The effect of electron backscatter and charge build up in media on beam current transformer signal for ultra-high dose rate (FLASH) electron beam monitoring

被引:2
|
作者
Pageot, Charles [1 ,2 ]
Zerouali, Karim [2 ]
Guillet, Dominique [2 ]
Muir, Bryan [3 ]
Renaud, James [3 ]
Lalonde, Arthur [2 ,4 ,5 ]
机构
[1] Ecole Polytech Montreal, Montreal, PQ, Canada
[2] Ctr Hosp Univ Montreal CHUM, Montreal, PQ, Canada
[3] CNR, Ottawa, ON, Canada
[4] Univ Montreal, Montreal, PQ, Canada
[5] Ctr Rech Ctr Hosp Univ Montreal CRCHUM, Montreal, PQ, Canada
关键词
ultra-high dose rate radiotherapy; flash radiotherapy; beam current transformer; electron UHDR radiotherapy; charge build-up; HIGH-ENERGY PHOTON; PHANTOM MATERIAL; IRRADIATION; CHAMBER; STORAGE; CELLS; WATER;
D O I
10.1088/1361-6560/ad40f7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Beam current transformers (BCT) are promising detectors for real-time beam monitoring in ultra-high dose rate (UHDR) electron radiotherapy. However, previous studies have reported a significant sensitivity of the BCT signal to changes in source-to-surface distance (SSD), field size, and phantom material which have until now been attributed to the fluctuating levels of electrons backscattered within the BCT. The purpose of this study is to evaluate this hypothesis, with the goal of understanding and mitigating the variations in BCT signal due to changes in irradiation conditions. Approach. Monte Carlo simulations and experimental measurements were conducted with a UHDR-capable intra-operative electron linear accelerator to analyze the impact of backscattered electrons on BCT signal. The potential influence of charge accumulation in media as a mechanism affecting BCT signal perturbation was further investigated by examining the effects of phantom conductivity and electrical grounding. Finally, the effectiveness of Faraday shielding to mitigate BCT signal variations is evaluated. Main Results. Monte Carlo simulations indicated that the fraction of electrons backscattered in water and on the collimator plastic at 6 and 9 MeV is lower than 1%, suggesting that backscattered electrons alone cannot account for the observed BCT signal variations. However, our experimental measurements confirmed previous findings of BCT response variation up to 15% for different field diameters. A significant impact of phantom type on BCT response was also observed, with variations in BCT signal as high as 14.1% when comparing measurements in water and solid water. The introduction of a Faraday shield to our applicators effectively mitigated the dependencies of BCT signal on SSD, field size, and phantom material. Significance. Our results indicate that variations in BCT signal as a function of SSD, field size, and phantom material are likely driven by an electric field originating in dielectric materials exposed to the UHDR electron beam. Strategies such as Faraday shielding were shown to effectively prevent these electric fields from affecting BCT signal, enabling reliable BCT-based electron UHDR beam monitoring.
引用
收藏
页数:18
相关论文
共 23 条
  • [1] Modeling ultra-high dose rate electron and proton FLASH effect with the physicochemical approach
    Tan, Hai Siong
    Teo, Kevin Boon Keng
    Dong, Lei
    Friberg, Andrew
    Koumenis, Constantinos
    Diffenderfer, Eric
    Zou, Jennifer Wei
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (14)
  • [2] Characterization of the PTB ultra-high pulse dose rate reference electron beam
    Bourgouin, Alexandra
    Knyziak, Adrian
    Marinelli, Marco
    Kranzer, Rafael
    Schueller, Andreas
    Kapsch, Ralf-Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08)
  • [3] Validation of an ultrahigh dose rate pulsed electron beam monitoring system using a current transformer for FLASH preclinical studies
    Goncalves Jorge, Patrik
    Grilj, Veljko
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Germond, Jean-Francois
    Bochud, Francois
    Bailat, Claude
    Moeckli, Raphael
    MEDICAL PHYSICS, 2022, 49 (03) : 1831 - 1838
  • [4] Ultra-high dose rate electron beams and the FLASH effect: From preclinical evidence to a new radiotherapy paradigm
    Schuler, Emil
    Acharya, Munjal
    Montay-Gruel, Pierre
    Loo, Billy W. Jr Jr
    Vozenin, Marie-Catherine
    Maxim, Peter G.
    MEDICAL PHYSICS, 2022, 49 (03) : 2082 - 2095
  • [5] Implementation and validation of a beam-current transformer on a medical pulsed electron beam LINAC for FLASH-RT beam monitoring
    Oesterle, Roxane
    Jorge, Patrik Goncalves
    Grilj, Veljko
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Germond, Jean-Francois
    Bochud, Francois
    Bailat, Claude
    Moeckli, Raphael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (11): : 165 - 171
  • [6] Stem effect-free (Y,Yb)AG-based detectors for ultra-high dose rate electron beam dosimetry
    Cova, Francesca
    Morrocchi, Matteo
    Fasoli, Mauro
    Ciarrocchi, Esther
    Pensavalle, Jake Harold
    Gallo, Salvatore
    Cavalieri, Andrea
    Zhang, Mingji
    Zhang, Kaihui
    Jia, Zhitai
    Tonelli, Mauro
    Di Martino, Fabio
    Vedda, Anna
    Bisogni, Maria Giuseppina
    Veronese, Ivan
    SENSORS AND ACTUATORS A-PHYSICAL, 2025, 390
  • [7] A dose rate independent 2D Ce-doped YAG scintillating dosimetry system for time resolved beam monitoring in ultra-high dose rate electron "FLASH"radiation therapy
    Vanreusel, Verdi
    Heinrich, Sophie
    De Kerf, Thomas
    Leblans, Paul
    Vandenbroucke, Dirk
    Vanlanduit, Steve
    Verellen, Dirk
    Gasparini, Alessia
    Nascimento, Luana de Freitas
    SENSORS AND ACTUATORS A-PHYSICAL, 2024, 371
  • [8] Commissioning of an ultra-high dose rate pulsed electron beam medical LINAC for FLASH RT preclinical animal experiments and future clinical human protocols
    Moeckli, Raphael
    Jorge, Patrik Goncalves
    Grilj, Veljko
    Oesterle, Roxane
    Cherbuin, Nicolas
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Germond, Jean-Francois
    Bochud, Francois
    Bailat, Claude
    MEDICAL PHYSICS, 2021, 48 (06) : 3134 - 3142
  • [9] Real-time monitoring of ultra-high dose rate electron beams using bremsstrahlung photons
    Kim, Hyun
    Jeong, Dong Hyeok
    Kang, Sang Koo
    Lee, Manwoo
    Lim, Heuijin
    Lee, Sang Jin
    Jang, Kyoung Won
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2023, 55 (09) : 3417 - 3422
  • [10] Consensus statement on the exploration of clinical translation and application of electron ultra-high dose rate FLASH radiotherapy
    Luo, Hui
    Yang, Chengliang
    Yue, Jinbo
    Ge, Hong
    PRECISION RADIATION ONCOLOGY, 2025, 9 (01): : 4 - 12