Folding Deformation of Open-Cell Stents in Carotid Artery Stenting: Report of Three Cases and Review of Literature
被引:1
|作者:
Ichikawa, Tomonori
[1
]
Yasuda, Ryuta
[1
]
Yamamoto, Atsushi
[1
]
Nishikawa, Hirofumi
[1
]
Fujimoto, Masashi
[1
]
Miura, Yoichi
[1
]
Toma, Naoki
[1
]
Suzuki, Hidenori
[1
]
机构:
[1] Mie Univ, Dept Neurosurg, Grad Sch Med, 2-174 Edobashi, Tsu, Mie 5148507, Japan
关键词:
carotid artery stenting;
folding deformation;
open-cell stent;
ENDARTERECTOMY;
PREVALENCE;
FRACTURES;
D O I:
10.5797/jnet.cr.2022-0028
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
Objective: We describe 3 cases with folding deformation of a PRECISE (Cordis, Miami, FL, USA) stent in carotid artery Case Presentations: The 3 cases with cervical carotid stenosis consisted of 3 males around 80 years old and included 2 symptomatic lesions. During CAS, distal embolic protection was established using a Mo.Ma (Medtronic, Minneapolis, Irvine, CA, USA) was employed. In all cases, a PRECISE stent was deployed after pre-dilation performed using a percutaneous transluminal angioplasty (PTA) balloon with the diameter of 2.5 to 3 mm. Post-dilation was performed after the stent deployment using a PTA balloon whose diameter was about 80% of that of the normal distal internal carotid artery. In all cases, cone-beam CT taken after the deployment of a stent showed folding deformation of the stent. In 2 cases, heavily calcified plaque hampered self-expansion of the stent, which resulted in the stent deformation. On the other hand, in the remaining 1 case, a distal shaft of the Mo.Ma caused the stent deformation, which was likely accelerated by head rotation and cervical compression that was performed to resolve difficulties for a filter retrieval device to pass through the stent, and post-dilation after the stenting. Conclusion: Heavily calcified plaque and a distal shaft of a Mo.Ma would result in stent deformation.
引用
收藏
页码:570 / 575
页数:6
相关论文