A comparison of intraocular pressure measurement using SUOER SW-500 rebound tonometer and conventional reusable Goldmann prisms

被引:1
作者
Chaung, Jia Quan [1 ,2 ]
Sangapillai, Thanendthire [1 ]
Quilat, Karen Kate [1 ]
Perera, Shamira [1 ,2 ,3 ]
机构
[1] Singapore Natl Eye Ctr, Singapore City, Singapore
[2] Singapore Eye Res Inst, Singapore City, Singapore
[3] Duke NUS Med Sch, Singapore City, Singapore
关键词
glaucoma; rebound tonometer; IOP; intraocular pressure; Goldmann applanation tonometry; APPLANATION TONOMETER; ICARE; GLAUCOMA; AGREEMENT;
D O I
10.3389/fmed.2024.1269332
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GA(1,2)T) and SUOER SW-500 Rebound Tonometer. Methods: This was a retrospective observational study where 205 eyes of 106 glaucoma patients had their IOPs measured by 2 fellowship trained ophthalmologists. Data were analyzed using the Bland-Altman method of differences. Correlation was measured using the Pearson coefficient. Results: Most of our patients were Chinese (88.7%) and female (51.9%). The average age was 66.9 years. The range of IOPs as measured by GAT was 2 to 58 mm Hg. Using the Bland-Altman method to compare GAT and SUOER SW-500 Rebound Tonometer. The tonometer overestimated the IOP by 0.5 mm Hg in the right eye and underestimated it by 0.1 mm Hg in the left eye. Overall, the tonometer overestimated the IOP by 0.2 mmHg. The Tonometer IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.89 (p < 0.001) for the right eye and 0.86 (p < 0.001) for the left eye, respectively. In patients with GAT IOP >= 21 mm Hg (n = 25), the Tonometer underestimated the IOP by 2.96 mm Hg. Discussion: The IOP measurements from the SUOER SW-500 Rebound Tonometer correlates well with the conventional GAT in measuring the IOP within normal ranges of IOP. SUOER SW-500 Rebound Tonometer may be of use, especially if the risk of transmission of infection is high considering that the probes are disposable. It is easy to use and its small size and portability makes it useful in situations where the patient is unable to be examined at the slit lamp.
引用
收藏
页数:7
相关论文
共 28 条
  • [21] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770
  • [22] A comparison of intraocular pressure values obtained with the Tono-Pachymeter NT530P iCare (R) rebound tonometer and Goldmann applanation tonometer
    Rampersad, N.
    Mashige, K. P.
    Jhetamt, S.
    [J]. AFRICAN VISION AND EYE HEALTH JOURNAL, 2011, 70 (03):
  • [23] Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer
    Mana Hagishima
    Kazutaka Kamiya
    Fusako Fujimura
    Tetsuya Morita
    Nobuyuki Shoji
    Kimiya Shimizu
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 257 - 262
  • [24] Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer
    Hagishima, Mana
    Kamiya, Kazutaka
    Fujimura, Fusako
    Morita, Tetsuya
    Shoji, Nobuyuki
    Shimizu, Kimiya
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (02) : 257 - 262
  • [25] Comparison of intraocular pressure adjusted by central corneal thickness or corneal biomechanical properties as measured in glaucomatous eyes using noncontact tonometers and the Goldmann applanation tonometer
    Yaoeda, Kiyoshi
    Fukushima, Atsushi
    Shirakashi, Motohiro
    Fukuchi, Takeo
    [J]. CLINICAL OPHTHALMOLOGY, 2016, 10 : 829 - 834
  • [26] Measurement of intraocular pressure using the Tono-Pen in comparison with Goldmann applanation tonometry - a clinical study in 100 eyes
    Deuter, CME
    Schlote, T
    Hahn, GA
    Bende, T
    Derse, M
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2002, 219 (03) : 138 - 142
  • [27] Measurement of intraocular pressure (IOP) in chickens using a rebound tonometer: Quantitative evaluation of variance due to position inaccuracies
    Prashar, A.
    Guggenheim, J. A.
    Erichsen, J. T.
    Hocking, Pm.
    Morgan, J. E.
    [J]. EXPERIMENTAL EYE RESEARCH, 2007, 85 (04) : 563 - 571
  • [28] Measurement of the difference in intraocular pressure between the sitting and lying body positions in healthy subjects: direct comparison of the Icare Pro with the Goldmann applanation tonometer, Pneumatonometer and Tonopen XL
    Barkana, Yaniv
    Gutfreund, Shay
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 42 (07) : 608 - 614