Effects of diameters of implant and abutment screw on stress distribution within dental implant and alveolar bone: A threedimensional finite element analysis

被引:5
|
作者
Shen, Yen -Wen [1 ,2 ]
Huang, Heng-Li [1 ,3 ,4 ]
Hsu, Jui-Ting [1 ,3 ]
Fuh, Lih-Jyh [1 ,2 ]
机构
[1] China Med Univ, Sch Dent, 91 Hsueh Shih Rd, Taichung 40402, Taiwan
[2] China Med Univ Hosp, Dept Dent, Taichung, Taiwan
[3] China Med Univ, Dept Biomed Engn, Taichung, Taiwan
[4] Asia Univ, Dept Bioinformat & Med Engn, Taichung, Taiwan
关键词
Abutment screw; Implant diameter; Bone; Stress; Finite element analysis; QUALITY;
D O I
10.1016/j.jds.2023.12.020
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background/purpose: Few studies have investigated the effects of abutment screw diameter in the stress of dental implants and alveolar bones under occlusal forces. In this study, we investigated how variations in implant diameter, abutment screw diameter, and bone condition affect stresses in the abutment screw, implant, and surrounding bone. Materials and methods: Three-dimensional finite element (FE) models were fabricated for dental implants with external hex -type abutments measuring 4 and 5 mm in diameter. The models also included abutment screws measuring 2.0 and 2.5 mm in diameter. Each implant model was integrated with the mandibular bone comprising the cortical bone and four types of cancellous bone. In total, 12 finite element models were generated, subjected to three different occlusal forces, and analyzed using FE software to investigate the stress distribution of dental implant and alveolar bone. Results: Wider implants demonstrated lower stresses in implant and bone compared with standard -diameter implants. The quality of cancellous bone has a minimal impact on the stress values of the implant, abutment screw, and cortical bone. Regardless of occlusal arrangement or quality of cancellous bone, a consistent pattern emerged: larger abutment screw diameters led to increased stress levels on the screws, while the stress levels in both cortical and cancellous bone showed comparatively minor fluctuations. Conclusion: Wider implants tend to have better stress distribution than standard -diameter implants. The potential advantage of augmenting the abutment screw diameter is unfavorable. It may result in elevated stresses in the implant system. <feminine ordinal indicator> 2024 Association for Dental Sciences of the Republic of China. Publishing services by Elsevier B.V. This is an open access article under the CC BY -NC -ND license (http://creativecommons. org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:1126 / 1134
页数:9
相关论文
共 50 条
  • [1] Influence of Implant/Abutment Connection on Stress Distribution to Implant-Surrounding Bone: A Finite Element Analysis
    Hanaoka, Marcia
    Alexandre Gehrke, Sergio
    Mardegan, Fabio
    Gennari, Cesar Roberto
    Taschieri, Silvio
    Del Fabbro, Massimo
    Corbella, Stefano
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2014, 23 (07): : 565 - 571
  • [2] Comparison of stress distribution in alveolar bone with different implant diameters and vertical cantilever length via the finite element method
    Sayyedi A.
    Rashidpour M.
    Fayyaz A.
    Ahmadian N.
    Dehghan M.
    Faghani F.
    Fasih P.
    Journal of Long-Term Effects of Medical Implants, 2019, 29 (01): : 37 - 43
  • [3] Finite Element Analysis of Stress Distribution in Immediately Loaded Dental Implant
    Bahrami, Babak
    Ghalichi, Farzan
    Mirzakouchaki, Behnam
    Ashtiani, Mohammed N.
    Marghoub, Arsalan
    2012 19TH IRANIAN CONFERENCE OF BIOMEDICAL ENGINEERING (ICBME), 2012, : 154 - 157
  • [4] Effects of custom-made and thread dental implant systems on the stress distribution in alveolar bone: a 3-dimensional finite element analysis
    Zhang, Xiao
    Xie, Zhangong
    Feng, Wei
    Chen, Xianshuai
    Chen, Jianyu
    SENSORS, MEASUREMENT AND INTELLIGENT MATERIALS II, PTS 1 AND 2, 2014, 475-476 : 1487 - +
  • [5] Comparative Evaluation of Three Abutment-Implant Interfaces on Stress Distribution in and Around Different Implant Systems: A Finite Element Analysis
    Kharsan, Vishwas
    Bandgar, Vaibhav
    Mirza, Aftab
    Jagtiani, Kiran
    Dhariwal, Neha
    Kore, Rahul
    CONTEMPORARY CLINICAL DENTISTRY, 2019, 10 (04) : 590 - 594
  • [6] Finite Element Analysis of Stress Distribution in Cancellous Bone During Dental Implant Pilot Drilling
    Chethan, K. N.
    Shetty, Nisha
    Shetty, Divya
    Keni, Laxmikant G.
    Futane, Mohan
    Adnur, Mahantesh
    PROSTHESIS, 2025, 7 (01):
  • [7] Stress Distribution in Cortical Bone around the Basal Implant - A Finite Element Analysis
    Roy, Anip Kumar
    Dixit, Nivedita
    Punde, Prashant
    Sinha, Koshika Tondon
    Jalaluddin, Mohammad
    Kumar, Ashish
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2021, 13 (05): : 633 - 636
  • [8] Finite element analysis of stress distribution around a dental implant with different amounts of bone loss: An in vitro study
    Jafarian, Mohammad
    Mirhashemi, Fatemeh Sadat
    Emadi, Naghmeh
    DENTAL AND MEDICAL PROBLEMS, 2019, 56 (01) : 27 - 32
  • [9] Finite-Element Analysis of Stress on Dental Implant Prosthesis
    Rubo, Jose Henrique
    Capello Souza, Edson Antonio
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (02) : 105 - 113
  • [10] Comparison of stress distribution on dental implant with different abutment types
    Jayakumar, N.
    Senthilkumar, G.
    Vigneshwaran, S.
    Kumar, K. G. Ajay
    MATERIALS TODAY-PROCEEDINGS, 2021, 45 : 7970 - 7979