Evaluation of Selective Laser Melted Ti6Al4V/ST316L Composite and Selective Laser Sintered Polyamide 12 Implants for Orthopedic Applications: Finite Element Analysis, Physical and Mechanical Characterization, in Vitro and in Vivo Biocompatibility

被引:0
|
作者
Ismaeel, Marwa M. [1 ,2 ]
Al-Dergazly, Anwaar A. [1 ]
Al-Jaburi, Khider [3 ]
Abdulkareem, Samah K. [4 ]
机构
[1] Al Nahrain Univ, Coll Engn, Dept Laser & Optoelect Engn, Baghdad 64040, Iraq
[2] Baghdad Coll Econ Sci Univ, Comp Tech Engn Dept, Baghdad 10001, Iraq
[3] Al Farahidi Univ, Coll Tech Engn, Dept Aeronaut Tech Engn, Baghdad 00965, Iraq
[4] Al Turath Univ Coll, Baghdad, Iraq
来源
ANNALES DE CHIMIE-SCIENCE DES MATERIAUX | 2024年 / 48卷 / 02期
关键词
Selective Laser Melting (SLM); Polyamide; 12; Selective Laser Sintering (SLS); Ti6Al4V; ST316L; BEHAVIOR; 316L;
D O I
10.18280/acsm.480209
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This study examined the compatibility of an orthopedic implant made from polyamide 12 sintering, coupled with a composite made from Ti6Al4V/ST316L laser fusion, both in vitro and in vivo. SLM 3D printing was used to construct the composite implant of Ti6Al4V/ST316L, while SLS 3D printing was used for the polyamide 12 implant. SEM analysis, X-ray diffraction analysis, and energy dispersive spectroscopy (EDS) were used to characterize the fabricated implants. Mechanical properties of the implants were evaluated using a universal testing machine. Using finite element modeling, von Mises stresses within implants and human bones could be examined. It involved increasing the cell count and implanting MC3T3-E1 preosteoclasts on implants for a week. Biocompatibility was determined using an AlamarBlue (R) fluorescence test. After six weeks of implantation, rabbit femurs were stained with Hematoxylin and Eosin to determine in vivo biocompatibility of the implant. Based on the findings, both the polyamide 12 implant and the Ti6Al4V/ST316L composite SLM-3D printed by SLS-3D printing were biocompatible. Finite element analysis revealed that the maximum Von Misses stress was within acceptable limits.
引用
收藏
页码:223 / 231
页数:9
相关论文
共 24 条
  • [21] Mechanical properties and in vitro cytocompatibility of dense and porous Ti-6Al-4V ELI manufactured by selective laser melting technology for biomedical applications
    Suresh, Santhosh
    Sun, Chen-Nan
    Tekumalla, Sravya
    Rosa, Vinicius
    Nai, Sharon Mui Ling
    Wong, Raymond Chung Wen
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 123
  • [22] Evaluation of Treatment Effect and Mechanism Analysis of Ti6AL4V Porous Scaffolds Prepared by Selective Laser Melting with Different Chemical Polishing Processes
    Peng, Wen
    Cheng, Cai
    Hu, Jinwang
    Liu, Yami
    Li, Minmin
    Song, Changhui
    Shi, Wenqing
    3D PRINTING AND ADDITIVE MANUFACTURING, 2023, 11 (05) : 1746 - 1757
  • [23] Rationally designed functionally graded porous Ti6Al4V scaffolds with high strength and toughness built via selective laser melting for load-bearing orthopedic applications
    Xiong, Yin-Ze
    Gao, Rui-Ning
    Zhang, Hang
    Dong, Lan-Lan
    Li, Jian-Tao
    Li, Xiang
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 104
  • [24] Corrosion Properties of Ceramic-Based TiO2 Films on Plasma Oxidized Ti6Al4V/316L Layered Implant Structured Manufactured by Selective Laser Melting
    Tekdir, Hilmi
    Yetim, Tuba
    Yetim, Ali Fatih
    JOURNAL OF BIONIC ENGINEERING, 2021, 18 (04) : 944 - 957