Radiation doses and diagnostic reference levels for common CT scans in adults in Northwest region of Iran

被引:1
作者
Mardfar, Sina [1 ]
Ghaziyani, Mona Fazel [1 ]
Mortezazadeh, Tohid [2 ]
Zamani, Hamed [2 ]
Rahimiyan, Melika [1 ]
Khezerloo, Davood [1 ]
机构
[1] Tabriz Univ Med Sci, Fac Allied Med Sci, Dept Radiol, Tabriz, Iran
[2] Tabriz Univ Med Sci, Fac Med, Dept Med Phys, Tabriz, Iran
关键词
Computed tomography; Effective dose; Organ dose; Cancer risk; Diagnostic reference levels; Achievable doses; COMPUTED-TOMOGRAPHY EXAMINATIONS; IMAGE QUALITY; ESTABLISHMENT; SOFTWARE;
D O I
10.1007/s00411-024-01074-3
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aims to estimate organ dose and cancer risks, establish region-specific diagnostic reference levels (DRLs), and determine achievable doses (ADs) for common CT procedures in adults in the northwest of Iran. Effective and organ doses were estimated using VirtualDoseCT software in a sample of 480 adult patients who underwent head, sinus, chest, and abdomen-pelvis (AP) CT scans. The guidelines provided by the BEIR VII report were utilized to estimate cancer risks. Effective and organ doses for specific procedures were determined, with the highest mean organ dose being observed in the brain during head CT examinations, with a value of 54.02 mGy. It was observed that the lungs in chest examinations and the colon in AP examinations had the highest risk of cancer, with rates of 30.72 and 21.37 per 100,000 persons, respectively. Higher cancer risk values were generally exhibited by females compared to males. The DRLs for common CT examinations were established as follows: Head CT (CTDIvol 41 mGy, DLP 760 mGy cm), Sinus CT (CTDIvol 16 mGy, DLP 261 mGy cm), Chest CT (CTDIvol 8 mGy, DLP 287 mGy cm), and AP CT (CTDIvol 9 mGy, DLP 508 mGy cm). Significant variations in dose distribution among facilities were identified, indicating the need for optimization. The study highlights the importance of minimizing radiation exposure to critical organs and promoting patient safety in CT examinations. The establishment of region-specific DRLs and ADs can help optimize radiation doses and reduce cancer risks for patients.
引用
收藏
页码:545 / 555
页数:11
相关论文
共 41 条
  • [1] Anam C, 2020, J PHYS C SERIES
  • [2] Establishment of diagnostic reference levels in computed tomography in two large hospitals in Oman
    Bouchareb, Yassine
    Al-Maimani, Amal
    Al-Balushi, Al-Yaqin
    Al-Kalbani, Maather
    Al-Maskari, Hasina
    Al-Dhuhli, Humoud
    Al-Kindi, Faiza
    [J]. RADIATION PROTECTION DOSIMETRY, 2023, 199 (17) : 2148 - 2155
  • [3] The risks of radiation exposure related to diagnostic imaging and how to minimise them
    Davies, H. E.
    Wathen, C. G.
    Gleeson, F. V.
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2011, 342 : 589 - 593
  • [4] Local Diagnostic Reference Levels for Common Adult Computed Tomography Procedures in Addis Ababa
    Dellie, Seife Teferi
    Tesfaw, Ambaye Fisehaw
    Kumsa, Marema Jebessa
    Tolawak, Birhanu Turi
    [J]. DOSE-RESPONSE, 2023, 21 (02):
  • [5] VirtualDose: a software for reporting organ doses from CT for adult and pediatric patients
    Ding, Aiping
    Gao, Yiming
    Liu, Haikuan
    Caracappa, Peter F.
    Long, Daniel J.
    Bolch, Wesley E.
    Liu, Bob
    Xu, X. George
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (14) : 5601 - 5625
  • [6] Diagnostic reference levels for common computed tomography (CT) examinations: results from the first Nigerian nationwide dose survey
    Ekpo, Ernest U.
    Adejoh, Thomas
    Akwo, Judith D.
    Emeka, Owujekwe C.
    Modu, Ali A.
    Abba, Mohammed
    Adesina, Kudirat A.
    Omiyi, David O.
    Chiegwu, Uche H.
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 525 - 535
  • [7] Establishment of CT diagnostic reference levels in Ireland
    Foley, S. J.
    McEntee, M. F.
    Rainford, L. A.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1018) : 1390 - 1397
  • [8] ALARA AND AN INTEGRATED APPROACH TO RADIATION PROTECTION
    HENDEE, WR
    EDWARDS, FM
    [J]. SEMINARS IN NUCLEAR MEDICINE, 1986, 16 (02) : 142 - 150
  • [9] Organ doses, effective dose, and cancer risk estimation from head and neck CT scans
    Kamrani, Samira
    Kasraie, Nima
    Jahangiri, Fatemeh
    Khezerloo, Davood
    Sheikhzadeh, Peyman
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2023, 212
  • [10] US Diagnostic Reference Levels and Achievable Doses for 10 Adult CT Examinations
    Kanal, Kalpana M.
    Butler, Priscilla F.
    Sengupta, Debapriya
    Bhargavan-Chatfield, Mythreyi
    Coombs, Laura P.
    Morin, Richard L.
    [J]. RADIOLOGY, 2017, 284 (01) : 120 - 133