A parametric study regarding structural design of a bioprosthetic aortic valve by 3D fluid-structure interaction simulations

被引:0
|
作者
Kim, Yongwoo [1 ]
Pyo, Won Kyung [2 ]
Kim, Wan Kee [3 ]
Suh, Ga-Young [4 ]
Kang, Keonwook [1 ]
Lee, Seung Hyun [2 ]
机构
[1] Yonsei Univ, Dept Mech Engn, Seoul, South Korea
[2] Yonsei Univ, Severance Hosp, Coll Med, Dept Thorac Cardiovasc Surg, Seoul, South Korea
[3] Yonsei Univ, Yongin Severance Hosp, Coll Med, Dept Thorac & Cariovasc Surg, Yongin, Gyeonggi Do, South Korea
[4] Calif State Univ Long Beach, Dept Biomed Engn, Long Beach, CA 90840 USA
基金
新加坡国家研究基金会;
关键词
Aortic valve; Fluid-structure interaction; Hemodynamics; Biomechanics; MECHANICAL-PROPERTIES; BOVINE PERICARDIUM; SIZE; REPLACEMENT; COAPTATION; MODEL;
D O I
10.1016/j.heliyon.2024.e27310
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Since the introduction of transcatheter aortic valve (AV) implantation as a viable option, surgical bioprosthetic AVs have recently started incorporating shorter struts considering future valve -invalve procedures. However, the effect of leaflet coaptation geometry on the longevity of these valves remains unexplored. To address this gap, we performed a finite element analysis on bioprosthetic AVs with varying strut heights using a two-way fluid-structure interaction method. To establish a baseline, we used a standard height based on a rendered platform image of the CE PERIMOUNT Magna Ease valve from Edward Lifesciences in Irvine, CA. Bovine pericardium properties were assigned to the leaflets, while normal saline properties were used as the recirculating fluid in hemodynamic simulations. The physiological pressure profile of the cardiac cycle was applied between the aorta and left ventricle. We calculated blood flow velocity, effective orifice area (EOA), and mechanical stress on the leaflets. The results reveal that as the strut height increases, the stroke volume increases, leakage volume decreases, and EOA improves. Additionally, the maximum mechanical stress experienced by the leaflet decreases by 62% as the strut height increases to 1.2 times the standard height. This research highlights that a low-strut design in bioprosthetic AVs may negatively affect their durability, which can be useful in design of nextgeneration bioprosthetic AVs.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] A three-dimensional computational analysis of fluid-structure interaction in the aortic valve
    De Hart, J
    Peters, GWM
    Schreurs, PJG
    Baaijens, FPT
    JOURNAL OF BIOMECHANICS, 2003, 36 (01) : 103 - 112
  • [32] NUMERICAL SIMULATIONS OF A 3D FLUID-STRUCTURE INTERACTION MODEL FOR BLOOD FLOW IN AN ATHEROSCLEROTIC ARTERY
    Kafi, Oualid
    El Khatib, Nader
    Tiago, Jorge
    Sequeira, Adelia
    MATHEMATICAL BIOSCIENCES AND ENGINEERING, 2017, 14 (01) : 179 - 193
  • [33] Fluid-structure interaction analysis of a healthy aortic valve and its surrounding haemodynamics
    Yin, Zhongjie
    Armour, Chloee
    Kandail, Harkamaljot
    O'Regan, Declan P.
    Bahrami, Toufan
    Mirsadraee, Saeed
    Pirola, Selene
    Xu, Xiao Yun
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2024, 40 (11)
  • [34] Fluid-Structure Interaction Analysis of Papillary Muscle Forces Using a Comprehensive Mitral Valve Model with 3D Chordal Structure
    Toma, Milan
    Jensen, Morten O.
    Einstein, Daniel R.
    Yoganathan, Ajit P.
    Cochran, Richard P.
    Kunzelman, Karyn S.
    ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (04) : 942 - 953
  • [35] On the stability of the coupling of 3d and 1d fluid-structure interaction models for blood flow simulations
    Formaggia, Luca
    Moura, Alexandra
    Nobile, Fabio
    ESAIM-MATHEMATICAL MODELLING AND NUMERICAL ANALYSIS-MODELISATION MATHEMATIQUE ET ANALYSE NUMERIQUE, 2007, 41 (04): : 743 - 769
  • [36] Modeling of aortic valve stenosis using fluid-structure interaction method
    Ghasemi Pour, Mohammad Javad
    Hassani, Kamran
    Khayat, Morteza
    Etemadi Haghighi, Shahram
    PERFUSION-UK, 2022, 37 (04): : 367 - 376
  • [37] INVESTIGATION OF LEAFLET GEOMETRY IN A PERCUTANEOUS AORTIC VALVE WITH THE USE OF FLUID-STRUCTURE INTERACTION SIMULATION
    Van Aswegen, K. H. J.
    Smuts, A. N.
    Scheffer, C.
    Weich, H. S. Vh
    Doubell, A. F.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2012, 12 (01)
  • [38] Fluid-structure interaction of deformable aortic prostheses with a bileaflet mechanical valve
    de Tullio, M. D.
    Afferrante, L.
    Demelio, G.
    Pascazio, G.
    Verzicco, R.
    JOURNAL OF BIOMECHANICS, 2011, 44 (09) : 1684 - 1690
  • [39] Biomechanical Assessment of Bicuspid Aortic Valve Phenotypes: A Fluid-Structure Interaction Modelling Approach
    de Oliveira, Diana M. C.
    Abdullah, Nazirul
    Green, Naomi C.
    Espino, Daniel M.
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2020, 11 (04) : 431 - 447
  • [40] The impact of calcification patterns in transcatheter aortic valve performance: a fluid-structure interaction analysis
    Luraghi, Giulia
    Matas, Jose Felix Rodriguez
    Beretta, Marta
    Chiozzi, Nicole
    Iannetti, Laura
    Migliavacca, Francesco
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (04) : 375 - 383