A comparison of the relative biological effectiveness of low energy electronic brachytherapy sources in breast tissue: a Monte Carlo study

被引:21
|
作者
White, Shane A. [1 ]
Reniers, Brigitte [1 ]
de Jong, Evelyn E. C. [1 ]
Rusch, Thomas [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Radiat Oncol MAASTRO, GROW Sch Oncol & Dev Biol, NL-6200 MD Maastricht, Netherlands
[2] Xoft, San Jose, CA USA
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
relative biological effectiveness; low energy brachytherapy; electronic brachytherapy; Monte Carlo; INTRAOPERATIVE RADIOTHERAPY TARGIT; CLUSTERED DNA LESIONS; X-RAYS; DAMAGE; RBE; IRRADIATION; SIMULATION; INDUCTION; DOSIMETRY; OXYGEN;
D O I
10.1088/0031-9155/61/1/383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electronic brachytherapy sources use low energy photons to treat the tumor bed during or after breast-conserving surgery. The relative biological effectiveness of two electronic brachytherapy sources was explored to determine if spectral differences due to source design influenced radiation quality and if radiation quality decreased with distance in the breast. The RBE was calculated through the number of DNA double strand breaks (RBEDSB) using the Monte Carlo damage simulator (MCDS) in combination with other Monte Carlo electron/photon spectrum calculations. 50kVp photons from the Intrabeam (Carl Zeiss Surgical) and Axxent (Xoft) through 40-mm spherical applicators were simulated to account for applicator and tissue attenuation in a variety of breast tissue compositions. 40kVp Axxent photons were also simulated. Secondary electrons (known to be responsible for most DNA damage) spectra at different distance were inputted into MCDS to calculate the RBEDSB. All RBEDSB used a cobalt-60 reference. RBEDSB data was combined with corresponding average photon spectrum energy for the Axxent and applied to model-based average photon energy distributions to produce an RBEDSB map of an accelerated partial breast irradiation (APBI) patient. Both Axxent and Intrabeam 50kVp spectra were shown to have a comparable RBEDSB of between 1.4 and 1.6 at all distances in spite of progressive beam hardening. The Axxent 40kVp also demonstrated a similar RBEDSB at distances. Most RBEDSB variability was dependent on the tissue type as was seen in rib (RBEDSB approximate to 1.4), gland (approximate to 1.55), adipose (approximate to 1.59), skin (approximate to 1.52) and lung (approximate to 1.50). RBEDSB variability between both sources was within 2%. A correlation was shown between RBEDSB and average photon energy and used to produce an RBEDSB map of a dose distribution in an APBI patient dataset. Radiation quality is very similar between electronic brachytherapy sources studied. No significant reductions in RBEDSB were observed with increasing distance from the source.
引用
收藏
页码:383 / 399
页数:17
相关论文
共 50 条
  • [1] A Monte Carlo study of the relative biological effectiveness in surface brachytherapy
    Valdes-Cortez, Christian
    Niatsetski, Yury
    Perez-Calatayud, Jose
    Ballester, Facundo
    Vijande, Javier
    MEDICAL PHYSICS, 2022, 49 (08) : 5576 - 5588
  • [2] Impact assessment of breast glandularity on relative biological effectiveness of low energy IORT X-rays through Monte Carlo simulation
    Shamsabadi, Reza
    Baghani, Hamid Reza
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 208
  • [3] Microdosimetry-based investigation of biological effectiveness of 252Cf brachytherapy source: TOPAS Monte Carlo study
    Chattaraj, Arghya
    Selvam, T. Palani
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (22)
  • [4] The difference of scoring dose to water or tissues in Monte Carlo dose calculations for low energy brachytherapy photon sources
    Landry, Guillaume
    Reniers, Brigitte
    Pignol, Jean-Philippe
    Beaulieu, Luc
    Verhaegen, Frank
    MEDICAL PHYSICS, 2011, 38 (03) : 1526 - 1533
  • [5] MONTE CARLO-BASED INVESTIGATION OF MICRODOSIMETRIC DISTRIBUTION OF HIGH ENERGY BRACHYTHERAPY SOURCES
    Chattaraj, Arghya
    Selvam, T. Palani
    Datta, D.
    RADIATION PROTECTION DOSIMETRY, 2019, 187 (01) : 115 - 128
  • [6] Monte Carlo calculation of the relative TG-43 dosimetry parameters for the INTRABEAM electronic brachytherapy source
    Alvarez, David Santiago Ayala
    Watson, Peter G. F.
    Popovic, Marija
    Heng, Veng Jean
    Evans, Michael D. C.
    Seuntjens, Jan
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (24)
  • [7] Changes in dose with segmentation of breast tissues in Monte Carlo calculations for low-energy brachytherapy
    Sutherland, J. G. H.
    Thomson, R. M.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2011, 38 (08) : 4858 - 4865
  • [8] Impact of spherical applicator diameter on relative biologic effectiveness of low energy IORT X-rays: A hybrid Monte Carlo study
    Shamsabadi, Reza
    Baghani, Hamid Reza
    Azadegan, Behnam
    Mowlavi, Ali Asghar
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 80 : 297 - 307
  • [9] A Monte Carlo study on tissue dose enhancement in brachytherapy: a comparison between gadolinium and gold nanoparticles
    Toossi, Mohammad Taghi Bahreyni
    Ghorbani, Mahdi
    Mehrpouyan, Mohammad
    Akbari, Fateme
    Sabet, Leila Sobhkhiz
    Meigooni, Ali Soleimani
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (02) : 177 - 185
  • [10] Monte Carlo aided design of an improved well-type ionization chamber for low energy brachytherapy sources
    Bohm, Tim D.
    Micka, John A.
    DeWerd, Larry A.
    MEDICAL PHYSICS, 2007, 34 (04) : 1274 - 1285