A comparison of the relative biological effectiveness of low energy electronic brachytherapy sources in breast tissue: a Monte Carlo study

被引:21
作者
White, Shane A. [1 ]
Reniers, Brigitte [1 ]
de Jong, Evelyn E. C. [1 ]
Rusch, Thomas [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Radiat Oncol MAASTRO, GROW Sch Oncol & Dev Biol, NL-6200 MD Maastricht, Netherlands
[2] Xoft, San Jose, CA USA
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
relative biological effectiveness; low energy brachytherapy; electronic brachytherapy; Monte Carlo; INTRAOPERATIVE RADIOTHERAPY TARGIT; CLUSTERED DNA LESIONS; X-RAYS; DAMAGE; RBE; IRRADIATION; SIMULATION; INDUCTION; DOSIMETRY; OXYGEN;
D O I
10.1088/0031-9155/61/1/383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electronic brachytherapy sources use low energy photons to treat the tumor bed during or after breast-conserving surgery. The relative biological effectiveness of two electronic brachytherapy sources was explored to determine if spectral differences due to source design influenced radiation quality and if radiation quality decreased with distance in the breast. The RBE was calculated through the number of DNA double strand breaks (RBEDSB) using the Monte Carlo damage simulator (MCDS) in combination with other Monte Carlo electron/photon spectrum calculations. 50kVp photons from the Intrabeam (Carl Zeiss Surgical) and Axxent (Xoft) through 40-mm spherical applicators were simulated to account for applicator and tissue attenuation in a variety of breast tissue compositions. 40kVp Axxent photons were also simulated. Secondary electrons (known to be responsible for most DNA damage) spectra at different distance were inputted into MCDS to calculate the RBEDSB. All RBEDSB used a cobalt-60 reference. RBEDSB data was combined with corresponding average photon spectrum energy for the Axxent and applied to model-based average photon energy distributions to produce an RBEDSB map of an accelerated partial breast irradiation (APBI) patient. Both Axxent and Intrabeam 50kVp spectra were shown to have a comparable RBEDSB of between 1.4 and 1.6 at all distances in spite of progressive beam hardening. The Axxent 40kVp also demonstrated a similar RBEDSB at distances. Most RBEDSB variability was dependent on the tissue type as was seen in rib (RBEDSB approximate to 1.4), gland (approximate to 1.55), adipose (approximate to 1.59), skin (approximate to 1.52) and lung (approximate to 1.50). RBEDSB variability between both sources was within 2%. A correlation was shown between RBEDSB and average photon energy and used to produce an RBEDSB map of a dose distribution in an APBI patient dataset. Radiation quality is very similar between electronic brachytherapy sources studied. No significant reductions in RBEDSB were observed with increasing distance from the source.
引用
收藏
页码:383 / 399
页数:17
相关论文
共 50 条
  • [41] Influence of photon energy spectra from brachytherapy sources on Monte Carlo simulations of kerma and dose rates in water and air
    Rivard, Mark J.
    Granero, Domingo
    Perez-Calatayud, Jose
    Ballester, Facundo
    MEDICAL PHYSICS, 2010, 37 (02) : 869 - 876
  • [42] Microdosimetric study on influence of low energy photons on relative biological effectiveness under therapeutic conditions using 6 MV linac
    Okamoto, Hiroyuki
    Kohno, Toshiyuki
    Kanai, Tatsuaki
    Kase, Yuki
    Matsumoto, Yoshitaka
    Furusawa, Yoshiya
    Fujita, Yukio
    Saitoh, Hidetoshi
    Itami, Jun
    MEDICAL PHYSICS, 2011, 38 (08) : 4714 - 4722
  • [43] The Relative Biological Effectiveness of Low-Dose Mammography Quality X Rays in the Human Breast MCF-10A Cell Line
    Mills, Caitlin E.
    Thome, Christopher
    Koff, David
    Andrews, David W.
    Boreham, Douglas R.
    RADIATION RESEARCH, 2015, 183 (01) : 42 - 51
  • [44] On the use of the absorbed depth-dose measurements in the beam calibration of a surface electronic high-dose-rate brachytherapy unit, a Monte Carlo-based study
    Valdes-Cortez, Christian
    Niatsetski, Yury
    Ballester, Facundo
    Vijande, Javier
    Candela-Juan, Cristian
    Perez-Calatayud, Jose
    MEDICAL PHYSICS, 2020, 47 (02) : 693 - 702
  • [45] Characteristics of miniature electronic brachytherapy x-ray sources based on TG-43U1 formalism using Monte Carlo simulation techniques
    Safigholi, Habib
    Faghihi, Reza
    Jashni, Somaye Karimi
    Meigooni, Ali S.
    MEDICAL PHYSICS, 2012, 39 (04) : 1971 - 1979
  • [46] Monte Carlo modeling of a High-Sensitivity MOSFET dosimeter for low- and medium-energy photon sources
    Wang, B
    Kim, CH
    Xu, XG
    MEDICAL PHYSICS, 2004, 31 (05) : 1003 - 1008
  • [47] Phantom scatter corrections of radiochromic films in high-energy brachytherapy dosimetry: a Monte Carlo study
    Subhalaxmi M.
    Palani Selvam T.
    Radiological Physics and Technology, 2015, 8 (2) : 215 - 223
  • [48] Analysis of relative biological effectiveness of high energy heavy ions in comparison to experimental data
    Sato, Y
    Soga, F
    JOURNAL OF RADIATION RESEARCH, 1997, 38 (02) : 103 - 110
  • [49] Propagation of polarized light in the biological tissue: a numerical study by polarized geometric Monte Carlo method
    Zhang, Yong
    Chen, Bin
    Li, Dong
    APPLIED OPTICS, 2016, 55 (10) : 2681 - 2691
  • [50] A calculation of the relative biological effectiveness of 125I and 103Pd brachytherapy sources using the concept of proximity function
    Wuu, CS
    Zaider, M
    MEDICAL PHYSICS, 1998, 25 (11) : 2186 - 2189