Evaluating Efficacy and Complications of Contour Intrasaccular Device in Cerebral Aneurysm Management: A Multicenter Analysis

被引:4
作者
Gallo-Pineda, Felix [1 ]
Fernandez-Gomez, Miriam [1 ]
Dominguez-Rodriguez, Carlos [2 ]
Gallego-Leon, Jose Ignacio [2 ]
Hidalgo-Barranco, Carlos [1 ]
Diaz-Marti, Teresa [3 ]
Romance-Garcia, Antonio [3 ]
机构
[1] Torrecardenas Univ Hosp, Div Intervent Neuroradiol, Almeria, Spain
[2] Alicante Gen Univ Hosp, Div Intervent Neuroradiol, Alicante, Spain
[3] Malaga Reg Univ Hosp, Div Intervent Neuroradiol, Malaga, Spain
关键词
Aneurysm; Contour; Disruptor; Flow; Intrasaccular; WEB;
D O I
10.1016/j.wneu.2024.01.017
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
- BACKGROUND: Intrasaccular devices provide a method for treating complex aneurysms without leaving metallic materials in the parent artery. Compared to other wellstudied devices in neurointervention, the Contour device is relatively new as an intrasaccular flow diverter. This study examines its use in cases of incidental aneurysms and its application in the acute treatment of ruptured aneurysms. Additionally, it covers potential complications that may arise and methods for prevention. - METHODS: We conducted a retrospective analysis of 25 patients who underwent treatment with the Contour device at 3 hospital centers. We collected data related to age, gender, baseline modified Rankin Scale, personal habits, medical history, procedure details, and angiographic results according to the Woven endobridge occlusion scale. - RESULTS: A total of 15 patients (65.5%) achieved a satisfactory angiographic result (grade 0 e0 0 ) 1 year after embolization. Contrast stagnation was observed in 14 patients (58.3%). Intraprocedural complications, such as device displacement, were documented in 3 patients (12%), while 2 patients (8%) had aggregates attached to the device. Regarding late complications, 5 patients (20%) experienced device displacement and 1 patient had a minor stroke (4%). Retreatment was necessary for 3 patients (12%), involving a flow diverter, stenting, and coiling. CONCLUSIONS: In summary, the Contour device offers a viable option for treating complex aneurysms. While initial results are promising, it is crucial to acknowledge a learning curve to minimize complications and achieve satisfactory angiographic results without the need for additional treatments.
引用
收藏
页码:E738 / E746
页数:9
相关论文
共 15 条
[1]   The safety and effectiveness of the Woven EndoBridge (WEB) system for the treatment of wide-necked bifurcation aneurysms: final 12-month results of the pivotal WEB Intrasaccular Therapy (WEB-IT) Study [J].
Arthur, Adam S. ;
Molyneux, Andy ;
Coon, Alexander L. ;
Saatci, Isil ;
Szikora, Istvan ;
Baltacioglu, Feyyaz ;
Sultan, Ali ;
Hoit, Daniel ;
Almandoz, Josser E. Delgado ;
Elijovich, Lucas ;
Cekirge, Saru ;
Byrne, James, V ;
Fiorella, David ;
Arthur, Adam ;
Woo, Henry ;
Coon, Alexander ;
Colby, Geoffrey ;
Cekirge, Saruhan ;
Asadov, Ruslan ;
Chavali, Ram ;
Frerich, Kai ;
Delgado, Josser ;
Kayan, Yasha ;
Hanel, Ricardo ;
Sauvageau, Eric ;
Turner, Raymond ;
Chaudry, Imran ;
Spiotta, Alejandro ;
Turk, Aquila ;
Rai, Ansaar ;
Carpenter, Jeffrey ;
Boo, Sohyun ;
Jabbour, Pascal ;
Tjoumakaris, Stavropoula ;
Rosenwasser, Robert ;
Woodward, Keith ;
Boulos, Alan ;
Dalfino, John ;
Yamamoto, Junichi ;
Vora, Nirav ;
Budzik, Ronald ;
Pema, Peter ;
Davis, Thomas ;
Kallmes, David ;
Lanzino, Giuseppe ;
Cloft, Harry ;
Siddiqui, Adnan ;
Levy, Elad ;
Snyder, Kenneth ;
Lopes, Demetrius .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (09) :924-930
[2]   The Contour-Early Human Experience of a Novel Aneurysm Occlusion Device [J].
Bhogal, P. ;
Lylyk, I ;
Chudyk, J. ;
Perez, N. ;
Bleise, C. ;
Lylyk, P. .
CLINICAL NEURORADIOLOGY, 2021, 31 (01) :147-154
[3]   Endosaccular flow disruption with the Contour Neurovascular System: angiographic and clinical results in a single-center study of 60 unruptured intracranial aneurysms [J].
Biondi, Alessandra ;
Primikiris, Panagiotis ;
Vitale, Giovanni ;
Charbonnier, Guillaume .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2023, 15 (09) :838-843
[4]   Occlusion assessment of intracranial aneurysms treated with the WEB device [J].
Caroff, Jildaz ;
Mihalea, Cristian ;
Tuilier, Titien ;
Barreau, Xavier ;
Cognard, Christophe ;
Desal, Hubert ;
Pierot, Laurent ;
Arnoux, Armelle ;
Moret, Jacques ;
Spelle, Laurent .
NEURORADIOLOGY, 2016, 58 (09) :887-891
[5]   Initial Experience with the Contour Device in the Treatment of Ruptured Intracranial Wide-necked Bifurcation Aneurysms: A Single-operator, Multicenter Study [J].
Dange, Nitin Narayan ;
Roy, Joanna Mary .
NEUROLOGY INDIA, 2022, 70 (05) :2059-2064
[6]   Intrasaccular Flow Disruptor-Assisted Coiling of Intracranial Aneurysms Using the Novel Contour Neurovascular Systems and NEQSTENT: A Single-Center Safety and Feasibility Study [J].
Diana, Francesco ;
de Dios Lascuevas, Marta ;
Peschillo, Simone ;
Raz, Eytan ;
Yoshimura, Shinichi ;
Requena Ruiz, Manuel ;
Hernandez Morales, David ;
Tomasello, Alejandro .
BRAIN SCIENCES, 2022, 12 (08)
[7]  
Gallo-Pineda F, 2023, J Neurosurg Case Lessons., V5
[8]   How to WEB: a practical review of methodology for the use of the Woven EndoBridge [J].
Goyal, Nitin ;
Hoit, Daniel ;
DiNitto, Julie ;
Elijovich, Lucas ;
Fiorella, David ;
Pierot, Laurent ;
Lamin, Saleh ;
Spelle, Laurent ;
Saatci, Isil ;
Cekirge, Saru ;
Arthur, Adam S. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2020, 12 (05) :512-520
[9]   A mini-review of intrasaccular flow diverters [J].
Hecker, Constantin ;
Broussalis, Erasmia ;
Griessenauer, Christoph J. ;
Killer-Oberpfalzer, Monika .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2023, 15 (01) :70-+
[10]   A new era in the treatment of wide necked bifurcation aneurysms: Intrasaccular flow disruption [J].
Heiferman, Daniel M. ;
Goyal, Nitin ;
Inoa, Violiza ;
Nickele, Christopher M. ;
Arthur, Adam S. .
INTERVENTIONAL NEURORADIOLOGY, 2024, 30 (01) :31-36