon-counting CT: modeling and compensating of spectral distortion effects
    Cammin, Jochen
    Kappler, Steffen
    Weidinger, Thomas
    Taguchi, Katsuyuki
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [3] Robustness of optimal energy thresholds in photon-counting spectral CT
    Zheng, Yifan
    Yveborg, Moa
    Gronberg, Fredrik
    Xu, Cheng
    Su, Qianqian
    Danielsson, Mats
    Persson, Mats
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 953
  • [4] Image quality in photon-counting CT images as a function of energy threshold
    Rajagopal, Jayasai R.
    Jones, Elizabeth C.
    Samei, Ehsan
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [5] Spectral Imaging: Dual-Energy, MultiEnergy and Photon-Counting CT
    Uhlenbrock, Detlev
    RADIOLOGIE, 2023, 63 (01): : 82 - 83
  • [6] Comparison of Flat Field Correction Methods for Photon-Counting Spectral CT Images
    Kim, Donghyeok
    Baek, Jongduk
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [7] Comparison of Threshold Energy Calibrations of a Photon-Counting Detector and Impact on CT Reconstruction
    Rodesch, Pierre-Antoine
    Richtsmeier, Devon
    Guliyev, Elmaddin
    Iniewski, Kris
    Bazalova-Carter, Magdalena
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (03) : 263 - 272
  • [8] Impact of energy threshold on material quantification of contrast agents in photon-counting CT
    Rajagopal, Jayasai R.
    Sahbaee, Pooyan
    Abadi, Ehsan
    Negussie, Ayele H.
    Pritchard, William F.
    Wood, Bradford J.
    Jones, Elizabeth C.
    Samei, Ehsan
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [9] On the Conditioning of Spectral Channelization (Energy Binning) and Its Impact on Multi-Material Decomposition Based Spectral Imaging in Photon-Counting CT
    Ren, Yan
    Xie, Huiqiao
    Long, Wenting
    Yang, Xiaofeng
    Tang, Xiangyang
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (09) : 2678 - 2688
  • [10] Energy resolution of a segmented silicon strip detector for photon-counting spectral CT
    Xu, Cheng
    Chen, Han
    Persson, Mats
    Karlsson, Staffan
    Danielsson, Mats
    Svensson, Christer
    Bornefalk, Hans
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 715 : 11 - 17