Introduction Synovial fluid (SF) cultures can yield false -positive or negative results when diagnosing periprosthetic joint infection (PJI). False -positives may arise during sample collection or from laboratory contamination. Understanding false -positive SF culture rates is crucial for interpreting PJI laboratory data, yet clinical laboratories rarely report these rates. This study aimed to define the false -positive SF culture rate at a major specialized clinical laboratory. Methods This study retrospectively analyzed prospectively collected data at a single clinical laboratory that receives SF for clinical testing for PJI. A total of 180,317 periprosthetic SF samples from the hip, knee, and shoulder were identified from January 2016 to December 2023, which met the inclusion criteria for this study. Samples were classified by both a modified 2018 International Consensus Meeting (ICM) score and an inflammation score that combined the SF -C -reactive protein, alpha-defensin, SF -white blood cell count, and SFpolymorphonuclear% into one standardized metric. Logistic regression was utilized to evaluate the impact of various collection -based characteristics on culture positivity, including inflammation biomarkers, the source joint, quality control metrics, and days of specimen transport to the laboratory. SF culture falsepositivity was calculated based on the ICM category of "not -infected" or low inflammation score. Results Overall, 13.3% (23,974/180,317) of the samples were associated with a positive culture result: 12.5% for knee samples, 20.3% for hip samples, and 14.7% for shoulder samples. The false -positive SF culture rate among 131,949 samples classified as "not -infected" by the modified 2018 ICM definition was 0.47% (95%CI: 0.43 to 0.51%). Stratification by joint revealed a false -positive rate of 0.34% (95%CI: 0.31 to 0.38%) for knee samples, 1.24% (95%CI: 1.05 to 1.45%) for hip samples, and 3.02% (95%CI: 2.40 to 3.80%) for shoulder samples, with p < 0.0001 for all comparisons. The false -positive SF culture rate among 90,156 samples, representing half of all samples with the lowest standardized inflammation scores, was 0.47% (95%CI: 0.43 to 0.52%). Stratification by joint revealed a false -positive rate of 0.33% (95%CI: 0.29 to 0.37%) for knee samples, 1.45% (95%CI: 1.19 to 1.77%) for hip samples, and 3.09% (95%CI: 2.41 to 3.95%) for shoulder samples, with p<0.0001 for all comparisons. Multivariate logistic regression demonstrated the joint source (hip, shoulder) and poor sample quality as collection -based factors associated with a false -positive culture. Evaluation of a cohort of samples selected to minimize collection -based causes of false -positive culture demonstrated a false -positive rate of 0.30%, representing the ceiling limit for laboratory -based SF culture false -positivity. Conclusions This study utilizes two methods to estimate the false -positive SF culture rate at a single specialized clinical laboratory, demonstrating an overall false -positive rate of approximately 0.5%. Stratification of samples by source joint demonstrated that periprosthetic SF from the shoulder and hip have a substantially higher falsepositive culture rate than that of the knee. The lowest false -positive SF culture rate (0.30%) was observed among samples from the knee -passing quality control. Culture positivity due to contamination at this specific laboratory is less than 0.30% because all specimens undergo identical processing.
引用
收藏
页数:11
相关论文
共 21 条
  • [1] Reduced culture time and improved isolation rate through culture of sonicate fluid in blood culture bottles
    Janz, Viktor
    Trampuz, Andrej
    Perka, Carsten F.
    Wassilew, Georgi I.
    TECHNOLOGY AND HEALTH CARE, 2017, 25 (04) : 635 - 640
  • [2] Synovial fluid culture: agar plates vs. blood culture bottles for microbiological identification
    Cohen, Daniel
    Natshe, Ayman
    Ben Chetrit, Eli
    Lebel, Ehud
    Breuer, Gabriel S.
    CLINICAL RHEUMATOLOGY, 2020, 39 (01) : 275 - 279
  • [3] False-positive blood culture results in patients with hematologic malignancies
    Ebihara, Yasuhiro
    Kobayashi, Kiyoko
    Watanabe, Noriyuki
    Taji, Yoshitada
    Maeda, Tomoya
    Takahashi, Naoki
    Ishida, Akaru
    Asou, Norio
    Ikebuchi, Kenji
    JOURNAL OF INFECTION AND CHEMOTHERAPY, 2019, 25 (05) : 404 - 406
  • [4] False-Positive Automated Synovial Fluid White Blood Cell Counting Is a Concern for Both Hip and Knee Arthroplasty Aspirates
    Deirmengian, Carl A.
    Kazarian, Gregory S.
    Feeley, Scott P.
    Sizer, Stephen C.
    JOURNAL OF ARTHROPLASTY, 2020, 35 (06) : S304 - S307
  • [5] Septic arthritis caused by Granulicatella adiacens: diagnosis by inoculation of synovial fluid into blood culture bottles
    Matthew J. Hepburn
    Susan L. Fraser
    Thomas A. Rennie
    Charles M. Singleton
    Bernardo Delgado
    Rheumatology International, 2003, 23 : 255 - 257
  • [6] Septic arthritis caused by Granulicatella adiacens:: diagnosis by inoculation of synovial fluid into blood culture bottles
    Hepburn, MJ
    Fraser, SL
    Rennie, TA
    Singleton, CM
    Delgado, B
    RHEUMATOLOGY INTERNATIONAL, 2003, 23 (05) : 255 - 257
  • [7] Equine synovial sepsis laboratory submissions yield a low rate of positive bacterial culture and a high prevalence of antimicrobial resistance
    Pearson, Garett B.
    Papa, Brittany
    Mosaddegh, Abdolreza
    Cooper, Hannah
    Aprea, Melissa
    Pigott, John
    Altier, Craig
    Cazer, Casey L.
    Reesink, Heidi L.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2023, 84 (08)
  • [8] The concordance between preoperative aspiration and intraoperative synovial fluid culture results: intraoperative synovial fluid re-cultures are necessary whether the preoperative aspiration culture is positive or not
    Li, Hao
    Xu, Chi
    Hao, LiBo
    Chai, Wei
    Jun, Fu
    Chen, Jiying
    BMC INFECTIOUS DISEASES, 2021, 21 (01)
  • [9] The concordance between preoperative aspiration and intraoperative synovial fluid culture results: intraoperative synovial fluid re-cultures are necessary whether the preoperative aspiration culture is positive or not
    Hao Li
    Chi Xu
    LiBo Hao
    Wei Chai
    Fu Jun
    Jiying Chen
    BMC Infectious Diseases, 21
  • [10] Utility of Synovial Fluid Biomarkers for Culture-Positive Septic Arthritis in a Lyme Disease-Endemic Region
    Dechnik, Andzelika
    Kahane, Caroline G.
    Nigrovic, Lise E.
    Lyons, Todd W.
    PEDIATRIC EMERGENCY CARE, 2024, 40 (07) : e82 - e88