A comparison of the relative biological effectiveness of low energy electronic brachytherapy sources in breast tissue: a Monte Carlo study

被引:21
作者
White, Shane A. [1 ]
Reniers, Brigitte [1 ]
de Jong, Evelyn E. C. [1 ]
Rusch, Thomas [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Radiat Oncol MAASTRO, GROW Sch Oncol & Dev Biol, NL-6200 MD Maastricht, Netherlands
[2] Xoft, San Jose, CA USA
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
relative biological effectiveness; low energy brachytherapy; electronic brachytherapy; Monte Carlo; INTRAOPERATIVE RADIOTHERAPY TARGIT; CLUSTERED DNA LESIONS; X-RAYS; DAMAGE; RBE; IRRADIATION; SIMULATION; INDUCTION; DOSIMETRY; OXYGEN;
D O I
10.1088/0031-9155/61/1/383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electronic brachytherapy sources use low energy photons to treat the tumor bed during or after breast-conserving surgery. The relative biological effectiveness of two electronic brachytherapy sources was explored to determine if spectral differences due to source design influenced radiation quality and if radiation quality decreased with distance in the breast. The RBE was calculated through the number of DNA double strand breaks (RBEDSB) using the Monte Carlo damage simulator (MCDS) in combination with other Monte Carlo electron/photon spectrum calculations. 50kVp photons from the Intrabeam (Carl Zeiss Surgical) and Axxent (Xoft) through 40-mm spherical applicators were simulated to account for applicator and tissue attenuation in a variety of breast tissue compositions. 40kVp Axxent photons were also simulated. Secondary electrons (known to be responsible for most DNA damage) spectra at different distance were inputted into MCDS to calculate the RBEDSB. All RBEDSB used a cobalt-60 reference. RBEDSB data was combined with corresponding average photon spectrum energy for the Axxent and applied to model-based average photon energy distributions to produce an RBEDSB map of an accelerated partial breast irradiation (APBI) patient. Both Axxent and Intrabeam 50kVp spectra were shown to have a comparable RBEDSB of between 1.4 and 1.6 at all distances in spite of progressive beam hardening. The Axxent 40kVp also demonstrated a similar RBEDSB at distances. Most RBEDSB variability was dependent on the tissue type as was seen in rib (RBEDSB approximate to 1.4), gland (approximate to 1.55), adipose (approximate to 1.59), skin (approximate to 1.52) and lung (approximate to 1.50). RBEDSB variability between both sources was within 2%. A correlation was shown between RBEDSB and average photon energy and used to produce an RBEDSB map of a dose distribution in an APBI patient dataset. Radiation quality is very similar between electronic brachytherapy sources studied. No significant reductions in RBEDSB were observed with increasing distance from the source.
引用
收藏
页码:383 / 399
页数:17
相关论文
共 50 条
  • [31] Effective energy assessment during breast cancer intraoperative radiotherapy by low-energy X-rays: A Monte Carlo study
    Shamsabadi, Reza
    Baghani, Hamid Reza
    Mowlavi, Ali Asghar
    Azadegan, Behnam
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2021, 60 (01) : 125 - 134
  • [32] The effect of finite patient dimensions and tissue inhomogeneities on dosimetry planning of 192Ir HDR breast brachytherapy:: A Monte Carlo dose verification study
    Pantelis, E
    Papagiannis, P
    Karaiskos, P
    Angelopoulos, A
    Anagnostopoulos, G
    Baltas, D
    Zamboglou, N
    Sakelliou, L
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (05): : 1596 - 1602
  • [33] Relative biological effectiveness measurements of low energy proton and lithium beams on tumor cells
    Schuff, JA
    Policastro, L
    Durán, H
    Kreiner, AJ
    Mazal, A
    Molinari, BL
    Burlón, A
    Debray, ME
    Kesque, JM
    Somacal, H
    Stoliar, P
    Valda, A
    Bernaola, OA
    de la Hoz, AP
    Saint-Martin, G
    O'Connor, S
    Davidson, J
    Davidson, M
    Naab, F
    Ozafrán, MJ
    Vázqueza, ME
    Cáneva, S
    Delacroix, S
    Favaudon, F
    Henry, Y
    Nauraye, C
    Brune, E
    Gautier, C
    Habrand, JL
    Palmieri, M
    Ruffolo, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2002, 187 (03) : 345 - 353
  • [34] Relative biological effectiveness of photon energies used in brachytherapy and intraoperative radiotherapy techniques for two breast cancer cell lines
    Marthinsen, Anne B. L.
    Gisetstadsetstad, Ragnhild
    Danielsen, Signe
    Frengen, Jomar
    Strickert, Trond
    Lundgren, Steinar
    ACTA ONCOLOGICA, 2010, 49 (08) : 1261 - 1268
  • [35] Monte Carlo simulations of therapeutic proton beams for relative biological effectiveness of double-strand break
    Wang, Chun-Ching
    Hsiao, Y.
    Lee, Chung-Chi
    Chao, Tsi-Chian
    Wang, Chun-Chieh
    Tung, Chuan-Jong
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2012, 88 (1-2) : 158 - 163
  • [36] A novel vertebrate system for the examination and direct comparison of the relative biological effectiveness for different radiation qualities and sources
    Szabo, E. R.
    Reisz, Z.
    Polanek, R.
    Tokes, T.
    Czifrus, Sz.
    Pesznyak, Cs.
    Biro, B.
    Fenyvesi, A.
    Kiraly, B.
    Molnar, J.
    Brunner, Sz.
    Daroczi, B.
    Varga, Z.
    Hideghety, K.
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2018, 94 (11) : 985 - 995
  • [37] Monte Carlo dose calculations of beta-emitting sources for intravascular brachytherapy: A comparison between EGS4, EGSnrc, and MCNP
    Wang, RQ
    Li, XA
    MEDICAL PHYSICS, 2001, 28 (02) : 134 - 141
  • [38] Relative biological effectiveness of high-energy iron ions for micronucleus formation at low doses
    Groesser, Torsten
    Chun, Eugene
    Rydberg, Bjorn
    RADIATION RESEARCH, 2007, 168 (06) : 675 - 682
  • [39] INTERLABORATORY COMPARISON OF DIFFERENT ALPHA-PARTICLE AND RADON SOURCES - CELL-SURVIVAL AND RELATIVE BIOLOGICAL EFFECTIVENESS
    SCHWARTZ, JL
    ROTMENSCH, J
    ATCHER, RW
    JOSTES, RF
    CROSS, FT
    HUI, TE
    CHEN, D
    CARPENTER, S
    EVANS, HH
    MENCL, J
    BAKALE, G
    RAO, PS
    HEALTH PHYSICS, 1992, 62 (05): : 458 - 461
  • [40] Monte Carlo radiation dose simulations and dosimetric comparison of the model 6711 and 9011 125I brachytherapy sources
    Rivard, Mark J.
    MEDICAL PHYSICS, 2009, 36 (02) : 486 - 491