Evaluation of the QMAC-dRAST System Version 2.5 for Rapid Antimicrobial Susceptibility Testing of Gram-Negative Bacteria From Positive Blood Culture Broth and Subcultured Colony Isolates

被引:2
|
作者
Kim, Tae Yeul [1 ]
Kang, Minhee [2 ,3 ]
Shim, Hyang Jin [4 ]
Kang, On-Kyun [1 ]
Huh, Hee Jae [1 ,3 ]
Lee, Nam Yong [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Lab Med & Genet, Seoul, South Korea
[2] Samsung Med Ctr, Smart Healthcare Res Inst, Biomed Engn Res Ctr, Seoul, South Korea
[3] Sungkyunkwan Univ, Samsung Adv Inst Hlth Sci & Technol, Dept Med Device Management & Res, Seoul, South Korea
[4] Samsung Med Ctr, Samsung Biomed Res Inst, Ctr Clin Med, Seoul, South Korea
关键词
antimicrobial resistance; antimicrobial susceptibility testing; bloodstream infections; broth microdilution; Gram-negative; positive blood culture broth; QMAC-dRAST; VITEK; 2; TURNAROUND TIME; VITEK-2; SYSTEM; IDENTIFICATION; ENTEROBACTERIACEAE; MORTALITY;
D O I
10.1002/jcla.25043
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Rapid antimicrobial susceptibility testing (AST) for bloodstream infections (BSIs) facilitates the optimization of antimicrobial therapy, preventing antimicrobial resistance and improving patient outcomes. QMAC-dRAST (QuantaMatrix Inc., Korea) is a rapid AST platform based on microfluidic chip technology that performs AST directly using positive blood culture broth (PBCB). This study evaluated the performance of QMAC-dRAST for Gram-negative bacteria using PBCB and subcultured colony isolates, comparing it with that of VITEK 2 (bioM & eacute;rieux, France) using broth microdilution (BMD) as the reference method. Methods We included 141 Gram-negative blood culture isolates from patients with BSI and 12 carbapenemase-producing clinical isolates of Enterobacterales spiked into blood culture bottles. QMAC-dRAST performance was evaluated using PBCB and colony isolates, whereas VITEK 2 and BMD were tested only on colony isolates. Results For PBCB, QMAC-dRAST achieved 92.1% categorical agreement (CA), 95.3% essential agreement (EA), with 1.8% very major errors (VMEs), 3.5% major errors (MEs), and 5.2% minor errors (mEs). With colony isolates, it exhibited 92.5% CA and 95.1% EA, with 2.0% VMEs, 3.2% MEs, and 4.8% mEs. VITEK 2 showed 94.1% CA and 96.0% EA, with 4.3% VMEs, 0.4% MEs, and 4.3% mEs. QMAC-dRAST yielded elevated error rates for specific antimicrobial agents, with high VMEs for carbapenems and aminoglycosides. The median time to result for QMAC-dRAST was 5.9 h for PBCB samples and 6.1 h for subcultured colony isolates. Conclusions The QMAC-dRAST system demonstrated considerable strengths and comparable performance to the VITEK 2 system; however, challenges were discerned with specific antimicrobial agents, underlining a necessity for improvement.
引用
收藏
页数:9
相关论文
共 48 条
  • [21] Evaluation of the Accelerate Pheno System for Fast Identification and Antimicrobial Susceptibility Testing from Positive Blood Cultures in Bloodstream Infections Caused by Gram-Negative Pathogens
    Marschal, Matthias
    Bachmaier, Johanna
    Autenrieth, Ingo
    Oberhettinger, Philipp
    Willmann, Matthias
    Peter, Silke
    JOURNAL OF CLINICAL MICROBIOLOGY, 2017, 55 (07) : 2116 - 2126
  • [22] Performance evaluation of the Specific Reveal system for rapid antibiotic susceptibility testing from positive blood cultures containing Gram-negative pathogens
    Ostermann, Greta
    Koerber-Irrgang, Barbara
    Krueger, Alexander
    Singh, Pragya
    Vo, Kenny
    Gielen, Joerg
    Aurbach, Ute
    Wisplinghoff, Hilmar
    Jazmati, Nathalie
    JOURNAL OF CLINICAL MICROBIOLOGY, 2024, 62 (12)
  • [23] Performance of the LifeScale automated rapid phenotypic antimicrobial susceptibility testing on Gram-negative rods directly from positive blood cultures
    Snyder, James W.
    Chaudhry, Nadia
    Hoffmann, Wesley
    JOURNAL OF CLINICAL MICROBIOLOGY, 2024, 62 (12)
  • [24] Systematic Evaluation of the Accelerate Pheno System for Susceptibility Testing of Gram-Negative Bacteria Isolated from Blood Cultures
    Patel, Yera A.
    Kirn, Thomas J.
    Weinstein, Melvin P.
    Uprety, Priyanka
    MICROBIOLOGY SPECTRUM, 2021, 9 (03):
  • [25] Rapid antimicrobial susceptibility testing by VITEK®2 directly from blood cultures in patients with Gram-negative rod bacteremia
    Hogan, C. A.
    Watz, N.
    Budvytiene, I.
    Banaei, N.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2019, 94 (02) : 116 - 121
  • [26] Comperative Evaluation of Vitek 2 and Etest Methods with the Referance Broth Microdilution for Antimicrobial Susceptibility Testing of Colistin Among Multi-Drug Resistant Gram-Negative Bacteria
    Ugrakli, Selin
    Dogan, Metin
    FLORA INFEKSIYON HASTALIKLARI VE KLINIK MIKROBIYOLOJI DERGISI, 2021, 26 (03): : 518 - 527
  • [27] Evaluation of Quantamatrix dRASTTM system for rapid antimicrobial susceptibility testing of bacterial isolates from positive blood cultures, in comparison with commercial Micronaut broth microdilution system
    Morecchiato, Fabio
    Coppi, Marco
    Chilleri, Chiara
    Antonelli, Alberto
    Giani, Tommaso
    Rossolini, Gian Maria
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2024, 110 (01)
  • [28] Multi-center evaluation of the Selux next-generation phenotyping system for gram-negative direct-from-positive blood culture antimicrobial susceptibility testing
    Streva, Vincent
    Gajewski, Joseph
    Pento, Jacqueline
    Chandrasekaran, Alamelu
    Green, Matt
    Lindley, Jill
    Huband, Micahel
    El Ganbour, Asmae
    Roberts, Kristen
    Flentie, Kelly
    Sherman, Jingzi
    Stern, Eric
    Berry, Gregory J.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2025,
  • [29] Antimicrobial susceptibility testing of Gram-positive and -negative bacterial isolates directly from spiked blood culture media with Raman spectroscopy
    Dekter, H. E.
    Orelio, C. C.
    Morsink, M. C.
    Tektas, S.
    Vis, B.
    te Witt, R.
    van Leeuwen, W. B.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2017, 36 (01) : 81 - 89
  • [30] Evaluation of a new Rapid Antimicrobial Susceptibility system for Gram-negative and Gram-positive bloodstream infections: speed and accuracy of Alfred 60AST
    Anton-Vazquez, Vanesa
    Adjepong, Samuel
    Suarez, Cristina
    Planche, Timothy
    BMC MICROBIOLOGY, 2019, 19 (01)