Influence of Sutureless Scleral Fixation Techniques With 3-Piece Intraocular Lenses on Dislocation Force

被引:2
作者
Zeilinger, Johannes [1 ]
Kronschlaeger, Martin [1 ]
Schlatter, Andreas [1 ]
Ruiss, Manuel [1 ]
Bayer, Natascha [1 ]
Findl, Oliver [1 ]
机构
[1] Hanusch Hosp, Vienna Inst Res Ocular Surg VIROS, Karl Landsteiner Inst, Vienna, Austria
关键词
EYES; EXTERNALIZATION; COMPLICATIONS; IMPLANTATION; OUTCOMES;
D O I
10.1016/j.ajo.2024.03.001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
center dot PURPOSE: To compare 4 different secondary intraocular lens (IOL) fixation techniques regarding the least required force to dislocate a scleral fixated 3-piece IOL in human corneoscleral donor tissue. center dot DESIGN: Experimental laboratory investigation. center dot METHODS: The least required dislocation force (LRDF) of 4 different secondary IOL fixation techniques, namely, the techniques using transscleral tunnels (TTs; as described by Scharioth), glued haptics (GHs; Agarwal), flanged haptics (FHs; Yamane), and bent haptic ends (BH; Behera/Bolz), were investigated using 40 threepiece IOLs (Sensar AR40) fixated to human scleral tissue. The main outcome of the study, dislocation force between different techniques, was measured with a tensiometer. center dot RESULTS: The force needed to dislocate the haptics was highest with the FH technique and was significantly higher than with all the other techniques (GH vs FH: - 1.02 +/- 0.02 N, P < .001; TT vs FH: -1.08 +/- 0.21 N, P < .001; BH vs FH: -1.00 +/- 0.25 N, P = .044). There was no significant difference regarding the dislocation force between the other techniques: GH vs TT (-0.06 +/- 0.100 N, P = .988), GH vs BH (-0.02 +/- 0.03 N, P = .60), TT vs BH (-0.08 +/- 0.04 N, P > .99). center dot CONCLUSIONS: The FH technique as described by Yamane proved to be the strongest form of secondary IOL fixation regarding dislocation force in this in vitro study. The other fixation techniques showed significantly less resistance to axial traction. (Am J Ophthalmol 2024;264: 229-234. (c) 2024 Elsevier Inc. All rights reserved.)
引用
收藏
页码:229 / 234
页数:6
相关论文
共 22 条
[1]   Fibrin glue-assisted sutureless posterior chamber intraocular lens implantation in eyes with deficient posterior capsules [J].
Agarwal, Amar ;
Kumar, Dhivya Ashok ;
Jacob, Soosan ;
Baid, Chandresh ;
Agarwal, Athiya ;
Srinivasan, Sridhar .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (09) :1433-1438
[2]   SIMPLIFIED TECHNIQUE FOR SUTURING DISLOCATED POSTERIOR CHAMBER INTRAOCULAR-LENS TO THE CILIARY SULCUS [J].
ANAND, R ;
BOWMAN, RW .
ARCHIVES OF OPHTHALMOLOGY, 1990, 108 (09) :1205-1206
[3]   Double-knot transscleral suture fixation technique for displaced intraocular lenses [J].
Azar, DT ;
Wiley, WF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 128 (05) :644-646
[4]   Scleral Fixation of Intraocular Lens in Aphakic Eyes without Capsular Support: Description of a New Technique [J].
Behera, Umesh Chandra ;
Thakur, Pratima Singh .
CLINICAL OPHTHALMOLOGY, 2021, 15 :4689-4696
[5]  
BLOOM SM, 1990, OPHTHALMIC SURG LAS, V21, P851
[6]  
Bolz M, 2023, 41 C ESCRS SEPT 8 12
[7]   Anterior Sclera Undergoes Thinning with Increasing Degree of Myopia [J].
Dhakal, Rohit ;
Vupparaboina, Kiran K. ;
Verkicharla, Pavan K. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (04)
[8]   A NEW TECHNIQUE FOR REPOSITIONING AND FIXATING A DISLOCATED INTRAOCULAR-LENS [J].
FRIEDBERG, MA ;
PILKERTON, AR .
ARCHIVES OF OPHTHALMOLOGY, 1992, 110 (03) :413-415
[9]   Sutureless intrascleral posterior chamber intraocular lens fixation [J].
Gabor, Scharioth G. B. ;
Pavlidis, Mitrofanis M. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (11) :1851-1854
[10]   Scleral fixation technique using 2 corneal tunnels for a dislocated intraocular lens [J].
Koh, HJ ;
Kim, CY ;
Lim, SJ ;
Kwon, OW .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (10) :1439-1441