A Critical Discussion of Alert Evaluations in the Context of Continuous Glucose Monitoring System Performance

被引:0
作者
Pleus, Stefan [1 ]
Eichenlaub, Manuel [1 ]
Waldenmaier, Delia [1 ]
Freckmann, Guido [1 ]
机构
[1] Univ Ulm, Inst Diabet Technol Forsch & Entwicklungsgesellsch, Lise Meitner Str 8-2, D-89081 Ulm, Germany
来源
JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY | 2024年 / 18卷 / 04期
关键词
alert evaluation; alert threshold; continuous glucose monitoring; predictive alert; rate of change; PROSPECTIVE MULTICENTER EVALUATION; ACCURACY; SENSOR; HYPOGLYCEMIA; RELIABILITY; PRECISE; CGMS;
D O I
10.1177/19322968241236504
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many continuous glucose monitoring (CGM) systems provide functionality which alerts users of potentially unwanted glycemic conditions. These alerts can include glucose threshold alerts to call the user's attention to hypoglycemia or hyperglycemia, predictive alerts warning about impeding hypoglycemia or hyperglycemia, and rate-of-change alerts. A recent review identified 129 articles about CGM performance studies, of which approximately 25% contained alert evaluations. In some studies, real alerts were assessed; however, most of these studies retrospectively determined the timing of CGM alerts because not all CGM systems record alerts which necessitates manual documentation. In contrast to assessment of real alerts, retrospective determination allows assessment of a variety of alert settings for all three types of glycemic condition alerts. Based on the literature and the Clinical and Laboratory Standards Institute's POCT05 guideline, two common approaches to threshold alert evaluation were identified, one value-based and one episode-based approach. In this review, a critical discussion of the two approaches, including a post hoc analysis of clinical study data, indicates that the episode-based approach should be preferred over the value-based approach. For predictive alerts, fewer results were found in the literature, and retrospective determination of CGM alert timing is complicated by the prediction algorithms being proprietary information. Rate-of-change alert evaluations were not reported in the identified literature, and POCT05 does not contain recommendations for assessment. A possible approach is discussed including post hoc analysis of clinical study data. To conclude, CGM systems should record alerts, and the episode-based approach to alert evaluation should be preferred.
引用
收藏
页码:847 / 856
页数:10
相关论文
共 47 条
  • [1] Abbott Diabetes Care, 2022, FREESTYLE LIBRE 3 US
  • [2] Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With Advanced Algorithm in Pediatric and Adult Population With Diabetes
    Alva, Shridhara
    Bailey, Timothy
    Brazg, Ronald
    Budiman, Erwin S.
    Castorino, Kristin
    Christiansen, Mark P.
    Forlenza, Gregory
    Kipnes, Mark
    Liljenquist, David R.
    Liu, Hanqing
    [J]. JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2022, 16 (01): : 70 - 77
  • [3] Bailey Timothy S, 2015, J Diabetes Sci Technol, V9, P209, DOI 10.1177/1932296814559746
  • [4] Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose Sensor
    Bailey, Timothy S.
    Ahmann, Andrew
    Brazg, Ronald
    Christiansen, Mark
    Garg, Satish
    Watkins, Elaine
    Welsh, John B.
    Lee, Scott W.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 (05) : 277 - 283
  • [5] Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia
    Bay, Christiane
    Kristensen, Peter Lommer
    Pedersen-Bjergaard, Ulrik
    Tarnow, Lise
    Thorsteinsson, Birger
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (05) : 371 - 377
  • [6] Bode Bruce, 2004, Diabetes Technol Ther, V6, P105, DOI 10.1089/152091504773731285
  • [7] Bode Bruce, 2008, Manag Care, V17, P40
  • [8] Effectiveness of adding alarms to flash glucose monitoring in adults with type 1 diabetes under routine care
    Boscari, Federico
    Ferretto, Sara
    Cavallin, Francesco
    Fadini, Gian Paolo
    Avogaro, Angelo
    Bruttomesso, Daniela
    [J]. ACTA DIABETOLOGICA, 2022, 59 (07) : 921 - 928
  • [9] Buckingham BA, 2006, DIABETES TECHNOL THE, V8, P318
  • [10] A New-Generation Continuous Glucose Monitoring System: Improved Accuracy and Reliability Compared with a Previous-Generation System
    Christiansen, Mark
    Bailey, Timothy
    Watkins, Elaine
    Liljenquist, David
    Price, David
    Nakamura, Katherine
    Boock, Robert
    Peyser, Thomas
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (10) : 881 - 888