From radiobiology to radiotherapy: dose homogeneity in cells after alpha irradiation in measurements and Monte Carlo simulations

被引:0
作者
Filipek, Mateusz [1 ]
Pszczolkowska-Kepa, Beata [1 ]
Kocinski, Bartlomiej [1 ]
Senkowska, Marta [1 ]
Szeflinski, Zygmunt [2 ]
Brzozowska, Beata [1 ]
机构
[1] Univ Warsaw, Inst Expt Phys, Fac Phys, Biomed Phys Div, Warsaw, Poland
[2] Univ Warsaw, Heavy Ion Lab, Warsaw, Poland
来源
POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING | 2024年 / 30卷 / 02期
关键词
Monte Carlo simulations; cell survival; dosimetry; radiation biology; high LET radiation; GEANT4; EMISSION;
D O I
10.2478/pjmpe-2024-0009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Proton radiotherapy offers an advantage in sparing healthy tissue compared to photon therapy due to the specific interaction of protons with the patient's body. In radiobiological experiments, alpha sources are commonly used instead of proton accelerators for convenience, but ensuring a uniform dose distribution is challenging. Properly designing the cell irradiation setup is crucial to reliably measure the average cellular response in such experiments. The objective of this research is to underscore the importance of dosimetric validation in radiobiological investigations. While Monte Carlo (MC) simulations offer valuable insights, their accuracy needs experimental confirmation. Once consistent results are obtained, the reliance on simulations becomes viable, as they are more efficient and less cumbersome compared to experimental procedures.Material and methods: The simulations are performed with three MC code-based tools: Geant4-DNA, GATE, and SRIM to model the alpha radiation source and calculate dose distributions for various cell irradiation scenarios. Dosimetric verification of the experimental setup containing a 241Am source is performed using radiochromic films. Additionally, a clonogenic cell survival assay is carried out for the DU145 cell line.Results: The study introduces a novel source simulation model derived from dosimetric measurements. The comparison between dosimetric results obtained with simulations and measured experimentally yields a gamma (3%/3mm) parameter value exceeding 99%. Furthermore, the LQ model parameters fitted to survival data of DU145 cells irradiated with particles emitted from 241Am source demonstrate consistency with previously published findings.Conclusions: Radiobiological experiments investigate cellular responses to various irradiation scenarios. Challenges arise with densely ionizing radiation used in clinical practice, particularly in ensuring uniform dose delivery for reliable experiments. MC codes aid in simulating dose distribution and designing irradiation systems for consistent cell treatment. However, experimental validation is essential before relying on simulation results. Once confirmed, these results offer a cost-effective and time-efficient approach to planning radiobiological experiments compared to traditional laboratory work.
引用
收藏
页码:79 / 89
页数:11
相关论文
共 21 条
  • [1] Track structure modeling in liquid water: A review of the Geant4-DNA very low energy extension of the Geant4 Monte Carlo simulation toolkit
    Bernal, M. A.
    Bordage, M. C.
    Brown, J. M. C.
    Davidkova, M.
    Delage, E.
    El Bitar, Z.
    Enger, S. A.
    Francis, Z.
    Guatelli, S.
    Ivanchenko, V. N.
    Karamitros, M.
    Kyriakou, I.
    Maigne, L.
    Meylan, S.
    Murakami, K.
    Okada, S.
    Payno, H.
    Perrot, Y.
    Petrovic, I.
    Pham, Q. T.
    Ristic-Fira, A.
    Sasaki, T.
    Stepan, V.
    Tran, H. N.
    Villagrasa, C.
    Incerti, S.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08): : 861 - 874
  • [2] DNA repair/pro-apoptotic dual-role proteins in five major DNA repair pathways: fail-safe protection against carcinogenesis
    Bernstein, C
    Bernstein, H
    Payne, CM
    Garewal, H
    [J]. MUTATION RESEARCH-REVIEWS IN MUTATION RESEARCH, 2002, 511 (02) : 145 - 178
  • [3] Freeware tool for analysing numbers and sizes of cell colonies
    Brzozowska, Beata
    Galecki, Maciej
    Tartas, Adrianna
    Ginter, Jozef
    Kazmierczak, Urszula
    Lundholm, Lovisa
    [J]. RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2019, 58 (01) : 109 - 117
  • [4] Elgqvist J, 2016, ANTICANCER RES, V36, P103
  • [5] LET dependent response of GafChromic films investigated with MeV ion beams
    Grilj, V
    Brenner, D. J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (24)
  • [6] A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems
    Hussein, Mohammad
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    Nisbet, Andrew
    Clark, Catharine H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) : 370 - 376
  • [7] Geant4-DNA example applications for track structure simulations in liquid water: A report from the Geant4-DNA Project
    Incerti, S.
    Kyriakou, I.
    Bernal, M. A.
    Bordage, M. C.
    Francis, Z.
    Guatelli, S.
    Ivanchenko, V.
    Karamitros, M.
    Lampe, N.
    Lee, S. B.
    Meylan, S.
    Min, C. H.
    Shin, W. G.
    Nieminen, P.
    Sakata, D.
    Tang, N.
    Villagrasa, C.
    Tran, H. N.
    Brown, J. M. C.
    [J]. MEDICAL PHYSICS, 2018, 45 (08) : E722 - E739
  • [8] THE GEANT4-DNA PROJECT
    Incerti, S.
    Baldacchino, G.
    Bernal, M.
    Capra, R.
    Champion, C.
    Francis, Z.
    Gueye, P.
    Mantero, A.
    Mascialino, B.
    Moretto, P.
    Nieminen, P.
    Villagrasa, C.
    Zacharatou, C.
    [J]. INTERNATIONAL JOURNAL OF MODELING SIMULATION AND SCIENTIFIC COMPUTING, 2010, 1 (02) : 157 - 178
  • [9] Comparison of GEANT4 very low energy cross section models with experimental data in water
    Incerti, S.
    Ivanchenko, A.
    Karamitros, M.
    Mantero, A.
    Moretto, P.
    Tran, H. N.
    Mascialino, B.
    Champion, C.
    Ivanchenko, V. N.
    Bernal, M. A.
    Francis, Z.
    Villagrasa, C.
    Baldacchino, G.
    Gueye, P.
    Capra, R.
    Nieminen, P.
    Zacharatou, C.
    [J]. MEDICAL PHYSICS, 2010, 37 (09) : 4692 - 4708
  • [10] Measurement of activity distribution in an Am-241 disc source using peeled-off Gafchromic EBT3 films
    Lee, Kwang-Ho
    Shin, Ji-Yong
    Kim, Eun-Hee
    [J]. APPLIED RADIATION AND ISOTOPES, 2018, 135 : 192 - 200