A Validated LC-MS/MS Assay for the Quantification of Cholate Isotopes in Human Serum

被引:2
|
作者
Helmke, Steve M. [1 ]
McRae, Michael P. [2 ]
Christians, Uwe [3 ]
Shokati, Touraj [1 ]
Everson, Gregory T. [1 ]
机构
[1] HepQuant LLC, Denver, CO USA
[2] Custom DX Solut LLC, Houston, TX USA
[3] Univ Colorado, Sch Med, iC42 Clin Res & Dev, Aurora, CO 80237 USA
来源
关键词
CHRONIC HEPATITIS-C; HEPQUANT-SHUNT TEST; SEVERITY INDEX DSI; DISEASE SEVERITY; LIVER-FUNCTION; FIBROSIS; IMPROVEMENT; IMPAIRMENT; CIRRHOSIS; PRESSURE;
D O I
10.1093/jalm/jfae094
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Current methods for evaluating liver health rely on nonspecific blood tests, elastography surrogates for fibrosis, and invasive procedures, none of which directly measure liver function and physiology. Herein we present the analytical validation of a unique, highly sensitive LC-MS/MS assay and dual-sample oral (DuO) cholate challenge test to reliably quantify serial serum concentrations of cholate isotopes administered to patients with liver diseases. The clearance of administered cholate isotopes measured by the assay provides information about liver function and physiology.Methods Analytical method validation of the cholate assay analytes (endogenous unlabeled cholic acid, 24-13C-cholic acid, and 2,2,4,4-D4-cholic acid) in terms of accuracy, precision, analytical sensitivity, analytical specificity, and range of reliable response was completed in human serum samples spiked with quality controls and calibrators in accordance with applicable guidelines. DuO test parameters were validated using samples from 48 subjects representing various liver disease etiologies.Results Accuracy (mean biases) for all analytes ranged from 0.1% to 3.7%. Using a nested components-of-variance design (20 days, 2 runs per day, 2 replicates per sample), total imprecision for all analytes ranged from 2.3% to 8.4%. Lower and upper limits of quantitation were established and validated at 0.1 to 10.0 mu M. Matrix effects and potential interferents did not affect assay performance. DuO test validation met all prespecified acceptance criteria.Conclusions The method validation studies described herein established the performance characteristics in terms of accuracy, precision, analytical sensitivity, analytical specificity, reportable ranges, and reference intervals of the LC-MS/MS cholate assay and DuO test.
引用
收藏
页码:1028 / 1039
页数:12
相关论文
共 50 条
  • [31] An LC-MS/MS Validated Method for Quantification of Chlorzoxazone in Human Plasma and Its Application to a Bioequivalence Study
    He, Jiake
    Li, Ning
    Xu, Jiaqiu
    Zhu, Jing
    Yu, Yang
    Chen, Xijing
    Lu, Yang
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2019, 57 (08) : 751 - 757
  • [32] An immuno-enrichment free, validated quantification of tau protein in human CSF by LC-MS/MS
    Self, Wade
    Awwad, Khader
    Savaryn, John Paul
    Schulz, Michael
    PLOS ONE, 2022, 17 (06):
  • [33] Determination of finasteride in human serum by LC-MS/MS
    Nam, Hye-Seon
    Nam, Kyong-Hee
    Jung, Su Hee
    Lee, Jang Woo
    Kang, Jin Yeong
    Hong, Soon Keun
    Kim, Tae Sung
    Kang, Tae Seok
    Yoon, Hae-Jung
    Lee, Kwang Ho
    Rhee, Gyu-Seek
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2011, 24 (05): : 345 - 351
  • [34] Analysis of dutasteride in human serum by LC-MS/MS
    Nam, Hye-Seon
    Nam, Kyong-Hee
    Jung, Su Hee
    Lee, Jang Woo
    Kang, Jin Yeong
    Hong, Soon Keun
    Kim, Tae Sung
    Jung, Ki Kyung
    Kang, Tae Seok
    Yoon, Hae-Jung
    Lee, Kwang Ho
    Rhee, Gyu-Seek
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2012, 25 (01): : 76 - 82
  • [35] Determination of isotretinoin in human serum by LC-MS/MS
    Nam, H. S.
    Yoon, H. J.
    Jung, K. K.
    Kang, I. H.
    Kim, T. S.
    Hong, S. K.
    Kang, T. S.
    TOXICOLOGY LETTERS, 2010, 196 : S39 - S39
  • [36] Determination of Clopidogrel in Human Plasma by a Validated LC-MS/MS Method
    Svinarov, Dobrin A.
    Kasabova, Lilia V.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 138 : A021 - A021
  • [37] Validated LC-MS/MS Assay for the Quantitative Determination of Fimasartan in Human Plasma: Application to Pharmacokinetic Studies
    Yoon, Seo Hyun
    Oh, Seul
    Kim, Hwa Suk
    Yi, SoJeong
    Yu, Kyung-Sang
    Jang, In-Jin
    Cho, Joo-Youn
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2015, 53 (08) : 1250 - 1256
  • [38] Determination of Anagrelide in Human Plasma by a Validated LC-MS/MS Method
    Svinarov, D.
    Kasabova, L.
    Kostova, K.
    THERAPEUTIC DRUG MONITORING, 2013, 35 (05) : 722 - 722
  • [39] A validated surrogate analyte LC-MS/MS assay for quantitation of endogenous kynurenine and tryptophan in human plasma
    Miller, Dennis
    Tan, Lei
    Dorshorst, Drew
    Morrissey, Kari
    Mahrus, Sami
    Milanowski, Dennis
    McKnight, Janine
    Cape, Stephanie
    Dean, Brian
    Liang, Xiaorong
    BIOANALYSIS, 2018, 10 (16) : 1307 - 1317
  • [40] DETERMINATION OF METHYLPHENIDATE IN HUMAN PLASMA BY A VALIDATED LC-MS/MS METHOD
    Kasabova, L.
    Svinarov, D.
    CLINICAL THERAPEUTICS, 2015, 37 (08) : E62 - E63