Validity and Reliability of the Orthelligent Pro Sensor for Measuring Single-Leg Vertical Jump Height in Healthy Athletic Adults

被引:0
作者
Pasquale, Caterina [1 ]
Pfister, Pierrette Baschung [2 ]
Kuhn, Manuel [3 ,4 ]
Stoeggl, Thomas [5 ,6 ]
机构
[1] Univ Hosp Zurich, Dept Physiotherapy & Occupat Therapy, CH-8091 Zurich, Switzerland
[2] Univ Hosp Zurich, Physiotherapy & Occupat Therapy Res Ctr, Directorate Res & Educ, CH-8091 Zurich, Switzerland
[3] Univ Zurich, Fac Med, CH-8006 Zurich, Switzerland
[4] Univ Hosp Zurich, Dept Pulmonol, CH-8091 Zurich, Switzerland
[5] Univ Salzburg, Dept Sport & Exercise Sci, A-5020 Salzburg, Austria
[6] Red Bull Athlete Performance Ctr, Salzburg, Austria
关键词
wearable sensor; single-leg countermovement jump; assessment; jumping ability; return to sport; test-retest; IDENTIFICATION; DEVICES; INDEX; APP;
D O I
10.3390/s24123699
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The Orthelligent Pro sensor is a practicable, portable measuring instrument. This study assessed the validity and reliability of this sensor in measuring single-leg countermovement jumps. Fifty healthy athletic adults participated in two measurement sessions a week apart in time. They performed single-leg countermovement jumps on the force plate while wearing the Orthelligent Pro sensor on their lower leg. During the first measurement session, Tester 1 invited the participants to make three single-leg countermovement jumps; subsequently, Tester 2 did the same. For assessing the sensor's intratester reliability, Tester 1 again invited the participants to make three single-leg countermovement jumps during the second measurement session. The sensor's validity was assessed by using the force plate results as the gold standard. To determinate the agreement between two measurements, Bland-Altman plots were created. The intertester reliability (ICC = 0.99; 0.97) and intratester reliability (ICC = 0.96; 0.82) were both excellent. The validity calculated (i) on the basis of the mean value of three jumps and (ii) on the basis of the maximum value of three jumps was very high, but it showed a systematic error. Taking this error into account, physiotherapists can use the Orthelligent Pro sensor as a valid and reliable instrument for measuring the jump height of countermovement jumps.
引用
收藏
页数:10
相关论文
共 26 条
[1]  
[Anonymous], Kistler Leistungsdiagnostik-System Quattro Jump
[2]  
[Anonymous], OPED Orthelligent Pro
[3]   Measurement errors when estimating the vertical jump height with flight time using photocell devices: the example of Optojump [J].
Attia, A. ;
Dhahbi, W. ;
Chaouachi, A. ;
Padulo, J. ;
wong, D. P. ;
Chamari, K. .
BIOLOGY OF SPORT, 2017, 34 (01) :63-70
[4]   The validity and reliability of an iPhone app for measuring vertical jump performance [J].
Balsalobre-Fernandez, Carlos ;
Glaister, Mark ;
Lockey, Richard Anthony .
JOURNAL OF SPORTS SCIENCES, 2015, 33 (15) :1574-1579
[5]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[6]  
Bloch H, 2017, TRAUMA BERUFSKRANKH, V19, P26, DOI 10.1007/s10039-017-0236-z
[7]   NOVEL TECHNOLOGIES FOUND TO BE VALID AND RELIABLE FOR THE MEASUREMENT OF VERTICAL JUMP HEIGHT WITH JUMP-AND-REACH TESTING [J].
Brooks, Edward R. ;
Benson, Amanda C. ;
Bruce, Lyndell M. .
JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2018, 32 (10) :2838-2845
[8]  
Cosmin, COSMIN Study Design Checklist for Patient-Reported Outcome Measurement Instruments
[9]  
de Vet HCW., 2011, Measurement in Medicine: A Practical Guide, DOI [10.1017/CBO9780511996214, DOI 10.1017/CBO9780511996214]
[10]   Inter-observer reproducibility of measurements of range of motion in patients with shoulder pain using a digital inclinometer -: art. no. 18 [J].
de Winter, AF ;
Heemskerk, MAMB ;
Terwee, CB ;
Jans, MP ;
Devillé, W ;
van Schaardenburg, DJ ;
Scholten, RJPM ;
Bouter, LM .
BMC MUSCULOSKELETAL DISORDERS, 2004, 5 (1)