Influence of superstructure geometry on the mechanical behavior of zirconia implant abutments: a finite element analysis

被引:10
作者
Geringer, Alexander [2 ]
Diebels, Stefan [2 ]
Nothdurft, Frank P. [1 ]
机构
[1] Univ Saarland, Dept Prosthet Dent & Dent Mat Sci, D-66421 Homburg, Germany
[2] Univ Saarland, D-66123 Saarbrucken, Germany
来源
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK | 2014年 / 59卷 / 06期
关键词
fatigue testing; finite element analysis; implant abutments; load-bearing capability; zirconia; zirconium dioxide; SINGLE-TOOTH REPLACEMENT; ALL-CERAMIC RESTORATIONS; FIXED DENTAL PROSTHESES; IN-VITRO; FRACTURE-RESISTANCE; PARTIAL DENTURES; STRAIGHT; FATIGUE; REGIONS; DESIGNS;
D O I
10.1515/bmt-2013-0088
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To predict the clinical performance of zirconia abutments, it is crucial to examine the mechanical behavior of different dental implant-abutment connection configurations. The international standard protocol for dynamic fatigue tests of dental implants (ISO 14801) allows comparing these configurations using standardized superstructure geometries. However, from a mechanical point of view, the geometry of clinical crowns causes modified boundary conditions. The purpose of this finite element (FE) study was to evaluate the influence of the superstructure geometry on the maximum stress values of zirconia abutments with a conical implant-abutment connection. Geometry models of the experimental setup described in ISO 14801 were generated using CAD software following the reconstruction of computerized tomography scans from all relevant components. These models served as a basis for an FE simulation. To reduce the numerical complexity of the FE model, the interaction between loading stamp and superstructure geometry was taken into account by defining the boundary conditions with regard to the frictional force. The results of the FE simulations performed on standardized superstructure geometry and anatomically shaped crowns showed a strong influence of the superstructure geometry and related surface orientations on the mechanical behavior of the underlying zirconia abutments. In conclusion, ISO testing of zirconia abutments should be accompanied by load-bearing capacity testing under simulated clinical conditions to predict clinical performance.
引用
收藏
页码:501 / 506
页数:6
相关论文
共 26 条
[1]  
[Anonymous], 2000, CRC MAT SCI ENG HDB, DOI DOI 10.1201/9781420038408
[2]  
[Anonymous], COMPEND CONTIN EDUC
[3]   Fracture resistance of single-tooth implant-supported all-ceramic restorations: An in vitro study [J].
Att, W ;
Kurun, S ;
Gerds, T ;
Strub, JR .
JOURNAL OF PROSTHETIC DENTISTRY, 2006, 95 (02) :111-116
[4]  
Canullo L, 2007, INT J PROSTHODONT, V20, P489
[5]   Influences of Internal Tapered Abutment Designs on Bone Stresses Around a Dental Implant: Three-Dimensional Finite Element Method With Statistical Evaluation [J].
Chu, Chun-Ming ;
Huang, Heng-Li ;
Hsu, Jui-Ting ;
Fuh, Lih-Jyh .
JOURNAL OF PERIODONTOLOGY, 2012, 83 (01) :111-118
[6]  
Dubbel H, 2001, TASCHENBUCH MASCHINE
[7]  
Glauser R, 2004, INT J PROSTHODONT, V17, P285
[8]   Biomechanical finite element analysis of small diameter and short dental implant [J].
Hasan, Istabrak ;
Heinemann, Friedhelm ;
Aitlahrach, Maria ;
Bourauel, Christoph .
BIOMEDIZINISCHE TECHNIK, 2010, 55 (06) :341-350
[9]  
IOf S., 2007, 14801 ISO
[10]   Prefabricated zirconium dioxide implant abutments for single-tooth replacement in the posterior region: evaluation of peri-implant tissues and superstructures after 12 months of function [J].
Nothdurft, Frank ;
Pospiech, Peter .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (08) :857-865