An in-vivo study of the safety of copper-containing intrauterine devices in 3.0 Tesla magnetic resonance imaging

被引:0
作者
Moy, Jeffrey [1 ]
Landon, Matthew [1 ]
Vigilante, John [1 ]
Lehmann, Benjamin [1 ]
DeChambeau, Amber [1 ,2 ]
Rohlfing, Frederick [1 ]
Marks, Robert [1 ,3 ]
机构
[1] Naval Med Ctr San Diego, Dept Radiol, 34800 Bob Wilson Dr,Suite 204, San Diego, CA 92134 USA
[2] Sharp Rees Stealy Med Grp, Dept Radiol, 300 Fir St, San Diego, CA 92101 USA
[3] Univ Calif San Diego, Dept Radiol, 200 West Arbor Dr, San Diego, CA 92103 USA
关键词
Magnetic resonance imaging; Intrauterine device; Paragard T380A; Copper IUD; MRI safety; CONTRACEPTIVE DEVICES; EFFICACY;
D O I
10.1007/s00261-024-04493-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe aim of this study is to prospectively evaluate whether women with copper-containing intrauterine devices (Cu-IUD), currently listed as MR conditional, can safely undergo 3.0 Tesla (3 T) magnetic resonance imaging (MRI).Methods73 women, age 18-54 years old, with a Cu-IUD who were undergoing MRI for any reason were included consecutively. Pre- and post-MRI standard pelvic ultrasound examinations were completed to determine the appropriate pre- and post-MRI positioning of the Cu-IUD. Displaced IUDs were defined by IUD crossbars not in the fundal portion of the endometrial cavity, a visualized tip in the mid or lower uterus, any part of the device located in the cervical canal or outside of the endometrial canal, a fractured device, or a non-visualized IUD. Additionally, a questionnaire was completed by participants to determine the level of pre- and post-MRI pelvic pain.ResultsThere were zero observed displaced Cu-IUDs on post-MRI pelvic ultrasounds (p = 0/70, 95% CI 0, .043). Three participants were dropped from the study due to malpositioned IUDs on pre-MRI pelvic ultrasound. Six patients reported new or worsening pelvic pain/discomfort during or after their MRI examination.ConclusionOur results suggest that performing 3 T MRI using a low SAR setting does not cause displacement of Cu-IUDs, with zero out of 70 patients demonstrating IUD displacement.
引用
收藏
页码:4592 / 4599
页数:8
相关论文
共 25 条
  • [1] [Anonymous], 1984, PARAGARD INTRAUTERIN
  • [2] Magnetic Field Interactions of Copper-Containing Intrauterine Devices in 3.0-Tesla Magnetic Resonance Imaging: In Vivo Study
    Berger-Kulemann, Vanessa
    Einspieler, Henrik
    Hachemian, Nilouparak
    Prayer, Daniela
    Trattnig, Siegfried
    Weber, Michael
    Ba-Ssalamah, Ahmed
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2013, 14 (03) : 416 - 422
  • [3] Safety of intrauterine devices in MRI
    Bussmann, Simon
    Luechinger, Roger
    Froehlich, Johannes M.
    von Weymarn, Constantin
    Reischauer, Carolin
    Koh, Dow Mu
    Gutzeit, Andreas
    [J]. PLOS ONE, 2018, 13 (10):
  • [4] The efficacy of intrauterine devices for emergency contraception: a systematic review of 35 years of experience
    Cleland, Kelly
    Zhu, Haoping
    Goldstuck, Norman
    Cheng, Linan
    Trussell, James
    [J]. HUMAN REPRODUCTION, 2012, 27 (07) : 1994 - 2000
  • [5] Magnetic resonance imaging and gynecological devices
    Correia, Lucia
    Ramos, Ana Beatriz
    Machado, Ana Isabel
    Rosa, Duarte
    Marques, Carlos
    [J]. CONTRACEPTION, 2012, 85 (06) : 538 - 543
  • [6] Continuation and satisfaction with intrauterine copper device inserted during caesarean delivery
    de Albuquerque, Clarisse Uchoa
    Rios, Camila Elena Gomes Filgueiras
    Figueiredo, Elfie Tomaz
    Bruno, Zenilda Vieira
    Feitosa, Francisco Edson de Lucena
    Peixoto, Raquel Autran Coelho
    [J]. EUROPEAN JOURNAL OF CONTRACEPTION AND REPRODUCTIVE HEALTH CARE, 2021, 26 (06) : 486 - 490
  • [7] Emily L., 2023, INTRAUTERINE DEVICE
  • [8] IF NOTHING GOES WRONG, IS EVERYTHING ALL RIGHT - INTERPRETING ZERO NUMERATORS
    HANLEY, JA
    LIPPMANHAND, A
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1983, 249 (13): : 1743 - 1745
  • [9] Safety of intrauterine contraceptive devices during MR imaging
    Hess, T
    Stepanow, B
    Knopp, MV
    [J]. EUROPEAN RADIOLOGY, 1996, 6 (01) : 66 - 68
  • [10] Copper containing intra-uterine devices versus depot progestogens for contraception
    Hofmeyr, G. Justus
    Singata, Mandisa
    Lawrie, Theresa A.
    [J]. COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2010, (06):