Radiation doses and diagnostic reference levels for common CT scans in adults in Northwest region of Iran

被引:1
作者
Mardfar, Sina [1 ]
Ghaziyani, Mona Fazel [1 ]
Mortezazadeh, Tohid [2 ]
Zamani, Hamed [2 ]
Rahimiyan, Melika [1 ]
Khezerloo, Davood [1 ]
机构
[1] Tabriz Univ Med Sci, Fac Allied Med Sci, Dept Radiol, Tabriz, Iran
[2] Tabriz Univ Med Sci, Fac Med, Dept Med Phys, Tabriz, Iran
关键词
Computed tomography; Effective dose; Organ dose; Cancer risk; Diagnostic reference levels; Achievable doses; COMPUTED-TOMOGRAPHY EXAMINATIONS; IMAGE QUALITY; ESTABLISHMENT; SOFTWARE;
D O I
10.1007/s00411-024-01074-3
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aims to estimate organ dose and cancer risks, establish region-specific diagnostic reference levels (DRLs), and determine achievable doses (ADs) for common CT procedures in adults in the northwest of Iran. Effective and organ doses were estimated using VirtualDoseCT software in a sample of 480 adult patients who underwent head, sinus, chest, and abdomen-pelvis (AP) CT scans. The guidelines provided by the BEIR VII report were utilized to estimate cancer risks. Effective and organ doses for specific procedures were determined, with the highest mean organ dose being observed in the brain during head CT examinations, with a value of 54.02 mGy. It was observed that the lungs in chest examinations and the colon in AP examinations had the highest risk of cancer, with rates of 30.72 and 21.37 per 100,000 persons, respectively. Higher cancer risk values were generally exhibited by females compared to males. The DRLs for common CT examinations were established as follows: Head CT (CTDIvol 41 mGy, DLP 760 mGy cm), Sinus CT (CTDIvol 16 mGy, DLP 261 mGy cm), Chest CT (CTDIvol 8 mGy, DLP 287 mGy cm), and AP CT (CTDIvol 9 mGy, DLP 508 mGy cm). Significant variations in dose distribution among facilities were identified, indicating the need for optimization. The study highlights the importance of minimizing radiation exposure to critical organs and promoting patient safety in CT examinations. The establishment of region-specific DRLs and ADs can help optimize radiation doses and reduce cancer risks for patients.
引用
收藏
页码:545 / 555
页数:11
相关论文
共 41 条
  • [11] Developing diagnostic reference levels in Japan
    Kanda, Reiko
    Akahane, Masaaki
    Koba, Yusuke
    Chang, Weishan
    Akahane, Keiichi
    Okuda, Yasuo
    Hosono, Makoto
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2021, 39 (04) : 307 - 314
  • [12] Intraindividual variation of dose parameters in oncologic CT imaging
    Lange, Isabel
    Alikhani, Babak
    Wacker, Frank
    Raatschen, Hans-Juergen
    [J]. PLOS ONE, 2021, 16 (04):
  • [13] Updated Australian diagnostic reference levels for adult CT
    Lee, Kam L.
    Beveridge, Toby
    Sanagou, Masoumeh
    Thomas, Peter
    [J]. JOURNAL OF MEDICAL RADIATION SCIENCES, 2020, 67 (01) : 5 - 15
  • [14] Lyon F, 2012, IARC MONOGRAPHS EVAL
  • [15] The application of diagnostic reference levels: General principles and an Irish perspective
    Matthews, Kate
    Brennan, Patrick C.
    [J]. RADIOGRAPHY, 2009, 15 (02) : 171 - 178
  • [16] Medicine American Association of Physicists, 2008, MEASUREMENT REPORTIN
  • [17] ACR Dose Index Registry
    Morin, Richard L.
    Coombs, Laura P.
    Chatfield, Mythreyi Bhargavan
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2011, 8 (04) : 288 - 291
  • [18] Estimation of effective dose and organ cancer risk from paediatric computed tomography thorax - Abdomen - Pelvis examinations
    Muhammad, N. A.
    Karim, M. K. A.
    Hassan, H. A.
    Kamarudin, M. A.
    Wong, J. H. D.
    Ibahim, M. J.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2019, 165
  • [19] Mullins ME, 2004, AM J NEURORADIOL, V25, P533
  • [20] Establishment of diagnostic reference levels for common multi-detector computed tomography examinations in Iran
    Najafi, Mohsen
    Deevband, Mohammad Reza
    Ahmadi, Mohsen
    Kardan, Mohammad Reza
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (04) : 603 - 609