Real-world evaluation of novel eye drop bottle sensors: Cloud-based AI support for eye drop adherence

被引:0
作者
Tabuchi, Hitoshi [1 ,2 ]
Nishimura, Kazuaki [2 ]
Akada, Masahiro [2 ]
Ishikami, Tomohiro [3 ]
Shirakami, Tomoki [2 ]
Kamiura, Naotake [3 ]
Kiuchi, Yoshiaki [4 ]
机构
[1] Hiroshima Univ, Dept Technol & Design Thinking Med, Hiroshima, Japan
[2] Tsukazaki Hosp, Dept Ophthalmol, Aboshi Ku Waku 68, Himeji, Japan
[3] Univ Hyogo, Grad Sch Engn, Himeji, Japan
[4] Hiroshima Univ, Dept Ophthalmol, Higashihiroshima, Japan
关键词
MEDICATION ADHERENCE; OCULAR HYPERTENSION; GLAUCOMA; INSTILLATION;
D O I
10.1016/j.heliyon.2024.e34167
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: To understand real-world eye drop adherence among glaucoma patients and evaluate the performance of our proposed cloud-based support for eye drop adherence (CASEA). Design: Prospective, observational case series. Methods: Setting: The Department of Ophthalmology at Tsukazaki Hospital. Patient or study population: Glaucoma patients treated at the hospital from May 2021 to September 2022, with 61 patients initially enrolled. Intervention or observation procedures: Pharmacists guided eye drop administration before the study. Changes in bottle orientation were detected using an accelerometer attached to the container, and acceleration waveforms and date/time data were recorded. Patients visited the clinic during the 4th and 8th weeks to report their eye drop administration, and the data were uploaded to the cloud. Main outcome measures: Two AI models (B-LSTM) were created to analyze the eye drop bottle movement time-series data for patients treating one or both eyes. The models were evaluated by comparing the true administration status with the AI model judgment. Results: Four of the 61 study subjects dropped out. The remaining 57 patients achieved recall, precision, and accuracy values of 98.6 %, 98.6 %, and 95.9 %, respectively, for the two-eyes model and 95.8 %, 98.8 %, and 95.6 % for the one-eye model. Three low-accuracy participants (77.1 %, 71.0 %, and 81.0 %) improved to 100 %, 99.1 %, and 100 %, respectively, after undergoing an additional 8-week performance validation using an aidtype container designed to ensure that the bottle was fully inverted during instillation. Conclusions: CASEA precisely monitored daily eye drop adherence and enhanced treatment efficacy by identifying patients with difficulty selfmedicating. This system has the potential to improve glaucoma patient outcomes by supporting adherence.
引用
收藏
页数:11
相关论文
共 31 条
  • [1] Smart Electronic Eyedrop Bottle for Unobtrusive Monitoring of Glaucoma Medication Adherence
    Aguilar-Rivera, Marcelo
    Erudaitius, Dieanira T.
    Wu, Vincent M.
    Tantiongloc, Justin C.
    Kang, Dae Y.
    Coleman, Todd P.
    Baxter, Sally L.
    Weinreb, Robert N.
    [J]. SENSORS, 2020, 20 (09)
  • [2] The influence of disease severity on quality of eye-drop administration in patients with glaucoma or ocular hypertension
    Aptel, F.
    Masset, H.
    Burillon, C.
    Robin, A.
    Denis, P.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (05) : 700 - 701
  • [3] Aids for eye drop administration
    Davies, Isaiah
    Williams, Andrew M.
    Muir, Kelly W.
    [J]. SURVEY OF OPHTHALMOLOGY, 2017, 62 (03) : 332 - 345
  • [4] Impact of the Hawthorne effect in a longitudinal clinical study: The case of anesthesia
    De Amici, D
    Klersy, C
    Ramajoli, F
    Brustia, L
    Politi, P
    [J]. CONTROLLED CLINICAL TRIALS, 2000, 21 (02): : 103 - 114
  • [5] Global causes of blindness and distance vision impairment 1990-2020: a systematic review and meta-analysis
    Flaxman, Seth R.
    Bourne, Rupert R. A.
    Resnikoff, Serge
    Ackland, Peter
    Braithwaite, Tasanee
    Cicinelli, Maria V.
    Das, Aditi
    Jonas, Jost B.
    Keeffe, Jill
    Kempen, John H.
    Leasher, Janet
    Limburg, Hans
    Naidoo, Kovin
    Pesudovs, Konrad
    Silvester, Alex
    Stevens, Gretchen A.
    Tahhan, Nina
    Wong, Tien Y.
    Taylor, Hugh R.
    [J]. LANCET GLOBAL HEALTH, 2017, 5 (12): : E1221 - E1234
  • [6] Prospects for lithium-ion batteries and beyond-a 2030 vision
    Grey, Clare P.
    Hall, David S.
    [J]. NATURE COMMUNICATIONS, 2020, 11 (01)
  • [7] Ishigami Tomohiro, 2021, 2021 Ninth International Symposium on Computing and Networking Workshops (CANDARW), P152, DOI 10.1109/CANDARW53999.2021.00032
  • [8] COMPLIANCE WITH TOPICAL PILOCARPINE TREATMENT
    KASS, MA
    MELTZER, DW
    GORDON, M
    COOPER, D
    GOLDBERG, J
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1986, 101 (05) : 515 - 523
  • [9] KASS MA, 1984, ARCH OPHTHALMOL-CHIC, V102, P1550
  • [10] The Increased Cost of Medical Services for People Diagnosed With Primary Open-Angle Glaucoma: A Decision Analytic Approach
    Kymes, Steven M.
    Plotzke, Michael R.
    Li, Jim Z.
    Nichol, Michael B.
    Wu, Joanne
    Fain, Joel
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 150 (01) : 74 - 81