Evaluation of radiation detectors for the determination of field output factors in Leksell Gamma Knife dosimetry using 3D printed phantom inserts
被引:0
作者:
Perrett, Ben
论文数: 0引用数: 0
h-index: 0
机构:
Princess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Perrett, Ben
[1
]
Seshadri, Venkatakrishnan
论文数: 0引用数: 0
h-index: 0
机构:
Princess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Seshadri, Venkatakrishnan
[1
]
Huxley, Cosmo
论文数: 0引用数: 0
h-index: 0
机构:
3DOne Australia Pty Ltd, 30 Nashos Pl, Wacol, Qld 4102, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Huxley, Cosmo
[2
]
Kumar, Sanjay
论文数: 0引用数: 0
h-index: 0
机构:
Elekta Australia & New Zealand Pty Ltd, Suite 10-02,146 Arthur St, North Sydney, NSW 2060, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Kumar, Sanjay
[3
]
Wawrzak, Michal
论文数: 0引用数: 0
h-index: 0
机构:
Princess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Wawrzak, Michal
[1
]
Ramachandran, Prabhakar
论文数: 0引用数: 0
h-index: 0
机构:
Princess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, AustraliaPrincess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
Ramachandran, Prabhakar
[1
]
机构:
[1] Princess Alexandra Hosp, Radiat Oncol, 199 Ipswich Rd, Woolloongabba, Qld 4102, Australia
[2] 3DOne Australia Pty Ltd, 30 Nashos Pl, Wacol, Qld 4102, Australia
[3] Elekta Australia & New Zealand Pty Ltd, Suite 10-02,146 Arthur St, North Sydney, NSW 2060, Australia
来源:
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS
|
2024年
/
121卷
关键词:
Gamma Knife;
microSilicon;
microDiamond;
RAZOR Nano;
Small fields;
Dosimetry;
Field output correction factors;
Field output factors;
CRYSTAL DIAMOND DETECTOR;
MONTE-CARLO;
PERFEXION;
RADIOSURGERY;
D O I:
10.1016/j.ejmp.2024.103370
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The Leksell Gamma Knife (R) PerfexionTM and IconTM have a unique geometry, containing 192 60Co sources with collimation for field sizes of 4 mm, 8 mm, and 16 mm. 4 mm and 8 mm collimated fields lack lateral charged particle equilibrium, so accurate field output factors are essential. This study performs field output factor measurements for the microDiamond, microSilicon, and RAZORTM Nano detectors. 3D printed inserts for the spherical Solid Water (R) Phantom were fabricated for microDiamond detector, the microSilicon unshielded diode and the RAZORTM Nano micro -ionisation chamber. Detectors were moved iteratively to identify the peak detector signal for each collimator, representing the effective point of measurement of the chamber. In addition, field output correction factors were calculated for each detector relative to vendor supplied Monte Carlo simulated field output factors and field output factors measured with a W2 scintillator. All field output factors where within 1.1 % for the 4 mm collimator and within 2.3 % for the 8 mm collimator. The 3D printed phantom inserts were suitable for routine measurements if the user identifies the effective point of measurement, and ensures a reproducible setup by marking the rotational alignment of the cylindrical print. Measurements with the microDiamond and microSilicon can be performed faster compared to the RAZORTM Nano due to differences in the signal to noise ratio. All detectors are suitable for field output factor measurements for the Leksell Gamma Knife (R) PerfexionTM and IconTM.